Hamilton Medical, Inc.
16 federal recalls on file, showing the 3 from the last 12 months. We monitor every CPSC, FDA, USDA, and NHTSA notice for Hamilton Medical, Inc. and translate it into plain English — free weekly digest, optional same-day alerts.
- 2026-07-08FDA-DeviceClass IIHamilton Medical IntelliCuff recalled for motor malfunction risk
Due to reported complaints, the cuff device may alarm with a "Cuff System Leakage" error. As a result, the motor may pump continuously and fail to maintain the cuff pressure set by the user.
- 2026-07-08FDA-DeviceClass IHamilton Medical ventilator breathing circuits recalled for valve obstruction risk
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
- 2025-08-27FDA-DeviceClass IHamilton C6 REF 160021 (ESM Board PN 10144626 UDI-DI code: 76300028PN160021ZW Hamilton C6 is a ventilator intended to provide positive pressure ventilatory support to adults and pediatrics and optionally infants and neonates.
A mechanical defect on a printed circuit board (PCB) as a result of PCB separating process. This defect might cause the ventilator to either fail at startup (no patient involvement) or during the use of the device enter ambient state, resulting in an interruption of ventilation.