Medical device recalls
38685 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2022-10-19FDA-DeviceHobbs Medical, Inc.Class IIHobbs Medical Polypectomy Snare Catalog Number: 4564
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.)
- 2022-10-19FDA-DeviceHobbs Medical, Inc.Class IIHobbs Medical Polypectomy Snare Catalog Number: 4563
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.)
- 2022-10-19FDA-DeviceHobbs Medical, Inc.Class IIHobbs Medical Cytology Brush Catalog Number: 4206
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.)
- 2022-10-19FDA-DeviceHobbs Medical, Inc.Class IIHobbs Medical Cytology Brush Catalog Number: 4204
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.)
- 2022-10-19FDA-DeviceHobbs Medical, Inc.Class IIHobbs Medical Flex-Ez Guide Wire Disposable For endoscopy procedures Ref: 3412
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.)
- 2022-10-19FDA-DeviceHobbs Medical, Inc.Class IIHobbs Medical Spray Catheter: The Mistifier: 2.6mm 165cm Disposable For endoscopy procedures Ref: 2191
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.)
- 2022-10-19FDA-DeviceHobbs Medical, Inc.Class IIHobbs Medical Spray Catheter: The Mistifier: 2.6mm 260cm Disposable For endoscopy procedures Ref: 2190
Device label display an incorrect expiration date, extends the shelf life of the product. Affected product distributed in the United States from 3/25/2018 to 9/28/2020, expiration dates between 2022-11-01 and 2025-09-01 (Devices have a lot number ending in R, or are kits containing at least 1 device with a lot number ending in R.)
- 2022-10-19FDA-DeviceJiangsu Well Biotech Co.,Ltd.Class IStop using SDI Labs COVID-19 rapid tests
Distribution of COVID-19 Ag Rapid Test kits in the U.S. without an Emergency Use Authorization, or a Pre-Market Approval or Clearance.
- 2022-10-12FDA-DeviceHologic, IncClass IIThe BioZorb LP Marker is an implantable radiopaque marker used to facilitate visualization of a soft tissue site. The BioZorb LP Marker is comprised of a bioabsorbable PLA (polylactic acid) component which resorbs completely in 1 year or more and a permanent component (titanium). The BioZorb LP Marker is provided sterile for single use and is implantable.
Hologic BioZorb LP Marker, (Product number F0221), lot number 22A0RL, contained removable labels with an expiration date that indicated the product had already expired. The expiration date on the removable label is actually manufacturing date, and it's different from the one on the shelf box label, which is the correct expiration date.
- 2022-10-12FDA-DevicePhilips Respironics, Inc.Class ITherapy Mask 3100 NC/SP
No contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.
- 2022-10-12FDA-DevicePhilips Respironics, Inc.Class IWisp and Wisp Youth Nasal Mask
No contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.
- 2022-10-12FDA-DevicePhilips Respironics, Inc.Class IDreamWisp Nasal Mask with Over the Nose Cushion
No contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.
- 2022-10-12FDA-DevicePhilips Respironics, Inc.Class IPhilips Respironics DreamWear Full Face Mask
No contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.
- 2022-10-12FDA-DevicePhilips Respironics, Inc.Class IPhilips Respironics Amara View Minimal Contact Full-Face Mask
No contraindication/warning regarding magnetic components in CPAP masks for patients with implantable devices or metallic splinters in the eyes.
- 2022-10-12FDA-DeviceB Braun Medical IncClass IIIntrocan Safety FEP 14G, 2.2x32MM-US, Product Code 4251890-02. A passive anti-needle device to provide venous or arterial access.
Potential for leakage at the catheter hub.
- 2022-10-12FDA-DeviceSynthes (USA) Products LLCClass IIHammertoe Continuous Compression Implant (CCI) Kit Part Number: 46.239.001
Drill Pin (drill bit) is not compatible with the Drill Template included in the kit due to a packaging error, result in a surgical delay and may cause soft tissue damage if a larger incision is required to accommodate an alternative form of fixation. Additionally, there is a potential for adverse tissue reaction if the surgeon attempts to use the guide regardless of tight fit and plastic debris is created and ends up in the wound
- 2022-10-12FDA-DeviceZimmer, Inc.Class IIG7 Acetabular System, Acetabular Shell, Item numbers Item Numbers: 010000663 (52 MM E), 010000667 (60 MM G). used in hip arthroplasty
The outer sterile package cavity has a corner wall thickness that is below the specification. The thin corner wall could potentially crack during transit. Potential risks include non-clinically or clinically significant extension of surgery, or infection leading to surgical intervention.
- 2022-10-12FDA-DeviceWright Medical Technology, Inc.Class IIstryker EVOLVE, Proline Stem, CoCr, Size: +4, OD: 9.5mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set configuration designed to address radial head fractures. It combines parts of two systems: EVOLVE Proline and EVOLVE TRIAD.
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.
- 2022-10-12FDA-DeviceWright Medical Technology, Inc.Class IIstryker EVOLVE, Proline Stem, CoCr, Size: +2, O: 705mm, REF 496S275, Wright s EVOLVE Proline Plus Radial Head and Repair System is a set configuration designed to address radial head fractures. It combines parts of two systems: EVOLVE Proline and EVOLVE TRIAD.
The outer label does not match the device inside the package. The affected lot of EVOLVE Proline Stem 7.5mm +2 contains a size 9.5mm +4 stem, while the affected lot of EVOLVE Proline Stem 9.5mm +4 contains a size 7.5mm +2 stem.
- 2022-10-12FDA-DevicePhilips North America LlcClass IIMR Patient Care Portal 5000 (Desktop Unit), Model Number 453564792561 This is a component of product number 866162.
Device may intermittently fail to produce audio. Turning the unit off and on does not resolve the issue; disconnecting and reconnecting to AC power does not reliably resolve the issue.
- 2022-10-12FDA-DeviceIN MY BATHROOM LLCClass IIThe product is a handheld ultraviolet-C germicidal wand and contains two, narrow-diameter, tubular lamps with a peak emission wavelength in the 254 nm range.
Excessive ultraviolet-C radiation
- 2022-10-12FDA-DeviceFresenius Kabi USA, LLCClass IStop using Ivenix Infusion System version 5.2.0
The display screen may become frozen and unresponsive to user input, triggering an alarm that is evidenced by a buzzer sound and flashing red LEDs as a secondary notification. However, as the screen goes blank it does not indicate an alarm condition or respond. This issue leads to temporary cessation of flow of a medication or fluid, requiring immediate intervention by the clinician.
- 2022-10-05FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMEDLINE CLEARPRO TRACH T-PIECE CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14; b. TRACH T-PIECE, 14 Fr, Item Number/REF: DYNCPTP14T
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
- 2022-10-05FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMEDLINE CLEARPRO TRACH DOUBLE SWIVEL CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. TRACH DOUBLE SWIVEL, 12 Fr, Item Number/REF: DYNCPDS12T b. TRACH DOUBLE SWIVEL, 14 Fr, Item Number/REF: DYNCPDS14T
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
- 2022-10-05FDA-DeviceMEDLINE INDUSTRIES, LP - NorthfieldClass IIMEDLINE CLEARPRO DOUBLE SWIVEL CLOSED SUCTION CATHETER, flat peel pouch EA, 10 EA per Inner Carton (IC), 20 EA/2 Inner cartons per Case (CS). Labeled as follows: a. ET LENGTH DOUBLE SWIVEL, 12 Fr, Item Number/REF: DYNCPDS12; b. ET LENGTH DOUBLE SWIVEL, 14 Fr, Item Number/REF: DYNCPDS14; c. ET LENGTH DOUBLE SWIVEL, 16 Fr, Item Number/REF: DYNCPDS16
Potential for the suction catheter to come apart from the green connector near the thumb valve. The connector is coming apart from the assembled device when pulling the catheter out of the artificial airway in the patient suctioning process.
- 2022-10-05FDA-DeviceBARCO NVClass IIBarco Medical Network Adapter, MNA-240, part of the NexxisOR system, Model Numbers: a) K9303078A, b) K9303097A
installation issues
- 2022-10-05FDA-DeviceSurgical Innovations LtdClass IIProduct/Part: YelloPort Plus 5x70 Cannula + Luer/YC0507001. For use in laparoscopic procedures.
Laparoscopic devices have a sterile symbol included on labels but devices are non-sterile.
- 2022-10-05FDA-DeviceSurgical Innovations LtdClass IIProduct/Part: YelloPort Elite 12x75 Pencil Point Trocar/ET1207503; YelloPort Elite Hasson Adaptor 10mm/EA10NH. For use in laparoscopic procedures.
Laparoscopic devices have a sterile symbol included on labels but devices are non-sterile.
- 2022-10-05FDA-DeviceLeica Microsystems, Inc.Class IILeica Surgical Operating Microscopes, PROvido Part Numbers: a. 10448976, b. 10448950
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.
- 2022-10-05FDA-DeviceLeica Microsystems, Inc.Class IILeica Surgical Operating Microscopes, M530 OHX, Part Number 10448737
During internal testing, Leica Microsystems has become aware of a component change on the photodiodes inside the M530 Optics Carrier which will result in an inaccurate adjustment of the illumination limits by the software of the surgical microscope when "BrightCare Plus" with Luxmeter is used.