Serious riskFDA DeviceClass I

Arrowg+ard Blue PICC catheters recalled for guidewire detachment risk

Arrow International recalled certain Arrowg+ard PICC catheters with hydrophilic-coated guidewires. The coating may swell unevenly when wetted with saline, which could cause clumping and potential detachment, and fragments could enter the bloodstream and may result in inflammation, clots, stroke, or further intervention. The notice reports no injuries.

From the FDA notice · Read the official notice ↗ · All Arrow recalls

What to do

  1. Set aside affected lots 000-08733129, 000-08792116, and 000-08880757 and do not use them.
  2. Check the lot numbers, expiration dates, and REF numbers on your stock against the full product list in the notice.
  3. Contact Arrow International, LLC for next steps, since the notice does not specify a remedy.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45541-HPHNM; Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45563-HPHNL; Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45552-HPHNL; Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45552-HPHNM, PR-45563-HPHNM; Pressure Injectable One-Lumen PICC, REF: PR-35541-NL; Pressure Injectable One-Lumen PICC with 130 cm Wire, REF: PR-34041-NL, PR-35541-HPHNL; Pressure Injectable One-Lumen PICC with 80 cm Wire, REF: PR-34041-HPHNM, PR-34041-NM, PR-35041-HPHNM, PR-35541-HPHNM; Pressure Injectable Three-Lumen PICC, REF: PR-34063-NM, PR-35563-NL; Pressure Injectable Three-Lumen PICC with 130 cm Wire, REF: PR-34063-NL, PR-35563-HPHNL; Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-35563-HPHNM; Pressure Injectable Two-Lumen PICC, REF: PR-34052-NM, PR-35552-NL; Pressure Injectable Two-Lumen PICC with 130 cm Wire, REF: PR-34052-NL, PR-35552-HPHNL; Pressure Injectable Two-Lumen PICC with 80 cm Wire, REF: PR-34052-HPHNM, PR-35552-HPHNM
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 6 products — see all of them.

Details

Units affected
90,426
Illnesses reported
None reported
Manufactured in
3015 Carrington Mill Blvd, Morrisville, North Carolina, United States
Classification
Class I
Recall number
Z-0500-2027

In the FDA's words

Description
Pressure Injectable Arrowg+ard Blue Advance One-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45541-HPHNM; Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45563-HPHNL; Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC with 130 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45552-HPHNL; Pressure Injectable Arrowg+ard Blue Advance Two-Lumen PICC with 80 cm Hydrophilic-Coated Nitinol Guidewire, REF: PR-45552-HPHNM, PR-45563-HPHNM; Pressure Injectable One-Lumen PICC, REF: PR-35541-NL; Pressure Injectable One-Lumen PICC with 130 cm Wire, REF: PR-34041-NL, PR-35541-HPHNL; Pressure Injectable One-Lumen PICC with 80 cm Wire, REF: PR-34041-HPHNM, PR-34041-NM, PR-35041-HPHNM, PR-35541-HPHNM; Pressure Injectable Three-Lumen PICC, REF: PR-34063-NM, PR-35563-NL; Pressure Injectable Three-Lumen PICC with 130 cm Wire, REF: PR-34063-NL, PR-35563-HPHNL; Pressure Injectable Three-Lumen PICC with 80 cm Wire, REF: PR-35563-HPHNM; Pressure Injectable Two-Lumen PICC, REF: PR-34052-NM, PR-35552-NL; Pressure Injectable Two-Lumen PICC with 130 cm Wire, REF: PR-34052-NL, PR-35552-HPHNL; Pressure Injectable Two-Lumen PICC with 80 cm Wire, REF: PR-34052-HPHNM, PR-35552-HPHNM Intravascular, peripheral, central catheters and guidewires have hydrophilic coated spring wire guide components with coating that may swell inconsistently when hydrated with saline leading to coating clumping and potential for detachment, which may lead to particle release into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, further intervention. Lot: 935670, Exp Date: 12/27/2026 Distribution: Worldwide Distribution: (US nationwide) CA, SC, GA, AZ, CO, MO, LA, AR, TX, UT, MT, IA, MA, AL, WY, NJ, VA, PA, WV, DE, OH, NE, IN, FL, TN, NY, NH, MS, MD, MI, KS, NM, KY, OK, ME, NC, ID, WA, HI, NV, RI, OR, VT, IL, WI, SD, MN, DC, CT; (OUS countries of): Australia, Brunei, Canada, Chile, China, Colombia, India, Japan, South Korea, Mexico, Malaysia, Nepal, New Zealand, Philippines, Singapore, Thailand, Taiwan, and Vietnam. Affected quantity: 90,426
Hazard
Intravascular, peripheral, central catheters and guidewires have hydrophilic coated spring wire guide components with coating that may swell inconsistently when hydrated with saline leading to coating clumping and potential for detachment, which may lead to particle release into the blood stream, which may result in inflammation, thrombosis, ischemia, tissue injury, stroke, further intervention.

6 products in this recall

Arrow PICC catheters, guidewires recalled for possible particle release

Arrow is recalling six groups of PICC and JACC catheter kits and guidewires, including Arrowg+ard Blue Advance kits and Arrow Marked Guidewire products. The hydrophilic coating on the spring wire guide may swell unevenly when hydrated with saline, which may cause it to clump and possibly detach, releasing particles into the bloodstream. The notice reports no injuries.