Moderate riskFDA DeviceClass II

Waldemar Link Endo-Model Replacement Plateau screw recalled for potential bushing detachment

Waldemar Link GmbH & Co. KG is recalling Endo-Model Replacement Plateau screws (Item Number 15-0027/12) because the bushing could detach from the screw shaft due to a crack. This could affect the stability of the implant. One affected unit has been identified outside the US.

From the FDA notice · Read the official notice ↗ · All Waldemar Link recalls

What to do

  1. If you received this implant, contact your surgeon or healthcare provider immediately
  2. Do not use the affected screw (Lot 2549036, UDI-DI: 04026575431960)
  3. Ask your provider whether replacement is needed

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Endo-Model Replacement Plateau; Item Number: 15-0027/12;
Sold in
Nationwide
Made for
Adults (0 months and older)

This recall covers 13 products — see all of them.

Details

Units affected
1 unit (OUS only)
Illnesses reported
None reported
Manufactured in
Ost Str. 4-10, N/A, Norderstedt, N/A, Germany
Classification
Class II
Recall number
Z-1522-2026

In the FDA's words

Description
Endo-Model Replacement Plateau; Item Number: 15-0027/12; The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture. Item Number: 15-0027/12; UDI-DI: 04026575431960; Serial/Lot Number: 2549036; Distribution: Worldwide - US Nationwide distribution in the state of NJ and the countries of Belarus, Canada, Switzerland , Colombia, Germany, Spain, United Kingdom, Hungary, India, Italy, Netherlands, Romania, Saudi Arabia. Affected quantity: 1 unit (OUS only)
Hazard
The potential for an implant with a plateau screw where the bushing could detach from the screw shaft due to a longitudinal fracture.

13 products in this recall

Waldemar Link Endo-Model implant parts recalled for fracture risk

Waldemar Link is recalling 13 Endo-Model implant components, including replacement plateaus and connection components with PE-Plateau. The bushing on these implants could detach from the screw shaft because of a longitudinal fracture. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

More from Waldemar Link

See all

Related topics