- 2025-10-29GE HealthcareClass IIOptima NX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIMillenium MPS; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIMillenium MPR; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIMillenium Myosight; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIMillenium MG/MC; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIMaxicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIElscint Helix; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIElscint Varicam; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIElscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIElscint SPX4; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIElscint Model 4XX; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29GE HealthcareClass IIElscint Cardial; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging
GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- 2025-10-29LeMaitre VascularClass IIArtegraft¿ Collagen Vascular Graft; REF: AG1015;
The device was incorrectly packed in the wrong size labeled outer packaging.
- 2025-10-29HeartSineClass IIHeartSine Pad-Pak, REF: PAD-PAK-01, PAD-PAK-02, PAD-PAK-03, PAD-PAK-04, PAD-PAK-07, PAD-PAK-03J, PAD-PAK-04J
Due to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.
- 2025-10-29SedecalClass IISM_40HF_BDC3, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IISM_40HF_BDC, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IISM_40HF_B_D_VIR3.004, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IISM_40HF_B_D_VIR3, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IISHIMADZU RF.004, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IISED URS LP, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIRADPRO MOBILE, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW70CDEMO.005, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW.005, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW.002, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW CANON READY.005, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW CANON READY, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW 80C, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW 70C READY, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.
- 2025-10-29SedecalClass IIModel Number SM-40HF-B-D-C; 40KW 70C EN, Mobile X-ray system
Sedecal sent an email to its consignees to remind the users that the equipment is not water-resistant and must be cleaned strictly according to the instructions in the manual, clearly outlining the possible consequences of improper cleaning.