Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2023-10-11FDA-DeviceZimmer Surgical IncClass IIZimmer Dermatome Blades- 10-pack boxes of blades. Intended for use with the dermatome handpiece which is a skin grafting instrument that is intended to provide variable graft thickness and width capabilities Part number: 00880000010
Skin grafts thin and non-uniform when using the affected blades. The issue would be identified at the time of use and may result in the need for additional harvests to adequately cover the area
- 2023-10-11FDA-DeviceBlue Belt Technologies, IncClass IIReal Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299
CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis
- 2023-10-11FDA-DeviceAB Sciex, LLCClass IICitrine Triple Quad Mass Spectrometer, Part Number 5063684
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- 2023-10-11FDA-DeviceAB Sciex, LLCClass IICitrine QTRAP Mass Spectrometer, Part Number 5063685
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- 2023-10-11FDA-DeviceAB Sciex, LLCClass II4500MD Triple Quad Mass Spectrometer, Part Number 5032522
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- 2023-10-11FDA-DeviceAB Sciex, LLCClass IIAPI3200MD Mass Spectrometer, Part Number 4466230
In rare occasions, the combination of a leak of a highly flammable solvent at the ion source probe, and an electrical discharge may result in a flame on top of the probe and possible ejection of the electrode assembly from the probe. There is a risk to the patient/user, customer property, or environment because there is a potential for a health risk and property damage should the flame be left unattended. The failure causes a delay in patient results.
- 2023-10-11FDA-DeviceGetinge Usa Sales IncClass IIMeera Mobile Operating Table-to support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Model Number: 7200.01F2
Under certain conditions, an issue might prevent the device from performing as intended. A specific sequence of commands on the control unit activates the traction drive and triggers an unintended driving (autodrive) of max. 7 sec. may result in the following reasonably foreseeable injury or damage to the health of patients and/or users
- 2023-10-11FDA-DeviceBard Peripheral Vascular IncClass IIBard Marquee Disposable Core Biopsy Instrument Kit, REF: MQK2016
Diameter mismatch between coaxial and biopsy needle (cutting cannula) in disposable core biopsy instrument kits, where the coaxial cannula inner diameter prohibits the biopsy needle (cutting cannula) from properly fitting into the coaxial cannula and accessing target tissue, which may lead to repeated procedure, and procedural complications, such as pain/discomfort, pneumothorax, and bleeding.
- 2023-10-11FDA-DeviceBecton Dickinson Infusion Therapy Systems Inc.Class IIREF: 386862, BD Cathena Safety IV catheter with BD Multiguard Technology, 20 GA 1.00 IN, 1.1 x 25 mm 64 mL/min , Rx Only, Sterile EO
There is the potential that the safety shield may not properly engage on IV catheters. The needle can pull through the safety shield and disconnect from the needle safety shield leaving the contaminated needle point exposed.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 5F 100CM MULTI PURPOSE, Catalog number M003101630, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 5F PRE-SHAPED 40 100CM, Catalog number M003101620, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 8F 100CM MULTI PURPOSE, Catalog number M003101480, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 7F 100CM MULTI PURPOSE, Catalog number M003101470, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 6F 100CM MULTI PURPOSE, Catalog number M003101460, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 8F PRE-SHAPED 40 100CM, Catalog number M003101440, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 7F PRE-SHAPED 40 100CM, Catalog number M003101430, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 6F PRE-SHAPED 40 100CM, Catalog number M003101420, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 5F 90CM MULTI PURPOSE, Catalog number M003100630, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 5F PRE-SHAPED 40 90CM, Catalog number M003100620, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 8F 90CM MULTI PURPOSE, Catalog number H965100480, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 7F 90CM MULTI PURPOSE, Catalog number H965100470, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 6F 90CM MULTI PURPOSE, Catalog number H965100460, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 8F PRE-SHAPED 40 90CM, Catalog number H965100440, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 7F PRE-SHAPED 40 90CM, Catalog number H965100430, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceBoston Scientific CorporationClass IIGuider Softip Guide Catheter 6F PRE-SHAPED 40 90CM, Catalog number H965100420, cardiac guide catheter
Stryker Neurovascular has observed that certain lots of Guider Softip XF Guide Catheters appear to have the incorrect tip curve shape. The impacted products were distributed with a tip curve shape which differs from the labeled tip curve shape.
- 2023-10-11FDA-DeviceOrtho-Clinical Diagnostics, Inc.Class IIProboscis/Piston Assembly, Part Number J55375, for VITROS 5,1 FS Chemistry System (Product Code 6801375), VITROS 5,1 Chemistry System (Refurbished) (Product Code 6801890), VITROS 3600 Immunodiagnostic System (Product Code 6802783), VITROS 3600 Immunodiagnostic System (Refurbished) (Product Code 6802914), VITROS 4600 Chemistry System (Product Code 6802445), VITROS 4600 Chemistry System (Refurbished) (Product Code 6900440), VITROS 5600 Integrated System (Product Code 6802413), VITROS 5600 Integrat
Ortho Clinical Diagnostics (QuidelOrtho) confirmed certain Proboscis/Piston Assemblies manufactured between 07-Apr-2022 (Date code: 040722) and 19-Dec-2022 (Date code: 121922) may not perform as intended due to an out-of-specification barrel component. This may cause VITROS VersaTips and/or VITROS MicroTips to be improperly seated on the proboscis, which may result in falsely elevated Gentamicin and Valproic Acid assay results as well as an increased frequency of condition codes resulting in delayed patient results. The issue was identified by internal testing when reported instruments failed to pass release testing and were traced to the molded Proboscis part J56639 with date codes from 040722 onward.
- 2023-10-11FDA-DeviceARROW INTERNATIONAL Inc.Class IArrow ErgoPack Pressure Injectable Arrowg+ard Blue Advance Three-Lumen PICC pre-loaded with Arrow VPS TipTracker Stylet, REF CDC-45563-TTS; catheter, intravascular, therapeutic, long-term greater than 30 days
This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits.
- 2023-10-11FDA-DeviceARROW INTERNATIONAL Inc.Class IArrow Pressure Injectable Arrowg+ard Blue Plus¿ Four-Lumen CVC, REF CDC-45854-P1A; catheter, intravascular, therapeutic, long-term greater than 30 days
This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits.
- 2023-10-11FDA-DeviceARROW INTERNATIONAL Inc.Class ICheck Arrow ErgoPack Complete catheter kit labels
This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits.
- 2023-10-11FDA-DeviceARROW INTERNATIONAL Inc.Class ICheck Arrow ErgoPack PICC catheter kits for labeling errors
This recall has been initiated due to reports that the product code and product name were incorrectly listed on the lidstock of affected kits.