- 2025-10-29Cardinal Health 200, LLCClass IISalem Sump PVC Tubes: Product Code/Product Name: 1180264408 Salem Sump" PVC Gastroduodenal Tube, 10 Fr/Ch (3.3 mm) x 90 cm 1180264416 Salem Sump" PVC Gastroduodenal Tube, 12 Fr/Ch (4.0 mm) x 114 cm 1180264424 Salem Sump" PVC Gastroduodenal Tube, 14 Fr/Ch (4.7 mm) x 114 cm 1180264432 Salem Sump" PVC Gastroduodenal Tube, 16 Fr/Ch (5.3 mm) x 114 cm 1180264440 Salem Sump" PVC Gastroduodenal Tube, 18 Fr/Ch (6.0 mm) x 114 cm 1180264457 Salem Sump" PVC Gastroduodenal Tube, 20 Fr/Ch (6.7 mm) x
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
- 2025-10-29Cardinal Health 200, LLCClass IISalem Sump PVC Tubes: Product Code/Product Name: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) 266130 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 16 Fr/Ch (5.3 mm) 266148 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 18 Fr/Ch 8888266106 Salem Sump" Dual Lumen Stomach Tube, Anti-Reflux Valve 10Fr/Ch (3.3 mm), 36" 266130CN Salem Sump" Tube 16FR with CONFIRM NOW, Nonsterile 266148CN Salem Sump" Tube 8 FR with CONFIRM
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
- 2025-10-29Cardinal Health 200, LLCClass IISalem Sump PVC Tubes: Product Code/Product Name: 266197 Salem Sump" Anti-Reflux Valve CE Marked 8888266197 Salem Sump" Anti-Reflux Valve
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
- 2025-10-29Cardinal Health 200, LLCClass IISalem Sump PVC Tubes: Product Code/Product Name: 8888265108 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36 (91 cm) 8888265116 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48 (122 cm) 8888265124 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48 (122 cm) 8888265132 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 16 Fr/Ch (5.3 mm), 48 (122 cm) 8888265140 Salem Sump Si
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
- 2025-10-29Cardinal Health 200, LLCClass IISalem Sump PVC Tubes: Product Code/Product Name: 264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 264960 Salem Sump" Stomach Tube, Dual Lumen, 16 Fr/Ch (5.3 mm) 264986 Salem Sump" Stomach Tube, Dual Lumen, 18 Fr/Ch (6.0 mm) 5558264957 Salem Sump" COILED TUBE 16Fr/Ch (5.3 mm) 5558264965 Salem Sump" TUBE 16 Fr/Ch (5.3 mm) 5558264973 Salem Sump" COILED TUBE 18Fr/Ch (6.0 mm) 5558264981 Salem Sump" TUBE 18 Fr/Ch 1
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
- 2025-10-29Cardinal Health 200, LLCClass IISalem Sump PVC Tubes: Product Code/Product Name: 8888264911E ENFit" SALEM SUMP" PVC TUBE 10FR 36IN 8888264929E ENFit" SALEM SUMP" PVC TUBE 12FR 48IN 8888264945E ENFit" SALEM SUMP" PVC TUBE 14FR 48IN 8888264960E ENFit" Salem Sump" PVC Tube 16Fr 48IN 8888264986E ENFit" SALEM SUMP" PVC TUBE 18FR 48IN 8888266106E PVC SALEM SUMP WITH ENFIT & ARV 10FR 36IN 8888266114E PVC SALEM SUMP WITH ENFIT & ARV 12FR 48IN 8888266122E PVC SALEM SUMP WITH ENFIT & ARV 14FR 48IN 8888266130E PVC SALEM SUM
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
- 2025-10-29ICU Medical, Inc.Class IIV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave,
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
- 2025-10-29Abiomed, Inc.Class IAutomated Impella Controller (AIC) labeled as the following with corresponding Product Codes: 1. AIC w/Impella Connect for ECP, Product Code: 1000432. 2. Dbl optical, AIC Impella Connect, Phg US, Product Code: 1000201. 3. Impella Controller, Packaged, AU, Product Code: 0042-0000-AU. 4. Impella Controller, Packaged, CA, Product Code: 0042-0000-CA. 5. Impella Controller, Packaged, EU, Product Code: 0042-0000-EU. 6. Impella Controller, Packaged, IN, Product Code: 0042-0000-IN. 7. Impella Con
Potential for Automated Impella Controller (AIC) purge pressure issues due to purge retainer failures.
- 2025-10-29B BRAUN MEDICAL INCClass IBrand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version: N/A Product Description: STREAMLINE BLOODLINE SET FOR DIALOG Component: N/A
The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.
- 2025-10-22Diagnostica Stago, Inc.Class IIProduct: STA Liatest D-Di; REF: 00515;
A potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM NON-DEHP CONTINU-FLO SOLN SET, 96" (2,4 m), Product Code: 2H8519
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK System Paclitaxel Set, Non-Vented with Polyethylene Lined Tubing and CLEARLINK Luer Activated Valve, 0.2 Micron Filter, CLEARLINK Luer Activated Valve, 10 drops/mL, 107-inch (2.7 meters), Product Code 2R8858
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, three CLEARLINK Luer Activated Valves, Backcheck Valve, 10 drops/mL, 120-inch (3.0 meters), Non-DEHP, Product Code 2R8537
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, two CLEARLINK Luer Activated Valves. 10 drops/mL, 96-inch (2.4 meters). Non-DEHP, Product Code 2R8519
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK System Non_DEHP Extenstion Set, Straight-Type Extension Set, Standard Bore, 1.2 Micron Filter, CLEARLINK Luer Activated Valve, 19-inch (47 centimeters), Product Code 2H8603
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM EXTENSION SET, Straight-Type Extension Set, Standard Bore, CLEARLINK Luer Activated Valve, Retractable Collar, 7.9-inch (20 centimeters), Product Code 2C8634
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM EXTENSION SET, Straight-Type Extension Set, Standard Bore, CLEARLINK Luer Activated Valve, Retractable Collar, 14-inch (35 centimeters), Product Code 2C8632
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM EXTENSION SET, Straight-Type Extension Set, Standard Bore, CLEARLINK Luer Activated Valve, Retractable Collar, 44-inch (111 centimeters), Product code 2C8610
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, 0.2 Micron Filter, three CLEARLINK Luer Activated Valves, Backcheck Valve, 10 drops/mL, 102-inch (2.6 meters), Product Code 2C8593
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, 0.2 Micron Filter, two CLEARLINK Luer Activated Valves, Backcheck Valve, 10 drops/mL, 104-inch (2.6 meters), Product Code 2C8571
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, three CLEARLINK Luer Activated Valves, Backcheck Valve, Retractable Collar, 10 drops/mL, 120-inch (3.0 meters), Product Code 2C8541
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, three CLEARLINK Luer Activated Valves, Backcheck Valve, Retractable Collar, 10 drops/mL, 120-inch (3.0 meters), Product Code 2C8537
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM CONTINU-FLO Solution Set, Non-Vented, two CLEARLINK Luer Activated Valves, Backcheck Valve, 10 drops/mL, 112-inch (2.8 meters), Product Code 2C8519
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK SYSTEM Solution Set, Non-Vented, two CLEARLINK Luer Activated Valves, 10 drops/mL, 100-inch (2.5 meters), Product Code 2C8425
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICLEARLINK System Solution Set, DUO-VENT, CLEARLINK Luer Activated Valve, 10 drops/mL, 92-inch (2.3 meters), Product code 2C8419
IV sets may leak.
- 2025-10-22Baxter Healthcare CorporationClass IICONTINU-FLO Solution Set, Non-Vented, two CLEARLINK Luer Activated Valves, Backcheck Valve, Plus 4-Way Stopcock Extension Set with two CLEARLINK Luer Activated Valve, Pfoduct code 2C6255
IV sets may leak.
- 2025-10-22American Contract Systems Inc.Class IIMedical convenience kits IHDC30R D AND C PACK
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- 2025-10-22American Contract Systems Inc.Class IIMedical convenience kits UDPG88W PICC G-TUBE PACK-230209
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- 2025-10-22American Contract Systems Inc.Class IIMedical convenience kits Item Number/Description BFCT04X CATARACT PACK - 205986 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 CMEY17AI EYE PACK - 213187 PERP25AQ PEI RETINAL PACK-206025 RFEY50AC EYE PACK-206026 UIVT47AS VITRECTOMY SUPPLY
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
- 2025-10-22American Contract Systems Inc.Class IIMedical convenience kits Item Number/Description ANSH14AF SHOULDER PACK - 205962 EVAH76H ANTERIOR HIP PACK FDSA15M SHOULDER ARTHROSCOPY GRTS90G TOTAL SHPULDER PACK HGHI26G HIP PINNING PACK HJTO25C TOTAL HIP ACCESSORY PACK HNEX60L EXTREMITY PACK IAHD53C HAND PACK IASH65A SHOULDER PACK IHUE40AI UPPER EXTREMITY PK| STRL F G ILTJ47AA TOTAL JOINT PACK LMHP14Z ORIF HIP, ST LUKES METHODIS LMSN44AE SPINAL PACK MHHP25H HAND PACK - 206002 MHTH90AJ TOTAL HIP PACK - 206016 MWAH32P ANTER
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.