Boston Scientific Neuromodulation Corporation
6 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Boston Scientific Neuromodulation Corporation and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-09-24FDA-DeviceClass IIVercise Deep Brain Stimulation Systems Surgical Implant Manual, Document Number: 92328632-09
Wire break(s) have occurred in rechargeable deep brain stimulation implantable pulse generators, implanted in the submuscular pectoral location. Wire breaks prevent successful stimulation, thus requiring device removal/replacement and users may experience high monopolar impedances, undesired sensation, loss of therapy, return of symptoms and/or Bluetooth connectivity challenges.
- 2025-09-10FDA-DeviceClass IISuperion IDS Kit Superion Indirect Decompression System 8MM, VF Implant 101-9808 00884662000529 Superion Indirect Decompression System 10MM, VF Implant 101-9810 00884662000536 Superion Indirect Decompression System 12MM, VF Implant 101-9812 00884662000543 Superion Indirect Decompression System 14MM, VF Implant 101-9814 00884662000550 Superion Indirect Decompression System 16MM, VF Implant 101-9816 00884662000567 Superion Indirect Decompression System Kit, VF Instrument
As a result of a comprehensive product performance review it was determine that the device and instruments are not consistently meeting expected performance levels. Issues include breakage of the instruments and implant/device.
- 2025-08-27FDA-DeviceClass IIBurr Hole Cover Kit, REF: M365DB4600C0; Burr Hole Cover Spares Kit, REF: M365DB4605C0. Components of Vercise Deep Stimulation System used with the Surgical Implant Manual
Resistance and difficulties could occur while closing the retaining clip locking mechanism, which is part of a burr hole cover kit, which is part of a Deep Brain Stimulation (DBS) system, which may lead to procedural delay, or lead-migration-which may require additional surgery.
- 2024-08-21FDA-DeviceClass IIWAVEWRITER ALPHA: Model: SC-1216, 16 IPG KIT, REF: M365SC12160; Model: SC-1232, 32 IPG KIT, REF: M365SC12320
Spinal Cord Stimulation (SCS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset, which may lead to transient loss of stimulation; patients may experience undesired sensations when therapy turns off for approx. 10-15 seconds and then back on, which may lead patient to request surgical intervention for replacement or revision.
- 2024-06-05FDA-DeviceClass IIVercise Genus Deep Brain Stimulation (DBS) Implantable Pulse Generator IPG: R16 IPG KIT, REF: M365DB12160, and R32 IPG KIT, REF: M365DB12320
Deep Brain Stimulation (DBS) Implantable Pulse Generator (IPG) may experience routine system check during IPG charging, which may cause device reset. The device reset could lead transient loss of stimulation; patients may experience undesired sensations, transient worsening of movement disorder symptoms, which may lead patient to request surgical intervention for replacement or revision.
- 2023-06-21FDA-DeviceClass IIDriver Instrument, part of the Superion Indirect Decompression System, REF: 102-9800, used with the Superion Indirect Decompression System IFU, Superion IDS Kit IFU, Surgical Technique Manual
Pending update to indirect decompression system instructions for use informing users that excessive force during the implant procedure may cause driver instrument tip breaks, which may result in metal fragments (Driver teeth/tips) within the implant location; and if metal fragments are not removed and remain in situ, MRI scans are NOT advised due to potential risk of patient injury.