FDA-Device2026-02-25Class IIPROCESSING DEFECT

Medline Convenience Kits recalled due to sterilization equipment calibration issues

OTHERNationwide distribution

Stop using Medline Convenience Kits

Medline identified calibration problems with equipment used to sterilize and package two models of convenience kits: the Drape Pack (Model DYNJ35363A) and Sick Kids-General Custom Split (Model SPTPCD10005). While the products went through sterilization cycles, the calibration issues may have affected whether the products remained properly sterile.

  • Check if you have Lot Number 21BBP020 (Drape Pack) or Lot Number 24JBB041 (Sick Kids-General).
  • Stop using affected kits and contact Medline Industries for instructions on return or replacement.
  • If these kits were already used in medical procedures, consult with your healthcare provider.
Hazard

Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

Sold states
Worldwide distribution - US Nationwide.
Affected count
240 units
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Products
Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-1409-2026