Moderate riskFDA DeviceClass II

GEM Premier 7000 analyzer barcode wands recalled for patient ID risk

Instrumentation Laboratory is recalling the GEM Premier 7000 with iQM3 portable blood analyzer. Bar code reader wands used with the GEM Premier 5000 and GEM Premier 7000 systems have the potential to assign patient results to an incorrect patient ID. No injuries have been reported.

From the FDA notice · Read the official notice ↗ · All Instrumentation Laboratory recalls

What to do

  1. Contact Instrumentation Laboratory before using affected GEM Premier 7000 units or bar code reader wands, and ask for instructions.
  2. Check whether your GEM Premier 7000 is serial number 26012062, 26012065, 26012069 or 26012070.
  3. Check whether a bar code reader wand (part 00024015859, UDI 08426950869634) received a configuration update on or after April 8, 2026.
  4. The notice does not specify a remedy; contact Instrumentation Laboratory for next steps.

Is yours affected?

Compare your label or packaging with the recalled product.

Product
Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer
Sold in
Nationwide
Made for
All ages (under 100 years)

This recall covers 6 products — see all of them.

Details

Units affected
4 US
Illnesses reported
None reported
Manufactured in
180 Hartwell Rd, N/A, Bedford, MA, United States
Classification
Class II
Recall number
Z-2822-2026

In the FDA's words

Description
Brand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Intelligent Quality Management 3 (iQM3) is used as the quality control (QC) and assessment system for the GEM Premier 7000. portable blood analyzer Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID. GEM Premier 7000: Part number: 00000015279 UDI: 0195226000924 Units distributed in US: GEM Premier 7000 Serial Nos. 26012062, 26012065, 26012069 and 26012070. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634 Distribution: US Nationwide distribution. Affected quantity: 4 US
Hazard
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

6 products in this recall

GEM Premier BCR wands recalled for patient ID error risk

Instrumentation Laboratory is recalling bar code reader (BCR) wands used with GEM Premier 5000 and GEM Premier 7000 portable blood analyzer systems. The wands could potentially assign patient results to an incorrect patient ID. The recall covers six product entries, and the notice reports no injuries.

The agency lists each product separately. Check every one you might have.

More from Instrumentation Laboratory

See all

Related topics

OtherSoftwareMedical devicesSold nationwideMore from FDA Device