Serious riskFDA DeviceClass II

MiniMed 780G loaner pumps recalled for possible insulin dosing errors

Some loaner MiniMed 780G insulin pumps were cleaned with chemically incompatible products that may have damaged the pump case or keypad. This damage could allow moisture to enter the pump, potentially causing it to deliver too much or too little insulin. These devices were distributed as loaners in Europe.

From the FDA notice · Read the official notice ↗ · All Medtronic recalls

What to do

  1. If you have a MiniMed 780G loaner pump, check its serial number against the FDA recall notice
  2. Contact Medtronic or your healthcare provider immediately if your pump is on the recall list
  3. Do not continue using an affected loaner pump without speaking to your provider
  4. Monitor your blood sugar closely if you have been using one of these pumps

Is yours affected?

Compare your label or packaging with the recalled product.

Product
MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW
Sold in
Nationwide
Made for
Adults (0 months and older)

This recall covers 3 products — see all of them.

Details

Units affected
257
Illnesses reported
None reported
Manufactured in
18000 Devonshire St, Northridge, CA, United States
Classification
Class II
Recall number
Z-2982-2026

In the FDA's words

Description
MiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLW Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis. Model/UDI-DI Serial Numbers: MMT-1885CLW/00763000442187, 00763000974374/NG3150737H, NG3155194H, NG3151561H, NG3154693H, NG3151716H, NG3153547H, NG3153637H, NG3155407H, NG3152666H, NG3155788H, NG3081975H, NG2901117H, NG3080842H, NG3081463H, NG3153416H, NG3081894H, NG3153781H, NG3155944H, NG3081982H, NG3081525H, NG3155909H, NG3090722H, NG3154897H, NG3088436H, NG3088708H, NG3090073H, NG3155908H, NG3150720H, NG3092957H, NG3128715H, NG3128784H, NG3152774H, NG3153929H, NG3514566H, NG3150780H, NG3151290H, NG3151508H, NG3507161H, NG3508693H, NG3081976H, NG3151460H, NG3152051H, NG3152085H, NG3152344H, NG3153213H, NG3152772H, NG3155586H, NG3150786H, NG3153214H, NG3153385H, NG3151840H, NG3082019H, NG3153405H, NG3153788H, NG3153908H, NG3154527H, NG3151598H, NG2854971H, NG3155196H, NG3155202H, NG3155341H, NG3091491H, NG3155557H, NG3081498H NG3155688H, NG3155020H, NG3508565H, NG3155419H, NG3507536H, NG3091563H, NG3082036H, NG3154689H, NG3154995H, NG3508504H, NG3514345H, NG3509850H, NG3527505H, NG3514547H, NG3527469H, NG3504333H, NG3504598H, NG3505071H, NG3505220H, NG3505273H, NG3514617H, NG3506727H, NG3085044H, NG3507542H, NG3507547H, NG3514448H, NG3508303H, NG3508350H, NG3508371H, NG3514612H, NG3508511H, NG3508617H, NG3508698H, NG3508707H, NG3508757H, NG3508799H, NG3508903H, NG3509138H, NG3509295H, NG3509299H, NG3509452H, NG3509685H, NG3509689H, NG3509836H, NG3509914H, NG3509925H, NG3514375H, NG3514441H, NG3505246H, NG3505166H, NG3505283H, NG3151570H, NG3154948H, NG3514667H, NG3527165H, NG3527243H, NG3527298H, NG3527468H, NG3527480H, NG3527491H, NG3527580H, NG3527529H, NG3527543H, NG3527571H, NG3527574H, NG3092722H, NG3527581H, NG3527611H, NG3527637H, NG3527642H, NG3527672H, NG3527710H, NG3527737H, NG3527760H, NG3527865H, NG3527918H, NG3527961H, NG3527989H, NG3528017H, NG3528023H, NG3528036H, NG3528061H, NG3528120H, NG3528147H, NG3528156H, NG3528158H, NG3528180H, NG3528218H, NG3528226H, NG3528378H, NG3539964H, NG3539969H, NG3539970H, NG3539999H, NG3540001H, NG3540013H, NG3540029H, NG3540046H, NG3540053H, NG3540084H, NG3540191H, NG3540270H, NG3540310H, NG3540320H, NG3540581H, NG3540604H, NG3540628H, NG3540629H MMT-1886CLW/00763000442194, 00763000974381/NG2966956H, NG2970938H, NG2986362H, NG3013390H, NG3057910H, NG3167493H, NG3466128H, NG3098147H, NG3136398H, NG3137113H, NG3139432H, NG3143920H, NG3144721H, NG3145486H, NG3145488H, NG3471528H, NG3167609H, NG3147120H, NG3147035H, NG3430059H, NG3438451H, NG3464758H, NG3163796H, NG3166778H, NG3409587H, NG3167450H, NG3167623H, NG3167967H, NG3169694H, NG3170253H, NG3321081H, NG3409583H, NG3409611H, NG3409656H, NG3409680H, NG3409697H, NG3409716H, NG3409807H, NG3409825H, NG3409831H, NG3409854H, NG3409925H, NG3409959H, NG3410005H, NG3410053H, NG3410092H, NG3410111H, NG3410166H, NG3410189H, NG3410227H, NG3410233H, NG3410243H, NG3410279H, NG3410304H, NG3410389H, NG3410398H, NG3429103H, NG3429377H, NG3429485H, NG3429881H, NG3430107H, NG3430448H, NG3437945H, NG3437957H, NG3438394H, NG3470788H, NG3466185H, NG3466619H, NG3466620H, NG3467368H, NG3467586H, NG3468759H, NG3469430H, NG3469907H, NG3470728H, NG3168104H, NG3472157H, NG3169669H, NG3170079H, NG3430111H, NG3429099H, NG3437621H, NG3167969H, NG3465813H, NG3167397H Distribution: International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland. Affected quantity: 257
Hazard
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

3 products in this recall

Medtronic MiniMed insulin pumps recalled for cleaning damage risk

Medtronic is recalling three MiniMed insulin pump models (640G, 670G, and 780G) that were loaned out and may have been cleaned with products that are incompatible with the pump. This could damage the case or keypad, let moisture in, and cause the pump to deliver too much or too little insulin. No injuries have been reported. The recall covers 280 affected pumps and was distributed in several European countries.

The agency lists each product separately. Check every one you might have.