- 2024-11-06FDA-DeviceBeckman Coulter, Inc.Class IIAccess Thyroglobulin Antibody II, REF A32898, containing Instructions for Use, IFU REF B14994, B63978, and A38168. The Access Thyroglobulin Antibody II (TgAb) assay is used for the quantitative determination of thyroglobulin antibody levels in human serum and plasma using the Access Immunoassay Systems.
IFUs for the device incorrectly states the biotin interference was remediated. The issue could result in potentially false-low results.
- 2024-11-06FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Electrosurgical Generator, Model Number ESG-410 (SEP) Version WA91327U
There was an increase in complaints indicating an occurrence of system error E1226 "Could not write to Audio Buffer" during procedures using the ESG-410 (SEP).
- 2024-11-06FDA-DeviceWestmed, Inc.Class II1) Flex Extenders, 5.2" x 15mm OD x 15mm ID REF S6411P 2) Flex Extender, 5.2" x 15mm OD x 22/15mm ID REF 7 5132 The Westmed Flex Extenders are non invasive, single use/disposable, large bore, corrugated plastic tubes that provide a flexible, low resistance lightweight connection between breathing circuits and patient interface (tracheal tube, tracheostomy tube, or mask).
Due to connector within the circuit flex extender not meeting specifications resulting in a low connection force and as a result, the connector may leak or disconnect and may cause life threating conditions or death.
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IIDuraform Dural Graft Implant, Size 3x3 , Model Number 80-1477US Lot CT007694
Out of specification endotoxin levels.
- 2024-11-06FDA-DeviceAmbu Inc.Class IIAmbu¿ VivaSight 2 DLT Endobronchial tube, 35 FR (Catalog Number 412351000), 37 Fr (Catalog Number 412371000), 39 Fr (Catalog Number 412391000), 41 Fr (Catalog Number 412411000)
Ambu has received complaints on Ambu¿ VivaSight 2 DLTs regarding hyper angulation of the distal end. The hyper angulation can lead to an increased risk of complications during intubation and potential airway injury.
- 2024-11-06FDA-DeviceTrackx Technology IncClass IIIProduct Name: Insight Base Flex - 8, Ref 608-3TX; The device consists of a sterile Snap that includes reflective features that facilitate imaging of the (attached) instrument using instrument tracking technology when used with a compatible tracking system. Each TrackX¿ Snap has a unique set of 2 or more reflective markers. By having unique configurations of reflective markers, it is possible to identify which TrackX Snap is in the field of view of the camera(s). As such, it is also possible to h
The patient label stickers may have the wrong expiration date recorded. The contract manufacturer entered the manufacturing date (2024-04-01) as the expiration date. The actual expiration date of the product is 2027-04-01. This is reflected on the carton label and sterile pouch label correctly.
- 2024-11-06FDA-DevicePrinceton Biomeditech CorpClass IIOsom Ultra hCG Combo Test- IVD immunoassay for the qualitative detection of human chorionic gonadotropin (hCG) in serum or urine for the early confirmation of pregnancy. Model/Catalog Number: 1004
This recall was initiated due to customer complaints of false negative of hCG test results when running patient urine samples. Use of the above listed product lot may result in false negative results. An end user reported that the OSOM Ultra hCG Combo Test, lot 563L13, produced false negative test results from 2 patient urine samples as confirmed by quantitative hCG and ultrasound. During further root cause testing and investigation, an additional 6 devices were found to produce false negative test results from hCG positive samples.
- 2024-11-06FDA-DeviceDRG International, Inc.Class III17-OH Progesterone ELISA 96 well kit- An enzyme immunoassay for the quantitative in vitro diagnostic measurement of 17-OH Progesterone (17-OHP) in serum or plasma Model/Catalog Number: EIA-1292
The High and Low Control acceptance ranges in the Certificate of Analysis (CofA) did not match the acceptance range information on the High and Low Control Component labels. Due to the defect, the device failed to provide results, which may result in a delay in obtaining test results and a remote chance of an adverse health risk.
- 2024-11-06FDA-DeviceCovidien LPClass IIChameleon PTA Balloon Catheter with Injection Port 5mm x 40mm x 75cm (CFN CH05-40-75US), 6mm x 40mm x 75cm (CFN CH06-40-75US), 7mm x 40mm x 75cm (CFN CH07-40-75US), 8mm x 40mm x 75cm (CFN CH08-40-75US), 9mm x 40mm x 75cm (CFN CH09-40-75US), 10mm x 40mm x 75cm (CFN CH10-40-75US), 12mm x 40mm x 75cm (CFN CH12-40-75US)
Specific lots of the Chameleon PTA Balloon Catheter failed to meet in-house balloon pressure testing.
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 3-1/2X6 Model/Catalog Number:801457. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 3X6 Model/Catalog Number: 801456. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 2X6 Model/Catalog Number: 801455. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 1X6-200 Model/Catalog Number: 801453. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 3/4X6 Model/Catalog Number: 801452. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 1/2X6 Model/Catalog Number: 801451. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 1/4X6 Model/Catalog Number: 801450. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG STRP 1/8X6 Model/Catalog Number: 801449. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 3X3 Model/Catalog Number: 801409. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1X3 Model/Catalog Number: 801408. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1/2X3 Model/Catalog Number: 801407. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1/2X2 Model/Catalog Number: 801406. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1/2X1-1/2 Model/Catalog Number: 801404. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1X1 Model/Catalog Number: 801403. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1/2X1 Model/Catalog Number: 801402. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1/2X1/2 Model/Catalog Number: 801400. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1/4X1/4 Model/Catalog Number: 801399. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1/4X3 Model/Catalog Number: 801398. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods
- 2024-11-06FDA-DeviceIntegra LifeSciences Corp.Class IISURG PAT XRAY 1/4X1-1/2 Model/Catalog Number: 801397. The surgical patties and surgical strips are indicated for the use in protection of tissue, including brain and other tissues of the central nervous system, during surgery.
Identified higher-than-expected levels of endotoxin within the raw material used to produce Codman Surgical Patties and Strips that may have resulted in out-of-specification levels of endotoxin in finished goods