Moderate riskFDA DeviceClass II

I.T.S. GmbH HLS Hand Locking Plates System recalled for higher MRI temperature risk

I.T.S. GmbH has recalled certain cortical screws, locking stabilization screws, and HLS plates used in hand surgery. New MRI safety testing found these implants may heat up more than expected during MRI scans under certain conditions. The original instructions did not reflect this updated safety information.

From the FDA notice · Read the official notice ↗

What to do

  1. If you received one of these implants, contact your surgeon to confirm your specific article and lot numbers.
  2. Inform your healthcare providers about this recall before any future MRI procedures.
  3. Ask your surgeon whether your implant requires special precautions or alternative imaging methods.
  4. Do not have an MRI without first discussing this recall with your surgical team.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-15, 32151-16, 32151-18, 32151-20. 2. HLS Plate, 2 T-Shape; Article Numbers: 21505-4, 21506-4. 3. HLS Plate, 3 T-Shape; Article Numbers: 21507-4, 21508-4. 4. HLS Plate, Square; Article Numbers: 21517. 5. HLS Plate, Straight; Article Numbers: 21502, 21503-4, 21503-6, 21503-8, 21503-10, 21504-6, 21504-8, 21504-10. 6. HLS Plate, Y-Shape; Article Numbers: 21513-3, 21513-5, 21514-3, 21514-5. 7. Locking Stabilization Screw; Article Numbers: 37151-5, 37151-8, 37151-9, 37151-10, 37151-12, 37151-13, 37151-15, 37151-16, 37151-18, 37151-24, 37251-8, 37251-9, 37251-11, 37251-12, 37251-13, 37251-14, 37251-16, 37251-20.
Sold in
Nationwide
Made for
All ages (18 years to 100 years)

This recall covers 17 products — see all of them.

Details

Units affected
1,307 units
Illnesses reported
None reported
Manufactured in
Autal 28, Lasnitzhohe, Austria
Classification
Class II
Recall number
Z-1607-2026

In the FDA's words

Description
HLS - Hand Locking Plates System with the below descriptions and corresponding article numbers. 1. Cortical Screw; Article Numbers: 32151-6, 32151-8, 32151-9, 32151-10, 32151-11, 32151-12, 32151-13, 32151-14, 32151-15, 32151-16, 32151-18, 32151-20. 2. HLS Plate, 2 T-Shape; Article Numbers: 21505-4, 21506-4. 3. HLS Plate, 3 T-Shape; Article Numbers: 21507-4, 21508-4. 4. HLS Plate, Square; Article Numbers: 21517. 5. HLS Plate, Straight; Article Numbers: 21502, 21503-4, 21503-6, 21503-8, 21503-10, 21504-6, 21504-8, 21504-10. 6. HLS Plate, Y-Shape; Article Numbers: 21513-3, 21513-5, 21514-3, 21514-5. 7. Locking Stabilization Screw; Article Numbers: 37151-5, 37151-8, 37151-9, 37151-10, 37151-12, 37151-13, 37151-15, 37151-16, 37151-18, 37151-24, 37251-8, 37251-9, 37251-11, 37251-12, 37251-13, 37251-14, 37251-16, 37251-20. Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). 1. Article Number (Lot Numbers): 32151-6 (18/408462, 18/500495, 18/691266), 32151-8 (18/408866, 18/531370), 32151-9 (18/351668, 18/541373, 18/741269), 32151-10 (18/408986, 18/551374), 32151-11 (18/371670, 18/561375), 32151-12 (18/409087, 18/571376, 18/771272), 32151-13 (18/581377), 32151-14 (18/409289, 18/591378, 18/791274), 32151-15 (18/409390), 32151-16 (18/409491), 32151-18 (18/821277), 32151-20 (18/831278). 2. Article Number (Lot Numbers): 21505-4 (86/350493, 86/677353), 21506-4 (86/642718, 86/677157). 3. Article Number (Lot Numbers): 21507-4 (26/F65050), 21508-4 (26/F63934, 86/677180). 4. Article Number (Lot Numbers): 21517 (26/F66911). 5. Article Number (Lot Numbers): 21502 (86/677136), 21503-4 (86/337934), 21503-6 (86/868870), 21503-8 (26/F63936), 21503-10 (86/337896, 86/677146), 21504-6 (26/F63364, 86/868885), 21504-8 (26/F63365), 21504-10 (86/642711, 86/677154). 6. Article Number (Lot Numbers): 21513-3 (86/677385), 21513-5 (86/677389), 21514-3 (26/F63935), 21514-5 (86/642722, 86/677183, 86/869183). 7. Article Number (Lot Numbers): 37151-5 (18/561228), 37151-8 (18/166019), 37151-9 (18/625633), 37151-10 (18/626634), 37151-12 (18/628636, 18/671232), 37151-13 (18/166423), 37151-15 (18/206189), 37151-16 (18/132035, 18/691234), 37151-18 (18/142036), 37151-24 (18/167068, 18/172039), 37251-8 (18/421197), 37251-9 (18/591223), 37251-11 (18/619644), 37251-12 (18/620645, 18/962052), 37251-13 (18/521703), 37251-14 (18/531704, 18/951141), 37251-16 (18/002056, 18/031952, 18/623648, 18/971143), 37251-20 (18/571708, 26/F61731). Distribution: US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W. Affected quantity: 1,307 units
Hazard
Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

17 products in this recall

I.T.S. bone plates and screws recalled for MRI heating risk

I.T.S. is recalling 17 orthopedic implant product lines, including bone plates and screw systems. Updated MRI safety testing found that these implants can heat up more during certain MRI conditions than the product instructions previously stated. No injuries have been reported.

The agency lists each product separately. Check every one you might have.

Related topics