Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2025-02-12FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD10C System Code: (1) 722001
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
- 2025-02-12FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAlluraXper FD10/10 System Code: (1) 722005 (2) 722011 (3) 722027
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
- 2025-02-12FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura Xper FD10 Catalog (1) 722003 (2) 722010 (3)722026
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
- 2025-02-12FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIAllura CV20 System Code: 722031
Potential safety issue with the AD7 and AD7X patient tables (both tilt and non-tilt versions) part of the Philips Allura and Azurion systems. During manual repositioning of the patient tabletop, a finger can get entrapped between the longitudinal guiding rails and the tabletop. This may result in an injury of the finger, applicable for operators and service personnel
- 2025-02-12FDA-DeviceOxoid LimitedClass IILev5, REF CT1587B, 50 Susceptibility discs Levofloxacin 5 ug, IVD, CE 2797
Mislabeling; Levofloxacin Antimicrobial susceptibility disc packaging may contain Norfloxacin Antimicrobial Susceptibility discs.
- 2025-02-12FDA-DeviceLivaNova USA, Inc.Class IISenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
- 2025-02-12FDA-DeviceLivaNova USA, Inc.Class IISenTiva, REF: 1000; SenTiva Duo, REF: 1000-D; Generators Part of the VNS Therapy System. Used for Vagus Nerve Stimulation (VNS).
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
- 2025-02-12FDA-DeviceROi CPS LLCClass Iregard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator. Convenience kit to be used in Labor and Delivery.
There is a potential for the Inline Controller to be using the spring out of specification causing a loss of positive pressure resulting in the Neo-Tee Resuscitator and/or Circuit will not reach the high and low ends of the PIP and PEEP pressure ranges impacting effective ventilation of the patient..
- 2025-02-12FDA-DeviceGE Medical Systems, LLCClass IIOptima XR646 HD. X-Ray imaging system.
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- 2025-02-12FDA-DeviceGE Medical Systems, LLCClass IIDiscovery XR656 HD. X-Ray imaging system.
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- 2025-02-12FDA-DeviceGE Medical Systems, LLCClass IIDiscovery XR656HD. X-Ray imaging system.
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- 2025-02-12FDA-DeviceGE Medical Systems, LLCClass IIDefinium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
- 2025-02-12FDA-DeviceFresenius Kabi USA, LLCClass IIvenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.
- 2025-02-05FDA-DeviceMurata Vios, Inc.Class IImuRata Vios Monitoring System Model 2050; bedside cardiac monitor, including cardiotachometer and alarm rate
During the set-up and workflow to begin Vios monitoring, it has been noticed that patient vital signs and ECG waveforms can become distorted and will flicker or flash very fast. This issue only occurs on the Central Station Monitor (CSM) in the patient tiles view and is not seen at any other point during patient monitoring & not on the Bedside Monitor.
- 2025-02-05FDA-DeviceLivaNova USA, Inc.Class IIBrand Name: SenTiva DUO" Product Name: VNS Therapy¿ SenTiva DUO", Model 1000-D Model/Catalog Number: 1000-D Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. The generator is an implantable, multi-programmable pulse generator that delivers electrical signals to the vagus nerve through the lead. The gener
Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
- 2025-02-05FDA-DeviceLivaNova USA, Inc.Class IIBrand Name: SenTiva" Product Name: VNS Therapy" SenTiva", Model 1000 Model/Catalog Number: 1000 Software Version: NA Product Description: The LivaNova VNS Therapy" System, used for Vagus Nerve Stimulation (VNS), consists of an implantable VNS Therapy generator, lead, and external programming system used to change stimulation settings. The generator is an implantable, multi-programmable pulse generator that delivers electrical signals to the vagus nerve through the lead. The generator is hous
Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.
- 2025-02-05FDA-DeviceDraeger Medical Systems, Inc.Class IIInfinity Central Station. Model Number: MS26800 The Infinity CentralStation (ICS) is intended for centralized monitoring of adult, pediatric and neonatal patient data within the hospital or clinical environment. Centralized monitoring involves the display and management of data from networked patient monitors including the annunciation of visual and audible physiologic parameter alarms at a central monitoring workstation.
Edifier speakers (shipped in or after April 2023) may not annunciate when configured in analog mode which can lead to delayed treatment to the patient.
- 2025-02-05FDA-DeviceAlphatec Spine, Inc.Class IICalibrate CCX Interbody System for spinal fusion procedures: Implants: (1) Calibrate CCX-O, 7 x 10 x 25mm 5 degrees, Part #325-07102505, nonsterile; (2) Calibrate CCX-O, 7 x 10 x 25mm 10 degrees, Part #325-07102510, nonsterile; (3) Calibrate CCX-O, 7 x 10 x 25mm 15 degrees, Part #325-07102515, nonsterile; (4) Calibrate CCX-O, 7 X 10 X 30mm 5 degrees, Part #325-07103005, nonsterile; (5) Calibrate CCX-O, 7 x 10 x 30mm 10 degrees, Part #325-07103010, nonsterile; (6) Calibrate CCX-O,
Complaints have been received regarding post-operative implant collapse.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27011K, Ureteroscope, 8 Fr., 6¿, 34cm, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27002K, Ureteroscope, 9.5 Fr., 6¿, 34 cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27001K, Ureteroscope, 9.5 Fr., 6¿, 34 cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27000K, Ureteroscope, 7 Fr., 6¿, 34 cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27010K, Ureteroscope, 7 Fr., 34 cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27002L, Uretero-Renoscope, 9.5 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27011L, Uretero-Renoscope, 8 Fr., 6¿, 43cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27001L, Uretero-Renoscope, 8 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27000L,Uretero-Renoscope, 7 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27010L, Uretero-Renoscope, 7 Fr., 43cm, NON STERILE, RxONLY, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27002KP, Pediatric Uretero-Renoscope 7.3 Fr., NON STERILE, CE 0123
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.
- 2025-02-05FDA-DeviceKarl Storz EndoscopyClass IIKARL STORZ - ENDOSKOPE, REF: 27030KB, Pediatric Cysto-Urethroscope, NON STERILE, CE
Scope IFUs contain a reprocessing modalities that have not been reviewed and/or approved for safety and efficacy by the FDA.