- 2018-06-20FDA-DeviceDatascope CorporationClass IIMaquet Cardiopulmonary Bypass Custom Tubing kit BO-TOP 38600 ADULT CPB, Part Number 701056035 Product Usage: The Maquet Cardiopulmonary Bypass Custom Tubing kit is a sterile, non-pyrogenic device for single use only and is not to be re-sterilized by the user. The product consists of tubing and connectors with connected oxygenators, reservoirs, filters and other cardiopulmonary bypass components assembled into user specified circuits for procedures requiring extracorporeal support for up to s
There is a potential for the over-pressure safety (OPS) valve in the tubing kits to not allow flow through.
- 2018-06-20FDA-DeviceZimmer Biomet, Inc.Class IIBiomet Modular Primary Tibial Tray Implants; Modular Interlok Primary Tray (size 75 mm); Item No. 141215; UDI: (01) 00880304005310 (17) 280306 (10) 385340 Product is used in Total Knee Arthroplasty
One lot of 75mm tibial tray is potentially etched and labeled as a 79 mm tibial tray. This may lead to delay in surgery, or immediate revision surgery to remove an incorrectly sized tibial component after it has been cemented.
- 2018-06-20FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrilliance 16 P with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728246. Computed Tomography X-ray system
The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.
- 2018-06-20FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrilliance 16 with DEPMED HARDENING KIT 12NC: 453567400741, Model Number 728240. Computed Tomography X-ray system
The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.
- 2018-06-20FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrilliance 64 with DEPMED HARDENING KIT 12NC: 459800191351, Model Number 728231. Computed Tomography X-ray system
The patient support head holder could contact the back ISO shelter wall during manual or motorized motion through the gantry bore. This may result in a serious injury that would require medical intervention to preclude permanent impairment in certain scenarios.
- 2018-06-20FDA-DeviceIon Beam Applications S.A.Class IIProteus 235 The PTS is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.
IBA is recalling to provide information through a Safety Notice to reduce risk for maintenance technicians during ion source replacement.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIOxidase Reagent (ref: 55635): This test is used to detect the production of the enzyme cytochrome oxidase by bacteria. This enzyme is characteristic of the genus Neisseria and most species of Pseudomonas. It enables the differentiation of Gram (-) bacilli (1, 2). A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips i
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIColor Gram 2 (COLOR GRAM 2 - F), these stains are used to stain bacterial and fungal microorganisms for direct examination, according to the Gram method. Dye and chemical solution stains for medical purposes are mixtures of synthetic or natural dyes or nondye chemicals in solutions used in staining cells and tissues for diagnostic histopathology, cytopathology, or hematology.
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIThe PREVI¿ Color Gram dyes are used with the PREVI¿ Color Gram instrument to stain bacterial and fungal microorganisms for direct examination, according to Gram s method. The PREVI¿ Color Gram instrument sprays the PREVI¿ Color Gram reagents onto microscope slides which have been smeared with specimens to be examined. The staining procedure is rapid and precise. Each reagent has a specific position (A, B, C, D, E) on the PREVI¿ Color Gram platform that is also indicated on the reagent label.
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIVIDAS TOXO IgG Avidity is an automated qualitative test for use on the VIDAS family instruments, for the determination of anti-toxoplasma IgG avidity in human serum or plasma (lithium heparin, sodium citrate, EDTA) using the ELFA technique (Enzyme Linked Fluorescent Assay). Toxoplasma gondii serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to Toxoplasma gondii in serum. Additionally, some of these reagents consist of anti
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIThe VITEK 2 Gram Negative Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of clinically significant aerobic gram negative bacilli to antimicrobial agents. Fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolat
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIThe VITEK¿ 2 Gram Positive Susceptibility Card is intended for use with the VITEK 2 Systems in clinical laboratories as an in vitro test to determine the susceptibility of Staphylococcus spp., Enterococcus spp., and S. agalactiae to antimicrobial agents. Fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIThe VITEK¿ 2 Anaerobic and Corynebacteria identification card (ANC) is intended for use with the VITEK¿ 2 Systems for the automated identification of most clinically significant anaerobic organisms and Corynebatcerium species. The VITEK¿ 2 ANC identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and p
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIThe VITEK 2 Neisseria-Haemophilus identification card (NH) is intended for use with the VITEK 2 Systems for the automated identification of most clinically significant fastidious organisms. The VITEK 2 NH identification card is a single-use disposable. A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips impregnated
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceBioMerieux SAClass IIThe VITEK 2 Gram-Positive identification card (GP) is intended for use with the VITEK 2 Systems for the automated identification of most significant Gram-positive organisms. The VITEK 2 GP identification card Is a single-use disposable. GP TEST KIT VTK2 20 CARTES A microorganism differentiation and identification device is a device intended for medical purposes that consists of one or more components, such as differential culture media, biochemical reagents, and paper discs or paper strips
A shipment to Guadeloupe suffered a temperature excursions up to 47.8¿C, exceeding the acceptable limits. After detection of the issue, Global Supply Chain QA asked the bioM¿rieux distributor in Guadeloupe to put products in quarantine in July 2017; After several reminders, the distributor informed in November 2017 that products were delivered to customers.
- 2018-06-20FDA-DeviceSmith & Nephew, Inc.Class IIsmith&nephew GENESIS(R) II TIBIAL Punch, 13 MM, REF 71440408, QTY: (1)
One lot of GENESIS II 13 mm tibial punches are 2 mm smaller than specification.
- 2018-06-20FDA-DeviceKeystone Dental IncClass IIPrima Plus ¿4.1 x 11.5mm Implant, Catalog Number 15737K Keystone Dental implants are intended for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including cement retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework.
The implants are mislabeled.
- 2018-06-20FDA-DeviceMaquet Cardiopulmonary AgClass IIAortic Arterial Cannula 24 Fr (O.D.); 23 cm (L); with 3/8-inch connector with LL, Part Number 701002267, Article Number A24-7107. For use in cardiopulmonary bypass
This lot of arterial cannula may have been supplied with 3/8-width de-airing plugs that could break during the process of removal from the arterial cannula.
- 2018-06-20FDA-DeviceMaquet Cardiopulmonary AgClass IIAortic Arterial Cannula 22 Fr (O.D.); 23 cm (L); with 3/8-inch connector with LL, Part Number 701002246, Article Number A22-7107. For use in cardiopulmonary bypass
This lot of arterial cannula may have been supplied with 3/8-width de-airing plugs that could break during the process of removal from the arterial cannula.
- 2018-06-20FDA-DeviceMaquet Cardiopulmonary AgClass IIAortic Arterial Cannula 20 Fr (O.D.); 23 cm (L); with 3/8-inch wide connector, Part Number 701002206, Article Number A20-0101. For use in cardiopulmonary bypass
This lot of arterial cannula may have been supplied with 3/8-width de-airing plugs that could break during the process of removal from the arterial cannula.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK, TUBE RESERVOIR, PEEP VALVE AND 7 FT. OXYGEN TUBING, REF CPRM1126P
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK, TUBE RESERVOIR, FILTER, PEEP VALVE, CO2 AND 7 FT. OXYGEN TUBING, REF CPRM1126FPC
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK, TUBE RESERVOIR, FILTER, PEEP VALVE AND 7 FT. OXYGEN TUBING, REF CPRM1126FP
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH LARGE ADULT MASK, TUBE RESERVOIR, FILTER AND 7 FT. OXYGEN TUBING, REF CPRM1126FL
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK, TUBE RESERVOIR, FILTER AND 7 FT. OXYGEN TUBING, REF CPRM1126F
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK, TUBE RESERVOIR, CO2 AND 7 FT. OXYGEN TUBING, REF CPRM1126C
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK, TUBE RESERVOIR AND 7 FT. OXYGEN TUBING, REF CPRM1126
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK 5, BAG RESERVOIR, PEEP VALVE AND 7 FT. OXYGEN TUBING, REF CPRM1116PD5
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK, BAG RESERVOIR, PEEP VALVE, CO2 AND 7 FT. OXYGEN TUBING, REF CPRM1116PC
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.
- 2018-06-20FDA-DeviceSun Med, LLCClass IIMEDLINE ADULT MANUAL RESUSCITATOR WITH MEDIUM ADULT MASK, BAG RESERVOIR, PEEP VALVE AND 7 FT. OXYGEN TUBING, REF CPRM1116P
There is a potential for the patient port retaining ring to not fully seat allowing the patient port to detach during use. This may impact air delivery to the patient.