Moderate riskFDA DeviceClass II

GEM Premier bar code wand recalled for wrong patient ID risk

Instrumentation Laboratory is recalling bar code reader (BCR) wands used with the GEM Premier 5000 and GEM Premier 7000 portable blood analyzers. The wands may assign patient results to an incorrect patient ID. The notice reports no injuries.

From the FDA notice · Read the official notice ↗ · All Instrumentation Laboratory recalls

What to do

  1. Check whether any BCR wand, part 00024015859 (UDI 08426950869634), received a configuration update on or after April 8, 2026, and contact Instrumentation Laboratory and your care team about it before using it with patient samples.
  2. Identify the affected systems by model and UDI: GEM Premier 5000, part 00024019255 (UDI 08426950870227), and GEM Premier 7000, part 00000015279 (UDI 0195226000924).
  3. The notice doesn't say how to get a fix or replacement; contact Instrumentation Laboratory for next steps.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 6 products — see all of them.

Details

Units affected
122
Illnesses reported
None reported
Manufactured in
180 Hartwell Rd, N/A, Bedford, MA, United States
Classification
Class II
Recall number
Z-2821-2026

In the FDA's words

Description
GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859 Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID. GEM Premier 5000 Part number: 00024019255, UDI: 08426950870227 GEM Premier 7000 Part number: 00000015279, UDI: 0195226000924 Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634 Distribution: US Nationwide distribution. Affected quantity: 122
Hazard
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

6 products in this recall

GEM Premier BCR wands recalled for patient ID error risk

Instrumentation Laboratory is recalling bar code reader (BCR) wands used with GEM Premier 5000 and GEM Premier 7000 portable blood analyzer systems. The wands could potentially assign patient results to an incorrect patient ID. The recall covers six product entries, and the notice reports no injuries.

The agency lists each product separately. Check every one you might have.

More from Instrumentation Laboratory

See all

Related topics

OtherSoftwareMedical devicesSold nationwideMore from FDA Device