Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2023-11-08FDA-DeviceCosmedent, Inc.Class IICosmedent FlexiCup Composite Finishing & Polishing System, Six Pack, Autoclavable Aluminum Oxide Polishing Instrument, Medium (Blue), REF #403-101, Rx.
Reports were received the elastomer cup material is crumbling.
- 2023-11-08FDA-DeviceMaquet Cardiovascular, LLCClass IIVasoview 6 Pro (VV6) - Endoscopic Vessel Harvesting System, Model VH-2400 printed on outer label, inner label and IFU, outer box artwork for Vasoview Hemopro 2 (HP2).
Device packaged inside an incorrect outer box; displaying artwork & illustrations for a different device.
- 2023-11-08FDA-DeviceAbbottClass IIAbbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator
Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.
- 2023-11-08FDA-DeviceNova Biomedical CorporationClass IINova Stat Profile Prime Plus POC Analyzer (CE)-Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 63368 Stat Profile Prime Plus POC Analyzer (2) PN 63369 Stat Profile Prime Plus POC Analyzer, Remanufactured P/N: 57400, 63368
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
- 2023-11-08FDA-DeviceNova Biomedical CorporationClass IINova Biomedical Prime Plus Analyzer -Intended to measure certain gases in blood, serum, plasma or pH of blood, serum, and plasma. (1) PN 57400 Stat Profile Prime Plus Analyzer (2) PN 59508 Stat Profile Prime Plus Analyzer, Remanufactured
Potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new BUN/Creatinine Sensor Cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional Quality Control practices.
- 2023-11-08FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayCare 5B and 6A including service packs- An oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care Product name (build number): (1)RayCare 5B SP1 (5.1.1.60246); (2)RayCare 5B SP3 (5.1.3.60023); (3) RayCare 6A (6.0.0.60553)
An issue was found in RayCare 5B, 6A, including service packs, where values will not be visible for documents in a read-only state when opened outside the Documents workspace when using document fields configured with single- or multi-selectable values.
- 2023-11-08FDA-DeviceSmiths Medical ASD Inc.Class IISmiths Medical Portex Bivona Pediatric Tracheostomy Tube, Cuffless, Model 60P045, Sterile.
The unit box is labeled with an incorrect expiration date (the manufacturing date of 3/17/2021) but the individual packaging contains the correct expiration date (3/17/2026).
- 2023-11-08FDA-DeviceCoapt LLCClass IICoapt ControlSeal Electrode, cutaneous EMG Electrode
Patient may experience minor burn or blistering if device is exposed to direct sunlight and/or warm temperatures for over 12 hours or battery is damaged.
- 2023-11-08FDA-DeviceCoapt LLCClass IICoapt Dome Electrode, Gen2 system, cutaneous electrode
Patient may experience minor burn or blistering if device is exposed to direct sunlight and/or warm temperatures for over 12 hours or battery is damaged.
- 2023-11-08FDA-DeviceRandox Laboratories Ltd.Class IIRX Monaco-IVD Clinical Chemistry Analyser Catalog Number: RX5000
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceRandox Laboratories Ltd.Class IIRX Modena-IVD Clinical Chemistry Analyser Catalog Number: RX9000
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceRandox Laboratories Ltd.Class IIRX Imola-IVD Clinical Chemistry Analyzer Catalog Number: RX4900. RX4900R (refurbished)
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceRandox Laboratories Ltd.Class IIRX Daytona Plus (without ISE)- IVD Clinical Chemistry Analyser (1) RX4041, (2) RX4041R (refurbished)
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceRandox Laboratories Ltd.Class IIRX Daytona Plus (with ISE)-IVD Clinical Chemistry Analyser Catalog Number: RX4040
RX Parameter rerun settings (Sample Volume Low/High), and Dilutions used for Auto reruns, not meeting sample re-run conditions when samples are above or below assay range and may result i issues with the re-run feature conditions when incorrectly. Potential to report incorrectly.
- 2023-11-08FDA-DeviceQuest Medical, Inc.Class IIQ2 Multiport Extension set, REF: 9520, Rx Only Sterile, single use non-pyrogenic intravenous fluid administration sets with a multiport IV manifold and integrated back-check valves, pre-attached needleless injection sites, drip chamber and roller clamps.
For administration of intravenous fluids to a patient s vascular system utilizing needle-free components and I.V. manifold for multiple simultaneous intravenous therapy via gravity, syringe, or infusion pump. Use of a needle-free system may aid in the prevention of needle-stick injuries.
- 2023-11-08FDA-DeviceSiemens Medical Solutions USA, IncClass II***Updated 12/4/23*** (1) Cios Flow (VA30) mobile fluoroscopic C-arm X-ray System. Designed for the surgical environment and provides comprehensive image acquisition modes. Material Number: 11108110 (2)Cios Spin- a mobile x-ray system designed to provide x-ray imaging of the anatomical structures of patients during clinical applications Material Number: 10308194
Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure
- 2023-11-08FDA-DeviceSiemens Medical Solutions USA, IncClass IISiemens Cios Alpha (VA30) mobile fluoroscopic C-arm X-ray System, designed for the surgical environment. 11105200
Software issue for Cios Alpha, Cios Flow, and Cios Spin VA30 systems, fulfillment of new requirements from DIN and IEC standards for Cios VA30 systems required may result in patient misdiagnosis/repeat exposure
- 2023-11-08FDA-DeviceMICROVENTION INC.Class IIAZUR CX 35 Peripheral Coil System Detachable 13mm x 24 CM, REF: 45-751324
Peripheral coil system detachable has a potential of unsealed pouch packaging.
- 2023-11-08FDA-DeviceAsensus Surgical IncClass ISenhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system
Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.
- 2023-11-08FDA-DeviceCovidienClass IIMcGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-084-000, Sterile EO, Rx Only
Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
- 2023-11-08FDA-DeviceCovidienClass IIMcGrath Mac 2, Disposable Laryngoscope Blade, REF: 350-017-000, Sterile EO, Rx Only
Laryngoscope blades may contain an ineffective application of anti-fog agent which may result in condensation accumulating causing the screen to be blurred or obscured.
- 2023-11-08FDA-DeviceCovidienClass IStolen McGRATH MAC Video Laryngoscopes, REF: 301-000-000
Defective video laryngoscopes that did not pass quality tests established for the product, that were designated for destruction, were stolen, and subsequently distributed for sale. Defective device use may result in failed laryngoscopy and/or intubation.
- 2023-11-01FDA-DeviceCarl Zeiss Suzhou Co., Ltd.Class IIOPMI LUMERA 300, REF 6137
Surgical microscope for use in ophthalmology may be missing a screw from the suspension arm, which may cause the suspended components to fall from the suspension arm, which may injure any person under the suspended components.
- 2023-11-01FDA-DeviceAmerican Contract Systems, Inc.Class IICath Cardiac Pack, REF IHCC03V
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
- 2023-11-01FDA-DeviceAmerican Contract Systems, Inc.Class IIPediatric Urology Wolson Pack, REF BPPU55A
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
- 2023-11-01FDA-DeviceAmerican Contract Systems, Inc.Class IISub Q Recorder Pack, REF BPRC87A
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
- 2023-11-01FDA-DeviceAmerican Contract Systems, Inc.Class IIRobotics Pack, REF SSRO22D
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
- 2023-11-01FDA-DeviceAmerican Contract Systems, Inc.Class IIRobotics Pack, REF SJVO56B
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
- 2023-11-01FDA-DeviceAmerican Contract Systems, Inc.Class IIVasc. Open CVOR SJH, REF SJVO56B
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.
- 2023-11-01FDA-DeviceAmerican Contract Systems, Inc.Class IIPediatric Heart SJH, REF SJPH34K
During an internal investigation, ACS identified that the components, I.V. Cath (part number 4252535-02) and Adhesive Dermabond (part number DHVM12) were inadvertently subjected to the ACS sterilization process and exposed to EO and higher temperatures than approved by the component manufacturer. As a result, the components may (1) have a loss/lack of functionality, (2) loss of drug efficacy, if applicable, and (3) have higher than specified EO residuals.