Moderate riskFDA DeviceClass II

Medline Convenience Kits recalled due to sterilization equipment calibration issues

Medline identified calibration problems with equipment used to sterilize and package two models of convenience kits: the Drape Pack (Model DYNJ35363A) and Sick Kids-General Custom Split (Model SPTPCD10005). While the products went through sterilization cycles, the calibration issues may have affected whether the products remained properly sterile.

From the FDA notice · Read the official notice ↗ · All Medline recalls

What to do

  1. Check if you have Lot Number 21BBP020 (Drape Pack) or Lot Number 24JBB041 (Sick Kids-General).
  2. Stop using affected kits and contact Medline Industries for instructions on return or replacement.
  3. If these kits were already used in medical procedures, consult with your healthcare provider.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005
Sold in
Nationwide
Made for
Children (under 18 years)

This recall covers 42 products — see all of them.

Details

Units affected
240 units
Illnesses reported
None reported
Manufactured in
3 Lakes Dr, Northfield, IL, United States
Classification
Class II
Recall number
Z-1409-2026

In the FDA's words

Description
Medline Convenience Kits: 1) DRAPE PACK, Model Number: DYNJ35363A; 2) SICK KIDS-GENERAL CUSTOM SPLIT, Model Number: SPTPCD10005 Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products. 1) DYNJ35363A, UDI-DI: 10884389516461(each), 40884389516462(case), Lot Number: 21BBP020; 2) SPTPCD10005, UDI-DI: 10080196228652(each), 40080196228653(case), Lot Number: 24JBB041 Distribution: Worldwide distribution - US Nationwide. Affected quantity: 240 units
Hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.

42 products in this recall

Medline convenience kits recalled for sterility assurance concerns

Medline is recalling 42 convenience kits and procedure packs that were sterilized and packaged on equipment with calibration problems. The kits went through the validated sterilization and packaging cycles, but the calibration issues could affect how reliably they are sterile. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Medline

See all

Related topics