- 2017-04-19FDA-DevicePerkinElmer Life and Analytical Sciences, Wallac, OYClass IISpecimen Gate Laboratory; Product Number: 5002-0180 Specimen Gate Laboratory is intended for use as a data processing software used in the storage, retrieving, and processing of laboratory data.
Potential for errors in patient results generated by the Specimen Gate Laboratory Software.
- 2017-04-19FDA-DeviceConMed CorporationClass IIHYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. --- There is no problem with the Hyfrecator 2000 Electrosurgical Units or with cases of P/N 7-796-19 sold in cases as P/N 7-796-19BX or P/N 7-796-19CS.
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were Sterile.
- 2017-04-19FDA-DeviceConMed CorporationClass IISterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons, 230 VOLTS AC, REF/Catalog Number 7-900-230, Rx Only. There is no problem with the Hyfrecator 2000 Electrosurgical Units. -- ASSEMBLED IN MEXICO -- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
- 2017-04-19FDA-DeviceConMed CorporationClass IISterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons , 220 VOLTS AC, REF/Catalog Number 7-900-220, Rx Only, There is no problem with the Hyfrecator 2000 Electrosurgical Units --- ASSEMBLED IN MEXICO -- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502 ---
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
- 2017-04-19FDA-DeviceConMed CorporationClass IISterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons, 115 VOLTS AC, REF/Catalog Number 7-900-115, Rx Only . There is no problem with the Hyfrecator 2000 Electrosurgical Units. -- ASSEMBLED IN MEXICO --- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502 The Hyfrecator 2000 Electrosurgical Unit is indicated for use in conjunction with an electrosurgical accessory handpiece for delivery of high frequency
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
- 2017-04-19FDA-DeviceConMed CorporationClass IISterile Handpiece Sheaths, P/N 7-796-19, included in the Accessory Packages contained in the HYFRECATOR 2000 Electrosurgical Unit (ESU) cartons, 100 VOLTS AC, REF/Catalog Number 7-900-100, Rx Only. There is no problem with the Hyfrecator 2000 Electrosurgical Unit. -- ASSEMBLED IN MEXICO --- CONMED CORPORATION 525 FRENCH ROAD UTICA, NY 13502
For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were sterile.
- 2017-04-19FDA-DeviceZimmer Biomet, Inc.Class IIMEDTECH ROSA Spine 1.0.2 The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments during spine surgery. Guidance is based on an intra-operative plan developed with three dimensional imaging software provided that the required fiducial markers and rigid patient anatomy can be identified on 3D CT scans. The device is indicated for the placement of pedicle screws in lumbar vertebr
Unapproved change made by the supplier.
- 2017-04-19FDA-DeviceZimmer Biomet, Inc.Class IIMEDTECH ROSA Brain 3.0 The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurological procedure in which the use of stereotactic surgery may be appropriate.
Unapproved change made by the supplier.
- 2017-04-19FDA-DeviceMedtronic NeurosurgeryClass IThe Medtronic StrataMR adjustable valves and shunts are designed to provide continuous cerebrospinal fluid (CSF) flow from the ventricles of the brain into the right atrium of the heart or the peritoneal cavity. The design enables the physician to non-invasive adjust valve pressure/performance level pre- and post-implantation by using magnetic adjustment tools without the need for radiographic confirmation. After surgical implantation, the Medtronic Neurosurgery StrataMR valves provide a non
Incorrect valve rotor orientation due to misalignment and/or off-center positioning of the StrataMR Locator Tool: When the StrataMR valve is set in the incorrect orientation, higher valve opening pressures result leading to a corresponding underdrainage in CSF fluid performance. This condition may result in the following adverse health consequences: ventriculomegaly, headaches, nausea, vomiting and lethargy. If left untreated, underdrainage has the potential to lead to coma and death.
- 2017-04-12FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIIBrightView X 3/8 model number 882478 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only). BrightV
The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
- 2017-04-12FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrightView X 3/4 model number 882479 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only). BrightV
The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
- 2017-04-12FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrightView XCT model number 882482 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only). BrightVie
The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
- 2017-04-12FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIBrightView SPECT model number 882480 BrightView and BrightView X are gamma cameras designed for single or dual detector nuclear imaging accommodating a broad range of Emission Computed Tomography (ECT) studies. The cameras can be used to perform planar static, dynamic, gated, and total body studies, as well as circular-orbit and noncircular orbit SPECT studies, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors only).
The brake hub was not engaging with the gear box shaft due to an error in the assembly of components within the brake hub mechanism after service, causing the detector to backdrive.
- 2017-04-12FDA-DeviceSiemens Healthcare Diagnostics, IncClass IISiemens ADVIA Centaur BR assay Cat No. 09010686, SMN 10340081 (5-pack/250 tests - REF) BR Assay for CA 27.29 The ADVIA Centaur¿ BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur¿, ADVIA Centaur¿ XP, and ADVIA Centaur¿ XPT systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of
Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems
- 2017-04-12FDA-DeviceSiemens Healthcare Diagnostics, IncClass IISiemens ADVIA Centaur BR assay Cat No. 03896216, SMN 10334837 (5-pack/250 tests) BR Assay for CA 27.29 The ADVIA Centaur¿ BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur¿, ADVIA Centaur¿ XP, and ADVIA Centaur¿ XPT systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patie
Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems
- 2017-04-12FDA-DeviceSiemens Healthcare Diagnostics, IncClass IISiemens ADVIA Centaur BR assay Cat No. 02419937, SMN 10333349 (1-pack/50 tests) BR Assay for CA 27.29 The ADVIA Centaur¿ BR assay is an in vitro diagnostic test for the quantitative serial determination of cancer antigen CA 27.29 in human serum using the ADVIA Centaur¿, ADVIA Centaur¿ XP, and ADVIA Centaur¿ XPT systems. The test is intended for use as an aid in monitoring patients previously treated for Stage II or Stage III breast cancer. Serial testing for CA 27.29 in the serum of patien
Potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) a assay on the ADVIA Centaur Systems
- 2017-04-12FDA-DeviceBecton Dickinson & CompanyClass IIBD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tubes, 4.5 mL, 13 x 75 mm, Mfd: Becton Dickinson and Company Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection Tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations.
BD is initiating this product correction of multiple lots of BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (13x75 mm x 4.5 mL). A limited number of tubes in these lots may exhibit stopper pullout from the tube as the user removes the needle from the stopper following specimen collection.
- 2017-04-12FDA-DeviceCovidien LLCClass IICovidien Curity Saline Dressing 8' x 4 (20.3cm x10.2 cm), Sterile Item Code: 3606 Curity sodium chloride dressing is saturated in a 20% hypertonic saline solution. The dressing is designed to help promote biological cleansing within the wound and aid in the autolytic debridement process.
Product sterility is compromised due to breach of sterile barrier
- 2017-04-12FDA-DeviceCovidien LLCClass IICovidien Curity Sodium Chloride Dressing , 6 x 6-3/4 (15.2 cm x 17.1 cm), Sterile Item Code: 3339 Curity Sodium Chloride Dressing is indicated for use on exudating and/or infected wounds
Product sterility is compromised due to breach of sterile barrier
- 2017-04-12FDA-DeviceCovidien LLCClass IICovidien Kerlix Super Sponge Saline Dressing, Sterile Item Code: 3338 Curity sodium chloride dressing is saturated in a 20% hypertonic saline solution. The dressing is designed to help promote biological cleansing within the wound and aid in the autolytic debridement process.
Product sterility is compromised due to breach of sterile barrier
- 2017-04-12FDA-DeviceCovidien LLCClass IICovidien Curity Wet Dressing 8 x 4 (20.3 cm x 10.2 cm), Sterile Item Code: 3337 Curity sodium chloride dressing is saturated in a 20% hypertonic saline solution. The dressing is designed to help promote biological cleansing within the wound and aid in the autolytic debridement process.
Product sterility is compromised due to breach of sterile barrier
- 2017-04-12FDA-DeviceCovidien LLCClass IICovidien Curity Eye Pad Item Code: 91650 The sterile eye pad is a device that consists of a pad made of various materials, such as gauze and cotton, intended for use as a bandage over the eye for protection or absorption of secretions. It is supplied sterile to allow for use in the operating room, and for application to the eye following injury and or surgery in order to reduce the likelihood of introduction of infection
Product sterility is compromised due to breach of sterile barrier
- 2017-04-12FDA-DeviceCovidien LLCClass IICovidien Curity Eye Pad Oval, 1-5/8 x 2-5/8 (4.2 cm x 6.7 cm), Sterile Item Code: 2841 The sterile eye pad is a device that consists of a pad made of various materials, such as gauze and cotton, intended for use as a bandage over the eye for protection or absorption of secretions. It is supplied sterile to allow for use in the operating room, and for application to the eye following injury and or surgery in order to reduce the likelihood of introduction of infection
Product sterility is compromised due to breach of sterile barrier
- 2017-04-12FDA-DeviceCovidien LLCClass IICovidien Curity Eye Pad Item Code: 03201 The sterile eye pad is a device that consists of a pad made of various materials, such as gauze and cotton, intended for use as a bandage over the eye for protection or absorption of secretions. It is supplied sterile to allow for use in the operating room, and for application to the eye following injury and or surgery in order to reduce the likelihood of introduction of infection
Product sterility is compromised due to breach of sterile barrier
- 2017-04-12FDA-DeviceAGFA Healthcare Corp.Class IIDX-D600 - DXD Imaging Package
One of the dent markers from the DX-D600 dropped down from the overhead rail.
- 2017-04-12FDA-DeviceAgilent Technologies, Inc.Class IIBond Elut OMS (Dried Matrix Spotting)' Product; Catalog model numbers: 5022-8226 - Auto DBS Card (LCMS Extraction Sys) 50 pk; A400150 - Bond Elut EMS Card 50/Pk; A400150K - Bond Elut DMS Card BULK 500/pk; A400150SK - Bond Elut DMS Card Starter Kit A42001 - Bond Elut MDS Card Accessory Pack Product Usage: For use in the analysis of dried blood spots and other biological matrices in research applications for drug metabolism pharmacokinetics/ absorption, distribution. metabolism and
The product was marketed without 510(k) clearance.
- 2017-04-12FDA-DeviceCenturion Medical Products CorporationClass IICUSTOM CADAVER KIT The Covidien Curity Oval Eye Pads are individually packaged and sterilized by the manufacturer; they are then placed into non-sterile Centurion Convenience Kits along with other components.
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- 2017-04-12FDA-DeviceCenturion Medical Products CorporationClass IIOPHTHALMIC DRESSING TRAY The Covidien Curity Oval Eye Pads are individually packaged and sterilized by the manufacturer; they are then placed into non-sterile Centurion Convenience Kits along with other components.
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- 2017-04-12FDA-DeviceCenturion Medical Products CorporationClass IIPOST-OPERATIVE OPHTHALMIC DRESSING TRAY The Covidien Curity Oval Eye Pads are individually packaged and sterilized by the manufacturer; they are then placed into non-sterile Centurion Convenience Kits along with other components.
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.
- 2017-04-12FDA-DeviceCenturion Medical Products CorporationClass IIEYE KIT The Covidien Curity Oval Eye Pads are individually packaged and sterilized by the manufacturer; they are then placed into non-sterile Centurion Convenience Kits along with other components.
Centurion Medical Products is recalling Convenience Kits that contain Covidien Curity Oval Eye Pads due to supplier's recall because of potential breach in sterile packaging.