GEM Premier 5000 barcode wands recalled for wrong patient ID risk
Instrumentation Laboratory is recalling barcode reader (BCR) wands used with the GEM Premier 5000 and GEM Premier 7000 systems. The wands may assign patient results to an incorrect patient ID. The notice reports no injuries.
From the FDA notice · Read the official notice ↗ · All Instrumentation Laboratory recalls
What to do
- Check whether your facility uses BCR wand part 00024015859 (UDI 08426950869634) that received a configuration update on or after 8 April 2026.
- Check the GEM Premier 5000 serial numbers listed in the notice: 25100466, 24097933 and 23064897.
- Contact Instrumentation Laboratory and your care team for instructions. The notice does not state a remedy.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system
- Sold in
- Nationwide
- Made for
- All ages (0 months and older)
This recall covers 6 products — see all of them.
Details
- Units affected
- 3
- Illnesses reported
- None reported
- Manufactured in
- 180 Hartwell Rd, N/A, Bedford, MA, United States
- Classification
- Class II
- Recall number
- Z-2826-2026
In the FDA's words
- Description
- Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID. GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Argentina: GEM Premier 5000 Serial Nos.: 25100466, 24097933 and 23064897. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634 Distribution: US Nationwide distribution. Affected quantity: 3
- Hazard
- Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
6 products in this recall
GEM Premier BCR wands recalled for patient ID error risk
Instrumentation Laboratory is recalling bar code reader (BCR) wands used with GEM Premier 5000 and GEM Premier 7000 portable blood analyzer systems. The wands could potentially assign patient results to an incorrect patient ID. The recall covers six product entries, and the notice reports no injuries.
The agency lists each product separately. Check every one you might have.
- GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer sys…Z-2821-2026 · 122
- Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is…Z-2822-2026 · 4 US
- Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer systemZ-2823-2026 · 27
- Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer systemZ-2824-2026 · 67
- Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer systemZ-2825-2026 · 21
- Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 Portable blood analyzer system (this page)Z-2826-2026 · 3
More from Instrumentation Laboratory
See all- HemosIL AcuStar HIT-IgG test recalled for possible result bias
SeriousFDAOct 21, 2026