Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2024-12-25FDA-DeviceCooperSurgical, Inc.Class IIBrand Name: K-Systems Product Name: T47 Warming Plate Model/Catalog Number: K23054 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
- 2024-12-25FDA-DeviceCooperSurgical, Inc.Class IIBrand Name: K-Systems Product Name: G85 Mini Incubator Model/Catalog Number: K22074 Software Version: N/A Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxid
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
- 2024-12-25FDA-DeviceCooperSurgical, Inc.Class IIBrand Name: K-Systems Product Name: G73 Dry Bath Incubator Model/Catalog Number: K22065 Product Description: An electrically-powered device intended to provide suitable warming and/or incubation conditions for reproductive biological materials (e.g., gametes and embryos), at closely regulated and monitored temperatures, during assisted reproduction examination and processing procedures. It may support a more comprehensive growth environment [e.g., temperature, carbon dioxide (CO2) and oxygen
The warming surfaces or incubator well temperatures of the listed products may exceed the desired set temperature. If temperatures exceed the desired set value, the device will alarm appropriately.
- 2024-12-25FDA-DevicePhilips North America LlcClass IIPhilips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.
- 2024-12-25FDA-DeviceSpacelabs Healthcare, Ltd.Class IIEclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patch recorder intended to record the patient s electrocardiogram.
Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
- 2024-12-25FDA-DeviceSpacelabs Healthcare, Ltd.Class IIEclipse PRO REF 98700 The Spacelabs model 98700 is a portable non-invasive Holter recorder intended to record the patient s ambulatory electrocardiogram.
Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.
- 2024-12-25FDA-DeviceSpectranetics CorporationClass IITurbo Elite Laser Atherectomy Catheter Model Number / UDI-DI code / Product Description 414-151 / 989930000531 / 1.4 mm OTW 417-152 / 989930000551 / 1.7 mm OTW 420-006 / 989930000571 / 2.0 mm OTW 423-001 / 989930000591 / 2.3 mm OTW 425-011 / 989930000601 / 2.5 mm OTW 423-135-01 / 989930004171 / 2.3 mm OTW 423-135-02 / 989930004181 / 2.3 mm OTW 425-135-01 / 989930004201 / 2.5 mm OTW 425-135-02 / 989930004211 / 2.5 mm OTW The Turbo-Elite devices are indicated for use in the treatment,
Due to complaints received for the marker band detaching from the distal tip of the catheter and remaining in the patient as laser catheter was retracted.
- 2024-12-25FDA-DeviceDraeger Medical, Inc.Class IBrand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
- 2024-12-25FDA-DeviceDraeger Medical, Inc.Class IBrand Name: Atlan Product Name: Atlan A300 Model/Catalog Number: 8211300 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
- 2024-12-25FDA-DeviceMedtronic NeurosurgeryClass IExacta External Drainage and Monitoring System: 27581 KIT 27581 EXACTA W/EDM VCATH 35CM 27636 EDM 27636 EXACTA W/BRAIDED CORD LOCK 27666 EDM 27666 EXCTA W GLUE STPCK GR PT LN 27732 EDM 27732 EXACTA W/NDL-LS INJ. ST W/CAP 27785 EDMS 27785 EXACTA 50ML BC NDLESS INJ 27795 EDMS 27795 EXACTA 100ML CATH 27861 EDMS 27861 EXACTA 50ML ANTIRFLUX GRNPL 46700 SYSTEM 46700 EXACTA DISP. DRAINAGE 46705 SYSTEM 46705 EXACTA DRAINAGE 100ML The Exacta external drainage and monitoring system (EDMS) is a
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
- 2024-12-25FDA-DeviceMedtronic NeurosurgeryClass IBecker External Drainage and Monitoring System: 24146 EDMS 24146 BECKER NO Y-SITE 25120 EDMS II 25120 BECKER W/1-WAY VALVE 26040 KIT 26040 BECKER EDMS II 27609 BECKER 27609 W/NEEDLESLESS INJ. SITE 27670 BECKER 27670 STOP BELOW DC NDL-LS IN 27672 BECKER 27672 ONEWAY VLV Y-SITE PT LN 27702 EDMS 27702 EURO BECKER II GR PT LN 27761 BECKER 27761 STOP BCNDLESS PATLN 24 27767 BECKER 27767 NO STPCOCK ON PTLN 63 27779 EDMS 27779 BECKER BC NDLESS BOND CONN 27811 EDMS 27811 BECKER GRN PL VCATH
Due to complaints received reporting cracks and/or leaks of the stopcocks associated with the external drainage and monitoring system (EDMS).
- 2024-12-25FDA-DeviceKirwan Surgical Products, LLCClass IOptikon Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: OPT14-5011. Product Description: The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
- 2024-12-25FDA-DeviceKirwan Surgical Products, LLCClass INidek Co. Disposable Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: 18241-0979. Product Description: The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
- 2024-12-25FDA-DeviceKirwan Surgical Products, LLCClass IWalcott Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: RX14-5011. Product Description: The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
- 2024-12-25FDA-DeviceKirwan Surgical Products, LLCClass IAmbler Surgical Bipolar Pencil 25 Gauge Straight. Model/Catalog Number: PD7226. The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge. Component: No
Potential for the outer stainless-steel tube to break or detach.
- 2024-12-25FDA-DeviceKirwan Surgical Products, LLCClass IAccutome/Microsurgical Disposable Bipolar Pencil 5 in. (12.7cm) 25 Gauge Straight. Model/Catalog Number: AB-7004 The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
- 2024-12-25FDA-DeviceKirwan Surgical Products, LLCClass IKirwan Surgical Products LLC Disposable 25-gauge bipolar pencil, straight, Long. Model/Catalog Number: 14-5011L The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
- 2024-12-25FDA-DeviceKirwan Surgical Products, LLCClass IKirwan Surgical Products LLC Disposable 25-gauge bipolar pencil, straight. Model/Catalog Number: 14-5011 The Disposable Microsurgical Bipolar Pencil are designed for use in eye surgery, and other soft tissue procedures. For instance, the pencils are designed for cauterization of the retinal area during surgical procedures. The current line of pencils all have the same molded handle. The pencils' tip outer diameter is 25 gauge.
Potential for the outer stainless-steel tube to break or detach.
- 2024-12-18FDA-DeviceNoah MedicalClass IIGalaxy bronchoscope (Product GALB-001) used with Instructions for use, Part: 10000752, part of the Noah Medical Galaxy System (GAL-001)
Not all biopsy tools meant to fit bronchoscope channel (2.1mm inner diameter/length greater or equal to 1100mm) will pass through if scope channel positioned in sharp bend (approx. greater than 100-deg.), excessive force may result in tissue injury/pneumothorax; instruments with long rigid distal sections may exhibit insertion resistance when bronchoscope bending section significantly articulated.
- 2024-12-18FDA-DeviceHermes Medical Solutions ABClass IIHybrid Recon software on Hermes Medical Imaging Suite version 6.1 Hybrid Recon software versions affected are 4.0.0x and 5.0.0
Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study.
- 2024-12-18FDA-DeviceDiasorin Inc.Class IIDiaSorin LIAISON Q.S.E.T. Device Plus, REF 319060; intended for use in the preparation of human stool specimens
Customer complaints were received alleging leaking devices and insufficient buffer volume. Diasorin's internal investigation determined that 0.14% of the LIAISON Q.S.E.T. Device Plus in the lots identified have loose clear caps, which may allow buffer to leak from the tube. A stool sample prepared with less than the correct amount of buffer may result an incorrectly high patient result.
- 2024-12-18FDA-DevicePHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Class IIPhilips Allura Xper Systems (R7.6-R8.1). Labeled as the following: 1. Allura Xper FD10, Model Number: 722010. 2. Allura Xper FD10/10, Model Number: 722011. 3. Allura Xper FD20, Model Number: 722012. 4. Allura Xper FD20 Biplane, Model Number: 722013. 5. Allura Xper FD10 OR Table, Model Number: 722022. 6. Allura Xper FD20 OR Table, Model Number: 722023. 7. Allura Xper FD20 Biplane OR Table, Model Number: 722025.
Potential safety issues in Allura Xper R7.6- R8.1 systems where the Lifetime Extension (LTE) kit is not installed in a timely manner. Root cause is under investigation, but the recall reason is listed as for systems over 10 years of life with LTE kits are experiencing propeller motor clamping bolt and x-ray tube locking bolt breaks. If this happens, it may cause erratic C-arm movements and potential collision with patients or bystanders, or detection of a false collisions and subsequent blocked movement of the c-arm. Movement may also cause image noise and black area on the x-ray image, and possible need to expose the patient to additional unnecessary radiation during re-examination.
- 2024-12-18FDA-DeviceCardinal Health 200, LLCClass IICardinal Health SMS Elastic Cuff Thumbhook Level 3 Isolation Gown, Full-Back, Product Codes AT6588-2XL, AT6588-BD, AT6588-XL, AT6688-2XL, AT6688-BD, and AT6688-XL
Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level 3 protection requirements.
- 2024-12-18FDA-DeviceCardinal Health 200, LLCClass IIConvertor Isolation Gown, AAMI Level 3 protection, Product Codes AT6100, AT6100-XL, AT6200, and AT6200-XL
Isolation Gowns were manufactured with raw materials that may fail the hydrostatic pressure specification and therefore would not comply with AAMI (The Association for the Advancement of Medical Instrumentation) Level 3 protection requirements.
- 2024-12-18FDA-DeviceBeckman Coulter Inc.Class IIAccess Intact PTH assay on the DxI 9000 Access Immunoassay Analyzer, Catalog number A16972, for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma.
The reason for the recall is that EDTA and lithium heparin plasma samples that are stored in the DxI 9000 Access Immunoassay Analyzer sample wheel for a prolonged period before performing the intact PTH (parathyroid hormone) test may produce falsely decreased intact parathyroid hormone results. This could lead to a missed diagnosis, a delay in treatment, or inappropriate clinical management, depending upon the patient s clinical condition.
- 2024-12-18FDA-DeviceCareFusion 303, Inc.Class IIBD Pyxis CII Safe ES, REF: 1116-00
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
- 2024-12-18FDA-DeviceCareFusion 303, Inc.Class IIBD Pyxis MedStation ES 7 Drawer Auxiliary Tower, REF: 324
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
- 2024-12-18FDA-DeviceCareFusion 303, Inc.Class IIBD Pyxis MedStation ES, REF: 323
Due to a software issue, automated dispensing cabinet devices may open the wrong cubie pocket or position, which can lead to: 1) Unintended and incorrect medication removal, or 2) Potential for a different medication (medication pulled in error) to be refilled/loaded, which may lead to may lead to low inventory or a stockout.
- 2024-12-18FDA-DeviceHologic, IncClass IIHologic Sizer Set- sterilization tray and 6 individual sizers, intended as a Surgical instrument guide Model/Catalog Number: F0101
IFU was not provided with the Sizer Set sterilization tray Revised Sterilization Instructions for Sizer Set
- 2024-12-18FDA-DeviceMicro-X Ltd.Class IIMICRO-X Rover Mobile X-ray System, # MXU-RV19
During internal testing it was found that if the system is Ready to expose and the hand switch is then dropped into the stowage unit such that it lands on the prep/expose switch, the prep switch can be activated for a very short time such that the orange ring of death is displayed in the tube cooling area of the screen. The system then becomes unresponsive to kV/mAs changes, prep/expose presses. The only way to proceed is to suspend the patient, Exit Nexus DR and relogin and continue the exam.