I.T.S. orthopedic surgical plates recalled over higher MRI heating risk

I.T.S. GmbH orthopedic plates used in bone surgery were tested for safety during MRI scans. Testing found these plates can get hotter than originally stated when exposed to certain MRI conditions. The manufacturer has updated safety information to reflect the actual heating levels.

From the FDA notice · Read the official notice ↗

What to do

  1. If you have received one of these plates, contact your surgeon or hospital with your plate article and lot number.
  2. Before any MRI scan, inform the radiologist and MRI technician that you have an I.T.S. orthopedic plate implanted.
  3. Review the updated Instructions for Use provided by the manufacturer for new MRI safety guidance.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9.
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 17 products — see all of them.

Details

Units affected
510 units
Illnesses reported
None reported
Manufactured in
Autal 28, Lasnitzhohe, Austria
Classification
Class II
Recall number
Z-1598-2026

In the FDA's words

Description
I.T.S. Straight Plate with Angular Stability with the below descriptions and corresponding article numbers. 1. Forearm Plate, Curved; Article Numbers: 21604-6, 21604-9. 2. Forearm Plate, Straight; Article Numbers: 21605-11, 21605-6, 21605-9. 3. Straight Compression Plate; Article Numbers: 21105-6, 21105-8. 4. Straight Plate, 1.5mm; Article Numbers: 21104-4, 21104-5, 21104-6, 21104-7, 21104-8, 21104-9. 5. Straight Plate, 2.0mm; Article Numbers: 21101-5, 21101-6, 21101-7, 21101-8, 21101-9, 21101-11, 21101-13. 6. Straight Plate, 3.5mm; Article Number: 21102-6. 7. Straight Plate, 4.5mm; Article Numbers: 21103-8, 21103-11, 21103-10, 21103-11, 21103-12, 21103-13, 21103-8, 21103-9. Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). 1. Article Number (Lot Numbers): 21604-6 (643/052301, 643/0652, 643/092201), 21604-9 (643/032311, 643/062308, 643/062309, 643/112310, 643/122201). 2. Article Number (Lot Numbers): 21605-11 (643/072314, 643/102207), 21605-6 (643/0654, 643/072306, 643/072316, 643/072338, 643/102208, 643/102212, 643/102217, 643/112306, 643/122202), 21605-9 (643/0643, 643/0662, 643/072307, 643/072339, 643/112309, 643/122203). 3. Article Number (Lot Numbers): 21105-6 (86/876787), 21105-8 (86/876788). 4. Article Number (Lot Numbers): 21104-4 (679/052234, 679/082305, 86/036334), 21104-5 (679/0722101, 679/082306, 86/036335), 21104-6 (679/012307, 86/876760), 21104-7 (679/082307, 86/036336, 86/876781), 21104-8 (679/082308, 86/876783), 21104-9 (86/876904). 5. Article Number (Lot Numbers): 21101-5 (86/876637), 21101-6 (679/121605, 00/1105-12, 26/F41923, 86/876638, 86/903610), 21101-7 (770/201905, 00/0518-06, 679/042303, 86/876640), 21101-8 (86/876721, 679/062224), 21101-9 (679/122264, 86/876727), 21101-11 (86/876633), 21101-13 (86/876635). 6. Article Number (Lot Numbers): 21102-6 (770/201905). 7. Article Number (Lot Numbers): 21103-8 (679/092216), 21103-9 (679/062215, 86/876758), 21103-10 (679/092219, 86/876741), 21103-11 (679/062214), 21103-12 (86/876743), 21103-13 (679/022303, 86/876745). Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W. Affected quantity: 510 units
Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

17 products in this recall

I.T.S. bone plates and screws recalled for MRI heating risk

I.T.S. is recalling 17 orthopedic implant product lines, including bone plates and screw systems. Updated MRI safety testing found that these implants can heat up more during certain MRI conditions than the product instructions previously stated. No injuries have been reported.

The agency lists each product separately. Check every one you might have.