Serious riskFDA DeviceClass II

MiniMed 670G insulin pumps recalled for possible dosing errors

Seven loaner MiniMed 670G insulin pumps were cleaned with chemically incompatible products that may have damaged the pump casing or keypad. This damage could allow moisture inside the pump and cause it to deliver incorrect insulin doses — either too much or too little — which can lead to serious blood sugar swings.

From the FDA notice · Read the official notice ↗ · All Medtronic recalls

What to do

  1. If you are using a loaner MiniMed 670G pump (model MMT-1781 or MMT-1782), contact Medtronic immediately for a replacement or instructions.
  2. Check the serial number on your pump against the list provided by Medtronic to see if yours is affected.
  3. Do not return a loaner pump to Medtronic without first consulting them about the cleaning issue.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
Sold in
Nationwide
Made for
Adults (0 months and older)

This recall covers 3 products — see all of them.

Details

Units affected
7
Illnesses reported
None reported
Manufactured in
18000 Devonshire St, Northridge, CA, United States
Classification
Class II
Recall number
Z-2981-2026

In the FDA's words

Description
MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782; Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis. Model/UDI-DI Serial Numbers: MMT-1781/00763000404536/NG2665223H, NG2748200H, NG2748664H, NG2755237H, NG2716176H, NG3547582H. MMT-1782/00763000404543/NG2768190H Distribution: International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland. Affected quantity: 7
Hazard
Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

3 products in this recall

Medtronic MiniMed insulin pumps recalled for cleaning damage risk

Medtronic is recalling three MiniMed insulin pump models (640G, 670G, and 780G) that were loaned out and may have been cleaned with products that are incompatible with the pump. This could damage the case or keypad, let moisture in, and cause the pump to deliver too much or too little insulin. No injuries have been reported. The recall covers 280 affected pumps and was distributed in several European countries.

The agency lists each product separately. Check every one you might have.