- 2018-04-04FDA-DeviceCovidien LLCClass IISignia" Power Handle (SIGPHANDLE, SIGNVAL1 and SIGVAL20 - Kits) The Signia" Power Handle, when used with Endo GIA" single-use reloads and Endo GIA" single-use reloads with Tri-Staple" Technology, has applications in open and minimally invasive general abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis.
Product reportedly shutting down during use.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIPhoenix Nail System, intended for use with an intramedullary nail system for alignment, stabilization and fixation of fractures. 3.5mm Inserter Connector, Short; Model Number: 14-441045; Kit Model Numbers: 14-400082, 14-400082S, 14-440065S, 14-442000-00, 14-442000S, 14-444000-00, 979105, 14-441045-00, and 14-444000S
Certain lots of the Phoenix Tibia Nail 3.5mm Inserter instruments are being recalled due to reports of fracturing and remaining in the device implant.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System STER L H TROC NAIL 320X9X125 STER L H TROC NAIL 340X9X125 STER L H TROC NAIL 360X9X125 STER L H TROC NAIL 380X9X125 STER L H TROC NAIL 400X9X125 STER L H TROC NAIL 420X9X125 STER L H TROC NAIL 440X9X125 STER L H TROC NAIL 460X9X125 STER L H TROC NAIL 320X11X125 STER L H TROC NAIL 340X11X125 STER L H TROC NAIL 360X11X125 STER L H TROC NAIL 380X11X125 STER L H TROC NAIL 400X11X125 STER L H TROC NAIL 420X11X125 STER L H TROC NAIL 440X11X125 STER L H TROC
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System ATN LAG SCREW 60MM STERILE ATN LAG SCREW 65MM STERILE ATN LAG SCREW 70MM STERILE ATN LAG SCREW 75MM STERILE ATN LAG SCREW 80MM STERILE ATN LAG SCREW 85MM STERILE ATN LAG SCREW 90MM STERILE ATN LAG SCREW 95MM STERILE ATN LAG SCREW 100MM STERILE ATN LAG SCREW 105MM STERILE ATN LAG SCREW 110MM STERILE ATN LAG SCREW 115MM STERILE ATN LAG SCREW 120MM STERILE ATN LAG SCREW 125MM STERILE ATN LAG SCREW 130MM STERILE The Trochanteric Nail System is inten
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System ATN AR SCREW 60MM STERILE ATN AR SCREW 65MM STERILE ATN AR SCREW 70MM STERILE ATN AR SCREW 75MM STERILE ATN AR SCREW 80MM STERILE ATN AR SCREW 85MM STERILE ATN AR SCREW 90MM STERILE ATN AR SCREW 95MM STERILE ATN AR SCREW 100MM STERILE The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of the
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System STER TROC NAIL 180X13X140 STER TROC NAIL 125 DEG X 13MM STER TROC NAIL 130 DEG X 13MM STER TROC NAIL 135 DEG X 13MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated wi
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System TROCH NAIL ST 13X180 125DG STRL TROCH NAIL ST 13X180 130DG STRL TROCH NAIL ST 13X180 135DG STRL The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, patho
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System STER TROC NAIL 180X11X140 STER TROC NAIL 125 DEG X 11MM STER TROC NAIL 130 DEG X 11MM STER TROC NAIL 135 DEG X 11MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with s
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System TROCH NAIL ST 11X180 125DG STRL TROCH NAIL ST 11X180 130DG STRL TROCH NAIL ST 11X180 135DG STRL The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, patho
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System STER TROC LAG SCR 70 MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fractures in osteoporotic bone of the trochanteric and diaphyseal ar
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System STER TROC NAIL 125 DEG X 9MM STER TROC NAIL 130 DEG X 9MM STER TROC NAIL 135 DEG X 9MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures associated with shaft fractures, pathologic fra
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System STER TROC END CAP STER TROC END CAP 5MM STER TROC END CAP 10MM STER TROC END CAP 15MM STER TROC END CAP 20MM STER TROC END CAP 25MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additionally indicated to treat pertrochanteric fractures
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIACE Trochanteric Nail System ACE TROCH NAIL 125 DEG. X 11MM ACE TROCH NAIL 130 DEG. X 11MM ACE TROCH NAIL 135 DEG. X 11MM ACE TROCH NAIL 125 DEG. X 13MM ACE TROCH NAIL 130 DEG. X 13MM ACE TROCH NAIL 135 DEG. X 13MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is add
Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
- 2018-04-04FDA-DeviceZimmer Biomet, Inc.Class IIPERSONA¿ THE PERSONALIZED KNEE SYSTEM "PSN TIB STM 5 DEG SZ D R" indicated for patients with severe knee pain and disability due to: - Rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
A complaint was received for Item 42-5320-067-02, lot 63236047. It was report that there was no set screw or taper plug included in the box. Complaint investigation confirmed that the entire lot of 63236047 was missing the set screw and taper plug.
- 2018-04-04FDA-DeviceCook Vascular Inc.Class IIVital-Port Vascular Access System, Subcutaneous, implanted, intravascular infusion port and catheter, IP, IP-S, Product usage: The devices are intended for use in patient therapy requiring long-term vascular access for infusion therapy and/or blood sampling.
Non-coring needle provided with the Cook Vital-Port¿ Vascular Assess System (Vital-Port) may cut or dislodge a core or sliver of material from the Vital-Port septum when the non-coring needle is inserted into the Vital-Port. This needle is used on the initial implant of the Vital-Port. Vital-Port products that have been successfully placed in patients are not impacted by this recall.
- 2018-04-04FDA-DeviceInova Diagnostics IncorporatedClass IIQUANTA-Lyser 240 EIA, Part #GS0241. The instrument is a fully automated, high-throughput, bench top system designed for clinical use to automate slide and reagent barcode reading, sample dilution and distribution, reagent pipetting and dispensing, slide and plate washing, dispensing slide mounting medium in the preparation for microscope analysis, and photometric measurement of microwell plates.
The device power supply was not manufactured according to specification which can result in the loss of the protective grounding of the instrument.
- 2018-04-04FDA-DeviceInova Diagnostics IncorporatedClass IIQUANTA-Lyser 240 IFA, Part #GS0242. The instrument is a fully automated, high-throughput, bench top system designed for clinical use to automate slide and reagent barcode reading, sample dilution and distribution, reagent pipetting and dispensing, slide and plate washing, dispensing slide mounting medium in the preparation for microscope analysis, and photometric measurement of microwell plates.
The device power supply was not manufactured according to specification which can result in the loss of the protective grounding of the instrument.
- 2018-04-04FDA-DeviceDePuy Orthopaedics, Inc.Class II(1) Sigma HP Cemented Trochlea Size 1 Narrow Left, Catalog Number: 102403100, GTIN: 10603295001614; (2) Sigma HP Cemented Trochlea Size 2 Narrow Left, Catalog Number: 102403200, GTIN: 10603295001621; (3) Sigma HP Cemented Trochlea Size 3 Narrow Left, Catalog Number: 102403300, GTIN: 10603295001638; (4) Sigma HP Cemented Trochlea Size 4 Narrow Left, Catalog Number: 102403400, GTIN: 10603295001645; (5) Sigma HP Cemented Trochlea Size 5 Narrow Left, Catalog Number: 102403500, GTIN: 10603295001652;
The SIGMA¿ HP PFJ Cemented Trochlear Implants, a standalone component of the partial knee system is being recalled based on elevated revision rates observed as part of the company s post market surveillance process.
- 2018-04-04FDA-DeviceDeerfield Imaging, Inc.Class IIIMRIS ORT300 Removable Operating Room Table, 50-60 Hz Version is intended for use during diagnostic examinations or surgical procedures to support and position a patient.
The rotational locking mechanisms on the ORT 200/300 tables fail to operate such that it cannot change states between locked and unlocked.
- 2018-04-04FDA-DeviceDeerfield Imaging, Inc.Class IIIMRIS ORT200 Removable Operating Room Table, 50-60 Hz Version is intended for use during diagnostic examinations or surgical procedures to support and position a patient.
The rotational locking mechanisms on the ORT 200/300 tables fail to operate such that it cannot change states between locked and unlocked.
- 2018-04-04FDA-DeviceStryker Instruments Div. of Stryker CorporationClass IIThe Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a pathway into the pedicle. The instrument has three types of buttons to remotely control the software: the UP, DOWN and SELECT buttons. The Pedicle Feeler is laser marked in 10 mm intervals to help indicate the depth in which the feeler tip has been inserted into the pedicle. Next to the navigational information, these depth indicators on the feeler tip are a
Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker¿ Navigation System because the settings on the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break.
- 2018-04-04FDA-DeviceKonica Minolta Medical Imaging USA, Inc.Class IISonimage HS1 Ultrasound Kit AC adapter, Product Number: A8AR, UDI: 04560141941588
There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.
- 2018-04-04FDA-DeviceKonica Minolta Medical Imaging USA, Inc.Class IISonimage HS1 Ultrasound Kit AC adapter, Product Number: 1. A9RP, UDI: 04560141945463; 2. AAEU, UDI: 04560141946385; 3. A9RR, UDI: 04560141945470
There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.
- 2018-04-04FDA-DeviceKonica Minolta Medical Imaging USA, Inc.Class IISonimage HS1 Ultrasound Kit AC adapter, Product Number: A8WA, UDI: 04560141944534
There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.
- 2018-04-04FDA-DeviceKonica Minolta Medical Imaging USA, Inc.Class IISonimage HS1 Ultrasound Kit AC adapter, Product Number: A7AP, UDI: 04560141940031
There is a risk of circuit board breakage within the main housing of the AC Adapter unit. Smoke may be emitted from the AC Adapter as a direct result.
- 2018-04-04FDA-DeviceJohnson & Johnson Vision Care, Inc.Class II1-DAY ACUVUE Moist for ASTIGMATISM Brand Contact Lenses
Some contact lenses were associated with complaints of a foreign matter caught between the blister package and foil.
- 2018-04-04FDA-DeviceJohnson & Johnson Vision Care, Inc.Class IIACUVUE ADVANCE Brand Contact Lenses with HYDRACLEAR
Some contact lenses were associated with complaints of a foreign matter caught between the blister package and foil.
- 2018-04-04FDA-DeviceJohnson & Johnson Vision Care, Inc.Class IIACUVUE OASYS Brand Contact Lenses with HYDRACLEAR PLUS
Some contact lenses were associated with complaints of a foreign matter caught between the blister package and foil.
- 2018-04-04FDA-DeviceCovidien LLCClass IIRechargeable Li-ion battery used with Puritan Bennett TM 980 Series Ventilator System including Pediatric-Adult Ventilator, Neonatal Ventilator, Universal Ventilator; Battery Product Number: 10086042; UDI: 20884521700861 The rechargeable lithium-ion batteries are intended to provide a backup power source to the ventilator in case AC power is lost.
Rechargeable lithium-ion batteries with incorrect firmware that are used in certain Puritan Bennett 980 ventilators may not fully charge after installation.
- 2018-04-04FDA-DeviceMedtronic IncClass I6F Taiga Guiding Catheter
Complaints regarding 6F Taiga guide catheter tip detachment and/or tip splitting.