Baxter TS7500 MOBIUS operating table column recalled for drop risk
Baxter is recalling TS7500 MOBIUS hybrid operating table columns because the main lift may unexpectedly drop to its lowest position during use. If this happens, the tabletop could not be raised again. An uncontrolled drop may result in abrupt patient displacement during surgery, and Baxter has received three complaints with no reported patient injuries.
From the FDA notice · Read the official notice ↗ · All Baxter recalls
What to do
- Set aside affected units and check them against the serial numbers listed in the notice.
- Identify affected columns by model 1704695, UDI-DI 00887761974425, and the listed serial numbers.
- Contact Baxter Healthcare Corporation. The notice does not specify a remedy.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 5 products — see all of them.
Details
- Units affected
- 77 units
- Illnesses reported
- None reported
- Manufactured in
- 1 Baxter Pkwy, Deerfield, Illinois, United States
- Classification
- Class II
- Recall number
- Z-0351-2027
In the FDA's words
- Description
- Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695 There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury. Serial numbers: 108094281, 108468075, 108484417, 108506318, 108516636, 108536785, 108544115, 108604154, 108463223, 108484404, 108489634, 108511411, 108516637, 108544106, 108544116, 108618964, 108463224, 108484405, 108489648, 108511412, 108521436, 108544107, 108604137, 108618965, 108468062, 108484406, 108506316, 108511413, 108521452, 108544110, 108604140, 108468063, 108484408, 108506317, 108516634, 108536776, 108544111, 108604153; UDI-DI number: 00887761974425 Distribution: Worldwide - US Nationwide distribution. Affected quantity: 77 units
- Hazard
- There is an issue with the main lift mechanism. The main lift may unexpectedly drop from its set position to its lowest position during use. In these cases, the tabletop could not be raised again after the drop occurred. The issue has been associated with unexpected wear of the main lift assembly, also known as the telescopic spindle. An uncontrolled drop of the surgical table may result in an abrupt patient displacement during surgery. Serious or critical adverse health consequences may occur, including traumatic injury, airway compromise, and surgical procedure complications, among others. Baxter has received three complaints related to this issue, with no reported patient injury.
5 products in this recall
Baxter TruSystem surgical tables recalled for unexpected drop risk
Baxter is recalling five surgical table and operating table column models: TruSystem 7500, TruSystem 7000 U, TS7500 MOBIUS, and TruSystem 7500 Hybrid Plus. The main lift may unexpectedly drop to its lowest position during use, and the tabletop could not be raised again after a drop. This has been linked to unexpected wear of the main lift assembly, also called the telescopic spindle. An uncontrolled drop may cause abrupt patient displacement, which could lead to traumatic injury, airway compromise, or surgical complications. Baxter has received three complaints and no patient injuries have been reported.
The agency lists each product separately. Check every one you might have.
- TruSystem 7500 surgical table, Model/Catalog Number: 4091000Z-0350-2027 · 129 units
- Operating table column TS7500 MOBIUS, Hybrid operating table column, Model/Catalog Number: 1704695 (this page)Z-0351-2027 · 77 units
- Operating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788Z-0352-2027 · 16 units
- Operating Table TruSystem 7000 U (dV), Model/Catalog Number: 1723633Z-0353-2027 · 62 units
- TruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087Z-0354-2027 · 1 unit
More from Baxter
See all- Baxter TruSystem 7000 U operating table recalled for possible unexpected drop
SeriousFDANov 3, 2026
- Baxter VOLARA System recalled for possible nebulizer flow rate error
ModerateFDASep 30, 2026
- Baxter Dextrose Injection 70% recalled for stainless steel particles
SeriousFDASep 16, 2026