Medical device recalls
38684 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2023-08-23FDA-DeviceCovidien LPClass IMAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 13 cm, Curved Extensions, IC Tray REF:8888101001HP
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus
- 2023-08-16FDA-DevicePhilips North AmericaClass IIMR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.
- 2023-08-16FDA-DevicePhilips North AmericaClass IIIngenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106
During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.
- 2023-08-16FDA-DevicePhilips North AmericaClass IIIngenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X)
During installation of the SmokeDetector Interlock, a connecting cable, between the SmokeDetector Interlock and the host PC, may not have been properly installed. This connection is used for service diagnostics and does not impact the functionality of the system and/or the SmokeDetector Interlock system.
- 2023-08-16FDA-DevicePhilips North AmericaClass IIMR 7700 Magnetic Resonance (MR) system, Product Numbers 782120 and 782130 (Upgrade to MR 7700)
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
- 2023-08-16FDA-DevicePhilips North AmericaClass IIIngenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
- 2023-08-16FDA-DevicePhilips North AmericaClass IIIngenia Elition X Magnetic Resonance (MR) system, Product Numbers 781358, 782107, and 782118 (SmartPath to Ingenia Elition X)
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
- 2023-08-16FDA-DeviceStryker Sustainability SolutionsClass IIHoverMatt, 39" W x 78" L, REF: HM39SPU-B
39-inche-wide inflatable air transfer mattress packages contain 34-Inch-wide mattresses, so there may be a decreased surface area available to distribute the patient's weight across, which may increase the effort exerted by staff for a lateral patient transfer, and there may be increased friction and shearing forces when the patient is moved resulting in skin tear injury.
- 2023-08-16FDA-DeviceCytrellis Biosystems, Inc.Class IIellacor System with Micro-Coring Technology-Indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. Model #: Gen 4
Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring
- 2023-08-16FDA-DeviceCanon Medical System, USA, INC.Class IIINFX-8000C XGCP-930AA TABLE SIDE CONSOLE, INFX-8000C XGCP-930BA TABLE SIDE CONSOLE
The C-arm, used to set the position and angle of the x-ray exposure field, may not function due to the tableside console PWB, which needs replacement. There may be a delay in the procedure, but procedures can continue using the operators console, the switches on the back of the column, switches on the side of the flat panel detector, and the emergency stop will continue to function normally.
- 2023-08-16FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IITotal Hip Kit, REF DYNJ903275S; surgical convenience kit
The prep solution included in the Total Hip Kit expires prior to the expiration date of the Kit.
- 2023-08-16FDA-DeviceMedtronic NeuromodulationClass IIMedtronic InterStim Ground Pad REF 041826
Lot of InterStim Self-Adhesive Ground Pad is being recalled due to the product being labeled with the incorrect expiration date.
- 2023-08-16FDA-DeviceScholly Fiberoptic GmbhClass IIUroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift¿ system. Instrument for insertion of a cystoscope shaft into the bladder. Model: UL-VO
There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.
- 2023-08-16FDA-DeviceMedtronic Sofamor Danek USA, IncClass IICD HORIZON Spinal System Solera Awl Tip Tap, REF NAV2006K
Incorrect Laser Marking - Solera Awl Tip Tap
- 2023-08-16FDA-DeviceRandox Laboratories Ltd.Class IIRandox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)
Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.
- 2023-08-16FDA-DeviceSiemens Medical Solutions USA, IncClass IIARTIS pheno-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 10849000
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
- 2023-08-16FDA-DeviceSiemens Medical Solutions USA, IncClass IIARTIS icono floor-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327700
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
- 2023-08-16FDA-DeviceSiemens Medical Solutions USA, IncClass IIARTIS icono biplane -Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327600
Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
- 2023-08-16FDA-DeviceORTHO-CLINICAL DIAGNOSTICSClass IIVITROS Immunodiagnostic Products Troponin I ES Calibrators
Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.
- 2023-08-16FDA-DeviceORTHO-CLINICAL DIAGNOSTICSClass IIVITROS Immunodiagnostic Products Troponin I ES Reagent Pack
Reagent packs could potentially contain incorrect wells, which will generate incorrect, and potentially false negative, results.
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 25 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-25/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 24 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-24/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 23 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-23/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 22 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-22/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 21 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-121/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 20 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-120/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 19-Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-119/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 18 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-118/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 17 - indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-117/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue
- 2023-08-16FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 16 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-116/00
Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue