Medical device recalls
39229 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2018-05-09FDA-DevicePhilips Medical Systems NederlandsClass IIMutlidiagnost-Eleva with Flat Detector Eleva with Flat Detector 708037 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- 2018-05-09FDA-DevicePhilips Medical Systems NederlandsClass IIMutlidiagnost-Eleva with Flat Detector 708034 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- 2018-05-09FDA-DevicePhilips Medical Systems NederlandsClass IIMultiDiagnost Eleva 708036 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- 2018-05-09FDA-DevicePhilips Medical Systems NederlandsClass II708033 URODiagnost As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- 2018-05-09FDA-DevicePhilips Medical Systems NederlandsClass IIMultiDim Eleva 708032 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.
The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.
- 2018-05-09FDA-DeviceSouthwest Technologies IncClass IIHYPOTHERMIA SLIPPERS Supportive in Cancer, packaged under the following brands: (a) Elasto-Gel(TM), SL3000 (b) Elasto-Gel(TM), SL3000LV (c) Elasto-Gel(TM), SL3001 (d) Elasto-Gel(TM), SL3007 (e) Elasto-Gel(TM), SL3008
Hypothermia mitts, hypothermia slippers, and hypothermia caps on the market are misbranded for use during chemotherapy treatments. Additionally, Southwest Technologies, Inc. was informed of 2 events of frostbite to digits in the last year while using the hypothermia mitts.
- 2018-05-09FDA-DeviceSouthwest Technologies IncClass IIHYPOTHERMIA MITTS Supportive in Cancer, packaged under the following brands: (a) Elasto-Gel(TM), TM7006 (b) Elasto-Gel(TM), TM7006LV (c) Elasto-Gel(TM), TM7007 (d) Elasto-Gel(TM), TM7008 (e) Elasto-Gel(TM), TM7008b (f) Elasto-Gel(TM), TM7009 (g) Elasto-Gel(TM), TM7010 (h) Elasto-Gel(TM), TM7011 (i) Elasto-Gel(TM), TM7012
Hypothermia mitts, hypothermia slippers, and hypothermia caps on the market are misbranded for use during chemotherapy treatments. Additionally, Southwest Technologies, Inc. was informed of 2 events of frostbite to digits in the last year while using the hypothermia mitts.
- 2018-05-09FDA-DeviceSouthwest Technologies IncClass IIHYPOTHERMIA CAP Supportive in Cancer, packaged under the following brands: (a) Elasto-Gel(TM), CAP610 (b) Elasto-Gel(TM), CAP610LV (c) Elasto-Gel(TM), CAP613 (d) Elasto-Gel(TM), CAP617 (e) Elasto-Gel(TM), CAP619 (f) Elasto-Gel(TM), CAP619LV (g) Elasto-Gel(TM), CAP622
Hypothermia mitts, hypothermia slippers, and hypothermia caps on the market are misbranded for use during chemotherapy treatments. Additionally, Southwest Technologies, Inc. was informed of 2 events of frostbite to digits in the last year while using the hypothermia mitts.
- 2018-05-09FDA-DeviceStryker Medical Division of Stryker CorporationClass II6381 M-1 Cot Fastener, catalog no. 6381000000
The Operations Manual and Maintenance Manual are missing various warnings and contain incorrect information.
- 2018-05-09FDA-DeviceStryker Medical Division of Stryker CorporationClass II6376 M-1 Cot Fastener, catalog no. 6376000000
The Operations Manual and Maintenance Manual are missing various warnings and contain incorrect information.
- 2018-05-09FDA-DeviceSiemens Medical Solutions USA, Inc.Class IIBiograph Horizon - PET/CT, PETsyngo VJ20A Software Nuclear medicine/ xray diagnostic scanner.
Error introduced into PET images acquired and reconstructed with VJ20A software. Array values are indexed improperly when the norm file is created during QC. During data reconstruction, incorrect values are being applied. This can lead to a gradient in the image. The severity of the error is directly related to the positioning of the PET QC phantom relative to the center of the field of view.
- 2018-05-09FDA-DeviceCook Inc.Class IIStoneBreaker Single Use Probe, Catalog numbers: SBP-010500; SBP-010605; SBP-016500; SBP-016605; SBP-020425 GPN : G52607; G52605; G52606; G52609; G52608
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- 2018-05-09FDA-DeviceCook Inc.Class IIStoneBreaker C02 Cartridge, Catalog number: SBC-10; GPN: G52603; UDI: (01)10827002526034
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- 2018-05-09FDA-DeviceCook Inc.Class IIStoneBreaker Sterilization Cap, Catalog number: SBA-SC; GPN: G52602; UDI: (01)00827002526020
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- 2018-05-09FDA-DeviceCook Inc.Class IIStoneBreaker Probe Cap, Catalog number:SBA-PC; GPN: G52601; UDI:(01)00827002526013
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- 2018-05-09FDA-DeviceCook Inc.Class IIStoneBreaker Exhaust Line, Catalog number: SBA-EL; GPN: G52600; UDI: (01)00827002526006
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- 2018-05-09FDA-DeviceCook Inc.Class IIStoneBreaker Exhaust Cap, Catalog number: SBA-EC; GPN: G52599; UDI: (01)00827002525993
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- 2018-05-09FDA-DeviceCook Inc.Class IIStoneBreaker Pneumatic Lithotripter, Catalog number: SBL-KIT1; GPN: G52604 UDI:(01)00827002526044
The interior of the product may not be sterilized to the appropriate Sterility Assurance Level when following the reprocessing instructions in the Instructions for Use.
- 2018-05-09FDA-DevicePhilips Electronics North America CorporationClass IIAllura Xper R8.x.25.5 and UNIZ systems (only with a FlexVision large screen monitor), Interventional fluoroscopic x-ray system Product Usage: The Allura Xper ED series is intended for use on human patients to perform: Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, eg peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placement, embolization and thrombolysis.
After continuous operation for more than one and a half days, the image on the large screen monitor may freeze for approximately 15 seconds after which the system will restore itself.
- 2018-05-09FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAnalyzer and Atellica IM 1600 Analyzer system for in vitro diagnostic testing of clinical specimens.
When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers
- 2018-05-09FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica Solution and Atellica Sample Handler Prime for in vitro diagnostic testing of clinical specimens.
When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers
- 2018-05-09FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtelia IM 1300 Analyzer for in vitro diagnostic testing of clinical specimens.
When more than one immunoassay analyzer is connected to a single solution system, reporting and conversion factors may not be applied to all analyzers
- 2018-05-09FDA-DeviceHitachi Ltd., Medical System Operations GroupClass IIMHI-TM2000 Linear Accelerator System Product Usage: MHI-TM2000 Linear Accelerator System is intended for radiation therapy of lesions, tumors and conditions anywhere in the body where radiation treatment is indicated.
Due to a system controller software anomaly, the patient positioning deviation correction may not be applied and may result in the wrong part of the patient being irradiated.
- 2018-05-09FDA-DeviceCooperSurgical, Inc.Class II5 mm Apple-Hunt Secondary Cannula/Pyramidal Tip Trocar - extended length (5 count); Model Number 900-840 Product Usage: The Apple-Hunt Secondary Cannulas are indicated for use in laparoscopic surgical procedures to provide a pathway for the introduction of endoscopic surgical instruments and allow for maintenance of pneumoperitoneum. For uninhibited laparoscopic access, the Apple-Hunt Cannula/Trocar provides greater instrument maneuverability while maintaining an airtight seal -with no bulky
Products have been packaged with the wrong Trocar length type.
- 2018-05-09FDA-DeviceDeRoyal Industries IncClass IIDeRoyal(R) SURGI MATE(R) Electrosurgical Pencil, Button Switch, with 15 ft (4.6 m) Cord and Holster, REF 88-001502 Product Usage: general surgical
Non-sterile product was packaged inside a case that was labeled as sterile.
- 2018-05-09FDA-DeviceRadiometer Medical ApSClass IIAQURE, Software version 2.3.0 and 2.3.1 Product Usage: The AQURE system is intended to let allow the management of analytical devices and operator profiles. The user can associate patient data with test data. The system shows test results. The system receives data from connected devices at the point-of care or laboratory. It can send test results to the HIS/LIS. The system lets the user send commands to selected devices. The system uses data related to the performance of devices, to tell use
There is a potential problem relating to the AQURE System, versions 2.3.0 and 2.3.1, that may result in patient mix-up.
- 2018-05-09FDA-DeviceCardinal Health 200, LLCClass IIMedi-Vac(TM) Flex Advantage(TM) Suction Canister Liner, Material Numbers: a) REF 6561-910C b) REF 6561-916K c) REF 6561-920C To remove blood and/or body fluids. These fluids may be suctioned during surgical procedures or other applications requiring the removal of built up fluids. The suction canisters are used in operating rooms but are also used outside the surgical field as well as post-operatively within patient suites. The collection containers hold the fluids until they are ready for
products may be missing a one-way valve on the fluid patient port.
- 2018-05-09FDA-DeviceBecton Dickinson & Co.Class IIIBD MAX DNA MMK Lab Use, catalog no. 442828
The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.
- 2018-05-09FDA-DeviceBecton Dickinson & Co.Class IIIBD MAX DNA MMK (SPC) For Laboratory Use, catalog no. 442829
The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.
- 2018-05-09FDA-DevicePolymer Technology Systems, Inc.Class IIPTS Panels¿ CHOL+GLU Test Strips (REF 1765)
Potential for loss of activity of the glucose analyte (under-recovery) prior to the full stated shelf life (August 2018)