Moderate riskFDA DeviceClass II

Medline Basic Nerve Block Tray recalled for potential sterility compromise

Medline Industries recalled its Basic Nerve Block Tray with Linen (Model PAIN0150C) because equipment used to sterilize and package the devices had calibration problems. While the products went through sterilization cycles, the calibration issues may have reduced the sterility assurance of these medical trays.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Stop using affected trays immediately if you have Model PAIN0150C, Lot 21JBV886
  2. Contact your Medline representative or the FDA with questions about your specific lot
  3. Do not use these trays for patient care until you confirm your lot is not affected

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number: PAIN0150C
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 42 products — see all of them.

Details

Units affected
448 units
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Classification
Class II
Recall number
Z-1387-2026

In the FDA's words

Description
Medline Convenience Kits: 1) BASIC NERVE BLOCK TRAY W/LINEN, Model Number: PAIN0150C Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. 1) PAIN0150C, UDI-DI: 10193489819571(each), 21JBV886(case), Lot Number: 21JBV886 Distribution: Worldwide distribution - US Nationwide. Affected quantity: 448 units
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

42 products in this recall

Medline convenience kits recalled for sterility assurance concerns

Medline is recalling 42 convenience kits and procedure packs that were sterilized and packaged on equipment with calibration problems. The kits went through the validated sterilization and packaging cycles, but the calibration issues could affect how reliably they are sterile. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Medline

See all