Moderate riskFDA DeviceClass II

GEM Premier 5000 wands recalled for potential patient ID mix-ups

Bar Code Reader (BCR) wands used with the GEM Premier 5000 and GEM Premier 7000 systems may assign patient results to an incorrect patient ID. The affected wands are those that received a configuration update on or after 8 April 2026. The notice reports 67 affected units distributed nationwide in the US, and no injuries.

From the FDA notice · Read the official notice ↗ · All Instrumentation Laboratory recalls

What to do

  1. Set aside affected BCR wands (Part No. 00024015859, UDI 08426950869634) that received a configuration update on or after 8 April 2026.
  2. Check the wand's update date against 8 April 2026 to identify affected units.
  3. The notice doesn't say how to get a remedy; contact Instrumentation Laboratory.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer system
Sold in
Nationwide
Made for
All ages (0 months and older)

This recall covers 6 products — see all of them.

Details

Units affected
67
Illnesses reported
None reported
Manufactured in
180 Hartwell Rd, N/A, Bedford, MA, United States
Classification
Class II
Recall number
Z-2824-2026

In the FDA's words

Description
Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer system Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID. GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in United Kingdom: GEM Premier 5000 Serial Nos.: 25110547, 25100462, 25110522, 25100472, 25101897, 25101920, 25101931, 25122029, 26012032, 25122030, 26012033, 26012043, 25110523, 25110536, 25120550, 25069638, 23101109, 24011137, 25081799, 25081797, 25091839, 25091840, 25091864, 25091870, 25091868, 25091873, 25091880, 25091876, 25101886, 25101889, 25101887, 25101891, 25101907, 25101904, 25101912, 25101905, 25101917, 25101921, 25101926, 25101929, 25101934, 25101927, 25111968, 25121986, 25121985, 25121981, 25121984, 25121994, 25121995, 25121982, 25121993, 25121991, 25121999, 25121996, 16030255, 17060782, 17060784, 17060785, 17060789, 17060790, 17060797, 17111126, 17121258, 18011414, 20054515, 20054516 and 22082271. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634 Distribution: US Nationwide distribution. Affected quantity: 67
Hazard
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

6 products in this recall

GEM Premier BCR wands recalled for patient ID error risk

Instrumentation Laboratory is recalling bar code reader (BCR) wands used with GEM Premier 5000 and GEM Premier 7000 portable blood analyzer systems. The wands could potentially assign patient results to an incorrect patient ID. The recall covers six product entries, and the notice reports no injuries.

The agency lists each product separately. Check every one you might have.

More from Instrumentation Laboratory

See all

Related topics

OtherSoftwareMedical devicesSold nationwideMore from FDA Device