Arrow International PICC catheters recalled for particle release risk
Arrow International is recalling Pressure Injectable Three-Lumen PICC with 80 cm Wire catheters, REF PR-34063-HPHNM and PR-35052-HPHNM. The hydrophilic coating on the wire guide may swell unevenly when wetted with saline, which may cause it to clump and detach. The notice says this may release particles into the bloodstream, which may lead to inflammation, blood clots, tissue injury, stroke, or further intervention.
From the FDA notice · Read the official notice ↗ · All Arrow recalls
What to do
- Clinicians: set aside unused affected catheters and do not use them.
- Check REF numbers PR-34063-HPHNM and PR-35052-HPHNM and expiration dates from 10/20/2026 through 03/15/2027. The notice lists no lot numbers, and its lot reference links to an unrelated press release.
- Contact Arrow International or your distributor for the remedy. The notice does not state one. For patients who already have these catheters, contact their care team rather than removing a device on your own.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-34063-HPHNM, PR-35052-HPHNM
- Sold in
- Nationwide
- Made for
- All ages (under 100 years)
This recall covers 6 products — see all of them.
Details
- Units affected
- 32,623
- Illnesses reported
- None reported
- Manufactured in
- 3015 Carrington Mill Blvd, Morrisville, North Carolina, United States
- Classification
- Class I
- Recall number
- Z-0501-2027
In the FDA's words
- Description
- Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-34063-HPHNM, PR-35052-HPHNM Intravascular, peripheral, central catheters and guidewires have hydrophilic coated spring wire guide components with coating that may swell inconsistently when hydrated with saline leading to coating clumping and potential for detachment, which may lead to particle release into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, further intervention. Lot: 920067, Exp Date: 10/25/2026; 948081, Exp Date: 11/22/2026 Distribution: Worldwide Distribution: (US nationwide) CA, SC, GA, AZ, CO, MO, LA, AR, TX, UT, MT, IA, MA, AL, WY, NJ, VA, PA, WV, DE, OH, NE, IN, FL, TN, NY, NH, MS, MD, MI, KS, NM, KY, OK, ME, NC, ID, WA, HI, NV, RI, OR, VT, IL, WI, SD, MN, DC, CT; (OUS countries of): Australia, Brunei, Canada, Chile, China, Colombia, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Philippines, Singapore, Thailand, Taiwan, and Vietnam. Affected quantity: 32,623
- Hazard
- Intravascular, peripheral, central catheters and guidewires have hydrophilic coated spring wire guide components with coating that may swell inconsistently when hydrated with saline leading to coating clumping and potential for detachment, which may lead to particle release into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, further intervention.
6 products in this recall
Arrow PICC catheters, guidewires recalled for possible particle release
Arrow is recalling six groups of PICC and JACC catheter kits and guidewires, including Arrowg+ard Blue Advance kits and Arrow Marked Guidewire products. The hydrophilic coating on the spring wire guide may swell unevenly when hydrated with saline, which may cause it to clump and possibly detach, releasing particles into the bloodstream. The notice reports no injuries.
The agency lists each product separately. Check every one you might have.
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet, REF: ASK-45541-CHS;
Pressure Injectable Arrowg+ar…Z-0498-2027 · 1,390
- Arrow Marked Guidewire Hydrophilic-Coated Nitinol, REF: PW-16130-HN, PW-18080-HN;
Arrow Marked Guidewire Hydrophilic-Coated Stainless Steel, REF: PW-1808…Z-0499-2027 · 79,678
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45541-HPHNM;
Pressure Injectable Arrowg+…Z-0500-2027 · 90,426
- Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-34063-HPHNM, PR-35052-HPHNM (this page)Z-0501-2027 · 32,623
- Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC Kit pre-loaded with Arrow VPS Precision Stylet, REF: ASK-45052-MCZ-0502-2027 · 180
- Pressure Injectable Arrowg+ard Blue Advance One-Lumen JACC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: JRT-43541-HPHNM;
Pressure Injectable Arro…Z-0503-2027 · 446