Arrow guidewires recalled for possible coating detachment risk
Certain Arrow Marked hydrophilic-coated guidewires may have coatings that swell unevenly when hydrated with saline. The coating may clump and detach, and the particles may enter the bloodstream. The FDA states this could cause inflammation, blood clots, reduced blood flow to tissue, tissue injury, stroke, or the need for further intervention.
From the FDA notice · Read the official notice ↗ · All Arrow recalls
What to do
- Set aside affected lot 311-08884630 (expiration 10/20/2026) and do not use it.
- Check REF numbers PW-16130-HN, PW-18080-HN, and PW-18080-HS for lot 311-08884630.
- The notice doesn't specify a remedy; contact Arrow International, LLC for return or replacement instructions.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Arrow Marked Guidewire Hydrophilic-Coated Nitinol, REF: PW-16130-HN, PW-18080-HN; Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel, REF: PW-18080-HS
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
This recall covers 6 products — see all of them.
Details
- Units affected
- 79,678
- Illnesses reported
- None reported
- Manufactured in
- 3015 Carrington Mill Blvd, Morrisville, North Carolina, United States
- Classification
- Class I
- Recall number
- Z-0499-2027
In the FDA's words
- Description
- Arrow Marked Guidewire Hydrophilic-Coated Nitinol, REF: PW-16130-HN, PW-18080-HN; Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel, REF: PW-18080-HS Intravascular, peripheral, central catheters and guidewires have hydrophilic coated spring wire guide components with coating that may swell inconsistently when hydrated with saline leading to coating clumping and potential for detachment, which may lead to particle release into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, further intervention. Lots 241029P9, 250505P9, 250520P9 & 260406P9, any Expiration Date between 9/26 and 8/28. Distribution: Worldwide Distribution: (US nationwide) CA, SC, GA, AZ, CO, MO, LA, AR, TX, UT, MT, IA, MA, AL, WY, NJ, VA, PA, WV, DE, OH, NE, IN, FL, TN, NY, NH, MS, MD, MI, KS, NM, KY, OK, ME, NC, ID, WA, HI, NV, RI, OR, VT, IL, WI, SD, MN, DC, CT; (OUS countries of): Australia, Brunei, Canada, Chile, China, Colombia, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Philippines, Singapore, Thailand, Taiwan, and Vietnam. Affected quantity: 79,678
- Hazard
- Intravascular, peripheral, central catheters and guidewires have hydrophilic coated spring wire guide components with coating that may swell inconsistently when hydrated with saline leading to coating clumping and potential for detachment, which may lead to particle release into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, further intervention.
6 products in this recall
Arrow PICC catheters, guidewires recalled for possible particle release
Arrow is recalling six groups of PICC and JACC catheter kits and guidewires, including Arrowg+ard Blue Advance kits and Arrow Marked Guidewire products. The hydrophilic coating on the spring wire guide may swell unevenly when hydrated with saline, which may cause it to clump and possibly detach, releasing particles into the bloodstream. The notice reports no injuries.
The agency lists each product separately. Check every one you might have.
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet, REF: ASK-45541-CHS;
Pressure Injectable Arrowg+ar…Z-0498-2027 · 1,390
- Arrow Marked Guidewire Hydrophilic-Coated Nitinol, REF: PW-16130-HN, PW-18080-HN;
Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel, REF: PW-1808… (this page)Z-0499-2027 · 79,678
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45541-HPHNM;
Pressure Injectable Arrowg+…Z-0500-2027 · 90,426
- Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-34063-HPHNM, PR-35052-HPHNMZ-0501-2027 · 32,623
- Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet, REF: ASK-45052-MCZ-0502-2027 · 180
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: JRT-43541-HPHNM;
Pressure Injectable Arro…Z-0503-2027 · 446