FDA-Device2026-03-18Class IIPROCESSING DEFECT

Waldemar Link Endo-Model SL implant screw recalled for potential bushing detachment

OTHERNationwide distribution

Check Waldemar Link implant screw lot numbers

Waldemar Link GmbH & Co. KG is recalling specific lots of Endo-Model SL Connection Component implant screws. The bushing on the plateau screw could detach from the screw shaft due to a crack, which could affect implant stability.

  • If you received this implant, contact your surgeon or healthcare provider with your implant lot number
  • Review the list of affected lot numbers (starting 250925 or 251110) with your medical records
  • Do not attempt to remove or adjust the implant yourself
Hazard

The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

Sold states
Worldwide - US Nationwide distribution in the state of NJ and the countries of Belarus, Canada, Switzerland , Colombia, Germany, Spain, United Kingdom, Hungary, India, Italy, Netherlands, Romania, Saudi Arabia.
Affected count
49 units (OUS only)
Manufactured in
Ost Str. 4-10, N/A, Norderstedt, N/A, Germany
Products
Endo-Model SL Connection Component incl. PE-Plateau; Item Number: 16-2840/02;