BD Needle Kit for Powered Driver recalled for obturator locking risk
BD Needle Kits used for emergency intraosseous (IO) access may have an obturator (the needle's internal plunger) that locks to the needle hub and becomes difficult or impossible to remove after placement. This locking is caused by out-of-tolerance manufacturing dimensions. If the obturator cannot be removed, a new needle may need to be placed, potentially delaying treatment.
From the FDA notice · Read the official notice ↗ · All Bard recalls
What to do
- Check your inventory for the affected needle kit models and lot numbers listed in the recall notice.
- Do not use affected kits for new procedures; contact BD or your distributor for replacement stock.
- If an obturator locks after needle placement, follow your facility's protocols for safe removal and secondary access placement.
- Report any obturator locking incidents to BD and the FDA MedWatch program.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK
- Sold in
- US: MA, FL, IA, SC, IL, NY, UT, NJ, ID, IN, PA, CA, TX, OH, MO, VA, MD, NC, AL, MN, CO, KY, LA, NV, OK, MS, AZ, MI, GA, AR, KS, TN. OUS: CA, SG
- Made for
- All ages (under 100 years)
Details
- Units affected
- 80,200
- Illnesses reported
- None reported
- Manufactured in
- 605 N 5600 W, Salt Lake City, UT, United States
- Classification
- Class I
- Recall number
- Z-2872-2026
In the FDA's words
- Description
- BD Needle Kit for Powered Driver 15mm x 15Ga, REF: D015151NK, BD Needle Kit for Powered Driver 25mm x 15Ga, REF: D015251NK, BD Needle Kit for Powered Driver 35mm x 15Ga, REF: D015351NK, BD Needle Kit for Powered Driver 45mm x 15Ga, REF: D015451NK, BD Needle Kit for Powered Driver 55mm x 15Ga, REF: D015551NK After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy REF/UDI-DI/Lot(Expiration): D015151NK/00801741163586/BSLC9275(2026-08-03), BSLC9563(2026-08-13), BSLC9567(2026-08-25), BSLC10103(2026-09-27), BSLC10516(2026-11-01), BSLC10526(2026-11-16), BSLC10872(2026-12-09), BSLC11216(2028-07-02), BSLC11456(2028-07-23); D015251NK/00801741163678/BSLC9467(2026-08-10), BSLC9575(2026-09-20), BSLC10105(2026-10-04), BSLC10522(2026-11-07), BSLC10868(2026-12-02), BSLC11212(2028-06-18), BSLC11454(2028-07-17), BSLC11463(2028-08-07), BSLC11884(2028-09-15), BSLC11890(2028-09-18), BSLC12082(2028-10-08), BSLC12115(2028-10-20), BSLC12123(2028-10-22); D015351NK/00801741163593/BSLC11458(2028-07-24); 00801741163678/BSLC9277(2026-08-04), BSLC9485(2026-08-12), BSLC9569(2026-08-26), BSLC10099(2026-09-07), BSLC10514(2026-10-16), BSLC10518(2026-10-21), BSLC10524(2026-11-08), BSLC10870(2026-12-03), BSLC10874(2028-06-12), BSLC11218(2028-07-07); D015451NK/00801741163609/BSLC9279(2026-08-06), BSLC9565(2026-08-17), BSLC9571(2026-09-07), BSLC10101(2026-09-17), BSLC10456(2026-10-14), BSLC10520(2026-10-22), BSLC10866(2026-11-21), BSLC11214(2028-06-23), BSLC11450(2028-07-09), BSLC11460(2028-07-31), BSLC11873(2028-08-13), BSLC11888(2028-09-25), BSLC12007(2028-10-02), BSLC12105(2028-10-15); D015551NK/00801741163616/BSLC9465(2026-08-07), BSLC9573(2026-09-26), BSLC10107(2026-10-02), BSLC10528(2026-11-20), BSLC11452(2028-07-10) Distribution: US: MA, FL, IA, SC, IL, NY, UT, NJ, ID, IN, PA, CA, TX, OH, MO, VA, MD, NC, AL, MN, CO, KY, LA, NV, OK, MS, AZ, MI, GA, AR, KS, TN. OUS: CA, SG Affected quantity: 80,200
- Hazard
- After placement of intraosseous (IO) needle sets, used when intravenous access is difficult or impossible to obtain in emergent, urgent, or medically necessary cases, users may experience difficulty/inability to remove obturator, due to it locking to the needle hub, due to out-of-tolerance dimensions and IFU unclarity, which may result in new needle placement, which may result in delayed therapy