Moderate riskFDA DeviceClass II

Medline LVAD driveline and central line trays recalled over sterilization calibration issue

Medline identified calibration problems with sterilization equipment used to process certain LVAD driveline and central line dressing change trays. These issues may have affected the sterility of the devices, even though they went through standard sterilization cycles. The recall affects specific lot numbers manufactured between 2020 and 2025.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Check your inventory for the affected lot numbers listed in the recall (DM1035A, DM1035, or DT8670A)
  2. Stop using any trays matching the affected lot numbers
  3. Contact Medline or your supplier for instructions on return or replacement
  4. Notify your facility's infection control and clinical teams immediately

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 42 products — see all of them.

Details

Units affected
9720 units
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Classification
Class II
Recall number
Z-1411-2026

In the FDA's words

Description
Medline Convenience Kits: 1) LVAD DRIVELINE TRAY, Model Number: DM1035A; 2) LVAD DRIVELINE TRAY, Model Number: DM1035; 3) CENTRAL LINE DRESSING CHANGE TRAY, Model Number: DT8670A Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. 1) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2025051290; 2) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2025031190; 3) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2023062390; 4) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2023101290; 5) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2022030390; 6) DM1035A, UDI-DI: 10653160345251(each), 00653160345254(case), Lot Number: 2022072590; 7) DM1035, UDI-DI: 10653160327752(each), 00653160327755(case), Lot Number: 2020042290; 8) DM1035, UDI-DI: 10653160327752(each), 00653160327755(case), Lot Number: 2020052290; 9) DM1035, UDI-DI: 10653160327752(each), 00653160327755(case), Lot Number: 2020072290; 10) DM1035, UDI-DI: 10653160327752(each), 00653160327755(case), Lot Number: 2020082690; 11) DT8670A, UDI-DI: 00653160177473(each), 10653160177470(case), Lot Number: 2020051390 Distribution: Worldwide distribution - US Nationwide. Affected quantity: 9720 units
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

42 products in this recall

Medline convenience kits recalled for sterility assurance concerns

Medline is recalling 42 convenience kits and procedure packs that were sterilized and packaged on equipment with calibration problems. The kits went through the validated sterilization and packaging cycles, but the calibration issues could affect how reliably they are sterile. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Medline

See all

Related topics

SterilityMedical devicesSold nationwideMore from FDA Device