Moderate riskFDA DeviceClass II

I.T.S. GmbH headless compression screws recalled for increased MRI heating risk

I.T.S. GmbH headless compression screws used in orthopedic surgery may heat up more than expected during MRI scans under certain conditions. Updated safety testing found higher temperature increases than the original instructions stated. This affects specific lot numbers of these implants distributed across the United States.

From the FDA notice · Read the official notice ↗

What to do

  1. If you have one of these screws implanted, contact your surgeon or orthopedic clinic with your implant lot number
  2. Ask your doctor about MRI safety precautions before scheduling an MRI scan
  3. Do not delay necessary medical imaging—discuss options with your healthcare provider

Is yours affected?

Compare your label or packaging with the recalled product.

Product
HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90.
Sold in
Nationwide
Made for
All ages (under 120 years)

This recall covers 17 products — see all of them.

Details

Units affected
623 units
Illnesses reported
None reported
Manufactured in
Autal 28, Lasnitzhohe, Austria
Classification
Class II
Recall number
Z-1606-2026

In the FDA's words

Description
HCS - Headless Compression Screw System, Headless Compression Screw, Cannulated; Article Numbers: 31207-10, 31207-12, 31207-14, 31207-16, 31207-18, 31207-20, 31207-24, 31207-26, 31207-28, 31207-30, 31257-12, 31257-14, 31257-16, 31257-18, 31257-20, 31257-22, 31257-24, 31257-28, 31257-30, 31307-18, 31307-32, 31307-34, 31307-36, 31307-38, 31358-30, 31437-30, 31437-38, 31437-40, 31437-50, 31757-100, 31757-120, 31757-45, 31757-80, 31757-85, 31757-90. Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). Article Number (Lot Numbers): 31207-10 (679/112002), 31207-12 (679/092133, 679/102343, 679/112006), 31207-14 (679/102050), 31207-16 (679/102345, 679/102049), 31207-18 (679/0323110, 679/102355, 679/112001), 31207-20 (679/0323111, 679/102051, 679/102356), 31207-24 (679/102028), 31207-26 (679/112008), 31207-28 (679/112009), 31207-30 (679/0323140, 86/483377), 31257-12 (679/081904), 31257-14 (679/042323, 679/081905), 31257-16 (679/042324), 31257-18 (679/112308, 679/022128), 31257-20 (679/022123, 679/042326, 679/112321), 31257-22 (679/022124, 679/042327, 679/112325), 31257-24 (679/022125, 679/112326), 31257-28 (679/022130, 679/052306), 31257-30 (679/022131), 31307-18 (679/061907), 31307-32 (679/062285), 31307-34 (679/062286), 31307-36 (679/062287), 31307-38 (679/032118), 31358-30 (26/F67790), 31437-30 (679/071916), 31437-38 (679/071918), 31437-40 (679/032406, 679/071903), 31437-50 (679/071920), 31757-100 (35/51892), 31757-120 (86/320792), 31757-45 (86/320793), 31757-80 (26/F55872), 31757-85 (26/F55873), 31757-90 (26/F55874). Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W. Affected quantity: 623 units
Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

17 products in this recall

I.T.S. bone plates and screws recalled for MRI heating risk

I.T.S. is recalling 17 orthopedic implant product lines, including bone plates and screw systems. Updated MRI safety testing found that these implants can heat up more during certain MRI conditions than the product instructions previously stated. No injuries have been reported.

The agency lists each product separately. Check every one you might have.