Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2025-03-12FDA-DeviceMedtronic MiniMed, Inc.Class IIParadigm insulin pump, REF: MMT-522, MMT-523, MMT-551, MMT-722, MMT-723, MMT-751
All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death.
- 2025-03-12FDA-DeviceMedtronic MiniMed, Inc.Class IIParadigm REF: MMT-712 and MMT-715
All Medtronic MiniMed pumps (Paradigm series, Minimed 530G, 6XX series, and 7XX series) were found to have the potential for abnormal insulin delivery due to changes in air pressure, which can be found during airplane take-off or landing. An unexpected additional dose of insulin, which can occur during take-off, can result in severe hypoglycemia, altered mental status, seizure, coma, or death. An unexpected under-delivery of insulin, which can occur during landing, can result in severe hyperglycemia, dehydration, diabetic ketoacidosis, or death. During a delivery accuracy dry run testing of the 780G NGP infusion pump under dynamic pressure conditions, the firm observed that when air pressure decreased during take-off simulation, on average 0.569U more insulin was delivered in 15 minutes than the expected delivery volume at ambient pressure at 1 U/hr delivery rate; when air pressure increased during landing simulation, on average 0.537 U less insulin was delivered in 25 minutes than the expected delivery volume at ambient pressure at 1 U/hr delivery rate. The firm also observed that the mechanisms causing the variations in delivery volumes are independent of the delivery rate.
- 2025-03-12FDA-DeviceBaxter Healthcare CorporationClass IBaxter Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009
There is the potential for missing motor mounting screws, which may have occurred during the servicing process.
- 2025-03-12FDA-DeviceBaxter Healthcare CorporationClass IBaxter Sigma Spectrum Infusion System (V6 Platform), Product Code 35700BAX
There is the potential for missing motor mounting screws, which may have occurred during the servicing process.
- 2025-03-12FDA-DevicePhilips Medical Systems Nederland B.V.Class IIPhilips IntelliSpace Cardiovascular software, Model 830089.
Study data is not able to be archived, copied, or exported with the cardiovascular software version.
- 2025-03-12FDA-DeviceHUMAN MED AGClass IIhuman med Biofill Infiltration Cannula, diameter 2.5mm, 15cm, REF 2024-012, REF 1580107, sterile, single use.
The cannula tubes do not comply with the specification. The water jet emerging from the nozzle of the nonconforming cannulas may emerge as a spot jet rather than a fan-shaped spray jet.
- 2025-03-12FDA-DeviceMedimaps GroupClass IIBrand Name: TBS Osteo Product Name: TBS iNsight Model/Catalog Number: TBS iNsight v3.x / pn: OS-TBSi Software Version: V3.x: (1) V3.1.2, (2) V3.1.1, (3) V3.1.0, (4)V3.0.0 Product Description: TBS iNsight is a class II Medical Device software that is installed on bone densitometers for analysis of bone microarchitecture and bone health management Component: No
Potential variability in calculations from fast array scans compared to array scans when operating on Hologic Horizon machines.
- 2025-03-12FDA-DeviceBeckman Coulter, Inc.Class IIAccess Erythropoietin (EPO), Catalog Number - A16364 a glycoprotein (~30,400 Daltons) produced primarily by the kidney, is the principal factor regulating red blood cell production (erythropoiesis) in mammals. Renal production of EPO is regulated by changes in oxygen availability. Under conditions of hypoxia, the level of EPO in the circulation increases and this leads to increased production of red blood cells.
Affected lot (439363) exhibited a negative dose drop of -22% with native patient samples compared to alternate reagent lots. Patient samples tested with the affected lot may demonstrate repeatable falsely decreased results, which may lead to improper diagnosis or repeat testing.
- 2025-03-12FDA-DeviceMicro Therapeutics, Inc.Class IMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-021-350-16, PED3-021-250-14, PED3-021-350-20, PED3-021-250-20, PED3-021-300-16, PED3-021-300-20, PED3-021-275-12, PED3-021-275-16, PED3-021-325-12, PED3-021-325-14, PED3-021-325-16, PED3-021-325-20, PED3-021-250-12, PED3-021-300-14, PED3-021-250-16, PED3-021-300-12, PED3-021-275-14, PED3-021-350-14, PED3-021-350-25, PED3-021-350-12, PED3-021-250-10, PED3-021-275-20, PED3-021-350-18, PED3-021-325-18, PED3-021-275-18,
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
- 2025-03-12FDA-DeviceMicro Therapeutics, Inc.Class IMedtronic Pipeline Vantage Embolization Device with Shield Technology, REF: PED3-027-350-12, PED3-027-350-14, PED3-027-350-16, PED3-027-350-20, PED3-027-350-25, PED3-027-400-12, PED3-027-400-14, PED3-027-400-16, PED3-027-400-18, PED3-027-400-20, PED3-027-400-25, PED3-027-400-30, PED3-027-450-12, PED3-027-450-14, PED3-027-450-16, PED3-027-450-18, PED3-027-450-20, PED3-027-450-25, PED3-027-450-30, PED3-027-450-40, PED3-027-500-14, PED3-027-500-16, PED3-027-500-18, PED3-027-500-20, PED3-027-500-25,
Use of embolization device with diameters greater-than-or-equal-to 4mm may lead to incomplete wall apposition and/or braid deformation (fish-mouthing/braid narrowing/braid collapse) noted during procedure and post-procedure, typically noted at 6-12 month imaging follow-up, potentially can lead to thrombosis and/or serious adverse events. Higher risk in females less-than-or-equal-to 45 years of age
- 2025-03-12FDA-DeviceBard Peripheral Vascular IncClass IRotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239
Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.
- 2025-03-12FDA-DeviceBiosense Webster, Inc.Class IVARIPULSE Bi-Directional Ablation Catheter REF D141201. The Field Catheter is indicated for use in catheter based cardiac electrophysiological mapping (stimulating and recording) and, when used for TRUPULSE Generator, for the treatment of drug refractory recurrent symptomatic paroxysmal atrial fibrillation. The catheter provides location information when used with the CARTO 3 System.
Due to an observed trend of neurovascular events
- 2025-03-05FDA-DeviceTERRAGENE S.A.Class IIASP BIOTRACE Auto Read 60 Steam BI Process Challenge Device, REF 73135.
Error in the IFU lists an incorrect incubation time on one occasion of 20 minutes rather than the correct incubation time of 60 minutes.
- 2025-03-05FDA-DeviceLEICA BIOSYSTEMS NUSSLOCH GMBHClass IIBrand Name: Leica Biosystems Cryostats Product Name: CM1950 Model/Catalog Number: 1491950C4US Software Version: firmware V3.01
Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue.
- 2025-03-05FDA-DeviceGE Healthcare (China) Co., Ltd.Class IISIGNA MR380, NMRI system, Non-US Only
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DeviceGE Healthcare (China) Co., Ltd.Class IISIGNA MR355, SIGNA MR360, NMRI system
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DeviceGE Healthcare (China) Co., Ltd.Class IIOptima MR360, NMRI system
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DeviceGE Healthcare (China) Co., Ltd.Class IIBrivo MR355, NMRI system
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DeviceGE Healthcare (China) Co., Ltd.Class IISIGNA Creator, SIGNA Explorer, NMRI systems
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
- 2025-03-05FDA-DevicePentax of America IncClass IIPentax Medical Video Processor- Intended to be used with PENTAX Medical endoscopes, monitors and other peripheral devices for endoscopic diagnosis, treatment, and video observation. Model: EPK-i8020c
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
- 2025-03-05FDA-DevicePentax of America IncClass IIPENTAX Medical Video Upper GI Scope-Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EG29-i20c
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
- 2025-03-05FDA-DevicePentax of America IncClass IIPENTAX Medical Video Colonoscope- Intended to provide optical visualization of (via a video monitor), and therapeutic access to the lower gastrointestinal tract Model: EC38-i20cL
During endoscopic procedures using a combination of the video processor EPK-i8020c and the i20c series video endoscope, the observed image can become reddish or dark (1.8%). Users have observed smoke-like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient, and may cause thermal injury to the patient's mucous membranes.
- 2025-03-05FDA-DeviceScientia Vascular, Inc.Class IIPlato 17 Microcatheter REF PL17-160-090 To assist in delivery of therapeutic devices and embolization materials and infusion of diagnostic agents to the neuro and peripheral vasculature during interventional or diagnostic procedures.
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
- 2025-03-05FDA-DeviceScientia Vascular, Inc.Class IISocrates 38 Aspiration Catheter 127 cm length REF SC038-127-001 The Socrates Aspiration System with a compatible suction pump is intended for use in the revascularization of patients with acute ischemic stroke
Due to manufacturing non-conformance that resulted in channels in the seal on the packaging pouch. As a result, this could potential create a breach in the sterile barrier.
- 2025-03-05FDA-DeviceBECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD.Class IIDxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.
Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.
- 2025-03-05FDA-DevicePhysio-Control, Inc.Class IILIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.
Incorrect keypad was incorrectly installed onto a defibrillator/monitor.
- 2025-03-05FDA-DeviceCareFusion 303, Inc.Class IIBD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration
Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.
- 2025-03-05FDA-DeviceCareFusion 303, Inc.Class II139088-01 BD PYXIS MEDBANK MEDPASS MODULE 169-151 BD PYXIS MEDBANK MINI 1FH-2HH MEDPASS 169-152 BD PYXIS MEDBANK MINI 2HH-1FM MEDPASS 169-150 BD PYXIS MEDBANK MINI 4HH MEDPASS 169-140 BD PYXIS MEDBANK TWR MN 3FH-8HH MEDPASS
Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.
- 2025-03-05FDA-DeviceCareFusion 303, Inc.Class II139054-01 BD PYXIS MEDFLEX MN 1000 10HH-1FM 139055-01 BD PYXIS MEDFLEX MN 1000 12HH 139053-01 BD PYXIS MEDFLEX MN 1000 2HH-2FH-3FM 139041-01 BD PYXIS MEDFLEX MN 1000 2HH-2FM 139046-01 BD PYXIS MEDFLEX MN 1000 2HH-5FM 139038-01 BD PYXIS MEDFLEX MN 1000 3FM 139040-01 BD PYXIS MEDFLEX MN 1000 4HH-1FM 139049-01 BD PYXIS MEDFLEX MN 1000 4HH-4FM 139044-01 BD PYXIS MEDFLEX MN 1000 6FM 139039-01 BD PYXIS MEDFLEX MN 1000 6HH 139051-01 BD PYXIS MEDFLEX MN 1000 6HH-3FM 139164-01 BD PYXIS MEDFLEX MN 1000 8H
Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.
- 2025-03-05FDA-DeviceCareFusion 303, Inc.Class II155288-01 BD PYXIS MEDBANK MINI 1FH-1FM 169-137 BD PYXIS MEDBANK MINI 1FH-1FM 155311-01 BD PYXIS MEDBANK MINI 1FH-2HH 138921-01 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 155159-01 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 169-101 BD PYXIS MEDBANK MINI 1HH-1FH-1HM 139122-01 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 155160-01 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 169-102 BD PYXIS MEDBANK MINI 1HH-1HM-1FM 138912-01 BD PYXIS MEDBANK MINI 1HH-3HM 155161-01 BD PYXIS MEDBANK MINI 1HH-3HM 169-103 BD PYXIS MEDBANK MINI 1HH-3HM 1389
Labeling update to include a contraindication statement against the use of medicine cabinets in procedure and operating rooms.