FDA DeviceClass II

Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, HeartMate II¿ Pocket Controller Software Version: N/A Product Description: The HeartMate II System Controller is a small computer that controls and monitors system operation. It sends power and operating signals to the Left Ventricular Assist Device and adjusts device operation to maintain programme

Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.

From the FDA notice · Read the official notice ↗ · All Thoratec recalls

What to do

The agency's notice has the remedy and who to contact.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, HeartMate II¿ Pocket Controller Software Version: N/A Product Description: The HeartMate II System Controller is a small computer that controls and monitors system operation. It sends power and operating signals to the Left Ventricular Assist Device and adjusts device operation to maintain programmed levels of cardiac support. The System Controller identifies alarm conditions and initiates Hazard and Advisory alarms. It records device performance and alarm data, and transfers the information to the HeartMate Touch Communication System. Component: Yes, HeartMate II¿ LVAS Implant Kit
Sold in
Nationwide
Made for
Adults (18 years to 100 years)

This recall covers 4 products — see all of them.

Details

Units affected
2424
Illnesses reported
None reported
Manufactured in
6035 Stoneridge Dr, Pleasanton, CA, United States
Classification
Class II
Recall number
Z-0508-2026

In the FDA's words

Description
Brand Name: HeartMate Product Name: HeartMate II¿ Controller (standalone) Model/Catalog Number: 106017, HeartMate II¿ Pocket Controller, US; 106762, HeartMate II¿ Pocket Controller, US; 107801, HeartMate II¿ Pocket Controller Software Version: N/A Product Description: The HeartMate II System Controller is a small computer that controls and monitors system operation. It sends power and operating signals to the Left Ventricular Assist Device and adjusts device operation to maintain programmed levels of cardiac support. The System Controller identifies alarm conditions and initiates Hazard and Advisory alarms. It records device performance and alarm data, and transfers the information to the HeartMate Touch Communication System. Component: Yes, HeartMate II¿ LVAS Implant Kit Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm. Lot Code: All units of the following models are impacted: Model 106017 UDI: 00813024011248 Model 106762 UDI: 00813024011286 Model 107801 UDI: 00813024011996 Note: Product bracketing is determined based on possible use of the controller. Since the backup battery has a 3 year life cycle, it can be replaced (per the IFU) and the controller will continue to be used. Distribution: Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV and the countries of Algeria, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, India, Iran, Iraq, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Lithuania, Macedonia, Malaysia, Martinique, Montenegro, Morocco, Netherlands, New Zealand, Norway, Oman, Palestine, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, UK, United Arab Emirates, Uzbekistan. Affected quantity: 2424
Hazard
Due to potential corrosion issues associated with the Backup Battery cable connector that may cause a Backup Battery Fault Alarm.

4 products in this recall

HeartMate LVAD controllers recalled for backup battery corrosion

Thoratec, a HeartMate company, is recalling four HeartMate heart pump system products because the backup battery cable connector may corrode. This can trigger a Backup Battery Fault alarm. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

Related topics