Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup s
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
From the FDA notice · Read the official notice ↗ · All Thoratec recalls
What to do
The agency's notice has the remedy and who to contact.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 LVAS
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
This recall covers 4 products — see all of them.
Details
- Units affected
- 35376
- Illnesses reported
- None reported
- Manufactured in
- 6035 Stoneridge Dr, Pleasanton, CA, United States
- Classification
- Class II
- Recall number
- Z-0506-2026
In the FDA's words
- Description
- Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3 Controller, OUS; 106531LF2, HeartMate 3 Low Flow 2.0 Controller Software Version: N/A Product Description: The HeartMate 3" System Controller acts as the central power and communication hub for the HeartMate 3 LVAS. It passes power from the Power Module, the Mobile Power Unit, Lithium-Ion Batteries, or its own integrated emergency backup supply, down to the LVAD via the Driveline. The HeartMate 3 System Controller constantly monitors system performance through communication with the implanted LVAD and Controller internal measurements and alerts the user to any alarm conditions by activating membrane panel LEDs and integrated audio annunciators. Further information on alarm conditions as well as system status can be attained by the user from the front panel LCD on the System Controller. When connected to a HeartMate Touch" Communication System, the System Controller sends information regarding the System Controller and Pump Status once per second to provide additional information to the user. This link also allows the clinician to set new patient operating parameters (e.g. pump speed) and provides a link for downloading trend and/or event recorder data. Component: Yes, HeartMate 3 LVAS Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm. Lot Code: All units of the following models are impacted: Model 106531US UDI: 05415067039446 Model 106531INT UDI: 05415067038982 Model 106531LF2 UDI: 05415067039422 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used. Distribution: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan. Affected quantity: 35376
- Hazard
- Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.
4 products in this recall
HeartMate LVAD controllers recalled for backup battery corrosion
Thoratec, a HeartMate company, is recalling four HeartMate heart pump system products because the backup battery cable connector may corrode. This can trigger a Backup Battery Fault alarm. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- Brand Name: HeartMate Product Name: HeartMate 3 Left Ventricular System (LVAS) Kits Model/Catalog Number: 106524US, HeartMate 3 LVAS KIT, US; 106524INT, He…Z-0505-2026 · 29059
- Brand Name: HeartMate Product Name: HeartMate 3 Controller (Standalone) Model/Catalog Number: 106531US, HeartMate 3 Controller, US; 106531INT, HeartMate 3… (this page)Z-0506-2026 · 35376
- Brand Name: HeartMate Product Name: HeartMate II¿ LVAS Implant Kit Model/Catalog Number: 106015, HeartMate II¿ LVAS Kit, with Pocket Controller, Sealed, US…Z-0507-2026 · 248
- Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartM…Z-0508-2026 · 2424