Medline homecare beds recalled for potential fire risk from pendant
Medline Industries has recalled certain Full Electric Low Lightweight Homecare Beds (model MDR107003LO) because the hand control pendant and its wires can overheat and catch fire in rare situations—such as when the bed is overloaded, blocked, or has jammed parts. The company has received 58 reports of the pendant sparking, burning, melting, smoking, or catching fire.
From the FDA notice · Read the official notice ↗ · All Medline recalls
What to do
- Stop using the hand control pendant immediately if you own this bed model
- Contact Medline Industries for instructions on repair or replacement
- Check your bed's reference number (REF MDR107003LO) and UDI (40080196325901) to confirm it is affected
- Watch for signs of sparking, burning, melting, or smoking near the pendant or wires, and do not use the bed if you notice these signs
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO
- Sold in
- Nationwide
- Made for
- Adults (0 months and older)
This recall covers 6 products — see all of them.
Details
- Units affected
- 15581 units
- Illnesses reported
- Yes
- Manufactured in
- 3 Lakes Dr, Northfield, IL, United States
- Classification
- Class I
- Recall number
- Z-1016-2026
In the FDA's words
- Description
- MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire. UDI/DI 40080196325901, All lots Distribution: US, CA, PA, GUAM, VI Affected quantity: 15581 units
- Hazard
- Medline Industries has identified a potential hazard involving the hand control pendant and associated wires. In rare and atypical scenarios such as operating the bed above its specified weight limit, encountering an obstruction to bed movement, or experiencing jammed components/motor the hand pendant and pendant cords may overheat and, in rare instances, pose a risk of fire. Medline Industries, LP has received a total of (58) complaints associated with reports of pendant sparking, burning, melting, smoking or fire.
6 products in this recall
Medline homecare beds recalled for overheating and fire risk
Medline is recalling six models of electric homecare beds, about 599,000 units in total. In rare cases, such as overloading the bed, an obstruction blocking its movement, or a jammed motor, the hand control pendant and its cords can overheat and, in rare instances, catch fire. Medline has received 58 complaints of pendants sparking, burning, melting, smoking, or catching fire. No injuries have been reported.
The agency lists each product separately. Check every one you might have.
- MEDLINE beds labeled as: 1) Semi Electric Basic Homecare Bed, REF MDR107002E; 2) Semi Electric Basic Homecare Bed, 4-pack,REF MDR107002E-4Z-1011-2026 · 373392 units
- MEDLINE beds labeled as: Semi Electric Lightweight Homecare Bed, REF MDR107002LZ-1012-2026 · 13245 units
- MEDLINE beds labeled as: 1) Full Electric Basic Homecare Bed, REF MDR107003E; 2) Full Electric Basic Homecare Bed, 4-pack,REF MDR107003E-4.Z-1013-2026 · 138412 units
- MEDLINE beds labeled as: Full Electric Low Basic Homecare Bed, REF MDR107003ELOZ-1014-2026 · 22565 units
- MEDLINE beds labeled as: Full Electric Lightweight Homecare Bed, REF MDR107003LZ-1015-2026 · 35694 units
- MEDLINE beds labeled as: Full Electric Low Lightweight Homecare Bed, REF MDR107003LO (this page)Z-1016-2026 · 15581 units
More from Medline
See all- Medline CABG Auxiliary Pack surgical kits recalled for cannula component defects
ModerateFDAOct 21, 2026
- Medline foot-ankle convenience kit recalled for shelf-life component discrepancy
ModerateFDAOct 21, 2026
- Medline bariatric rollator recalled for front wheel detachment risk
SeriousFDAOct 14, 2026