Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2020-12-09FDA-DeviceCME America, LLCClass IICME America T-Syringe Pump - T34L PCA - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- 2020-12-09FDA-DeviceCME America, LLCClass IICME America T-Syringe Pump - NeoThrive - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- 2020-12-09FDA-DeviceCME America, LLCClass IICME America T-Syringe Pump -T34L - Product Usage: infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, intra-arterial, epidural, in close proximity to nerves, and into an intra-operative site (soft tissue/body cavity/surgical wound site).
Multiple reasons: 1) Due to wear and tear of pump motor block mechanism - may cause under infusion situations without the pump alarm being generated. 2) Due to cracks, pump hardware may be damaged by fluid used during the cleaning process.
- 2020-12-09FDA-DeviceROi CPS LLCClass IIRegard - Product Usage: Obstetric Gynecologic specialized manual instrument. Labor and delivery kits.
Labeling error, incorrect expiration date.
- 2020-12-09FDA-DeviceChromsystems Instruments & Chemicals GmbHClass IIMassChrom Cortisol, Cortisone Saliva Control Level II, Order No. 0350 - Product Usage: The kit is used as a test for patients to monitor their cortisol and cortisone levels.
Control lots may not meet stability criteria and could result in increased results of patient samples for cortisone.
- 2020-12-09FDA-DeviceChromsystems Instruments & Chemicals GmbHClass IIMassChrom Cortisol, Cortisone Saliva Control Level I, Order No. 0349 - Product Usage: The kit is used as a test for patients to monitor their cortisol and cortisone levels.
Control lots may not meet stability criteria and could result in increased results of patient samples for cortisone.
- 2020-12-09FDA-DeviceChromsystems Instruments & Chemicals GmbHClass II6PLUS1 Multilevel Saliva Calibrator Set Cortisol Cortisone in Saliva, Order No. 73039 - Product Usage: The kit is used as a test for patients to monitor their cortisol and cortisone levels.
Calibrator lots may not meet stability criteria and could result in increased results of patient samples for cortisone.
- 2020-12-09FDA-DeviceK2M, IncClass IIYukon OCT Spinal System Yukon Anti-Torque Alignment Tube - Product Usage: The Anti-Torque Alignment Tube (the Alignment Tube ) provides counter rotation to reduce the torque applied to the anatomy when the size 15 Torque limiting shaft (the shaft is used to tighten the Yukon Set Screw to the specified torque.
Shafts do not pass through the inner cannula of the Anti-Torque Alignment Tube from Lot KTGG due to a manufacturing nonconformance. This could result in a delay of surgery or result in surgical intervention later on.
- 2020-12-09FDA-DeviceBausch & Lomb Surgical, Inc.Class IIMX60ET (enhanced enVista Toric) enVista¿ One-Piece Hydrophobic Acrylic Toric Model MX60ET - Product Usage: The lens is intended for placement in the capsular bag.
Product may be missing toric axis marks.
- 2020-12-09FDA-DeviceMedtronic Inc.Class IIMiniMed 640G, insulin pump, Ref/Model # MMT-1711K, MMT-1752K, MMT-1752, Rx Only - Product Usage: indicated for the continuous delivery of insulin, at set and variable rates, for the management of diabetes mellitus in persons requiring insulin.
Missing solder battery connection, which could interrupt power of insulin pump.
- 2020-12-09FDA-DeviceMedtronic Inc.Class IIMiniMed 630G with SmartGuard, insulin pump, Ref/Model # MMT-1754K, MMT-1714K, MMT-1755K - Product Usage: intended for continuous delivery of basal insulin (at user selectable rates) and administration of insulin boluses (in user selectable amounts) for the management of diabetes mellitus in persons, fourteen years of age and older, requiring insulin.
Missing solder battery connection, which could interrupt power of insulin pump.
- 2020-12-09FDA-DeviceMedtronic Inc.Class IIMiniMed 670G, insulin pump, Ref/Model # MMT-1760KPK, MMT-1762KCN, MMT-1762WWK - Product Usage: indicated for continuous or periodic monitoring of glucose levels in the fluid under the skin, and detecting possible low and high glucose episodes. When using a sensor and transmitter, the pump displays continuous sensor glucose values and stores this data so that it can be analyzed to track patterns and improve diabetes management. This data can be uploaded to a computer for analysis of historical gl
Missing solder battery connection, which could interrupt power of insulin pump.
- 2020-12-09FDA-DeviceCovidien LlcClass IICapnostream 35 Portable Respiratory Monitor, Product numbers PM35MN02, PM35MN01, PM35MN05, DLPM35MN02 - Product Usage: use during both no motion and motion conditions and for patients who are well or poorly perfused.
The firm has released software update V01.05.02.16 (also known as V1.5.2) in response to customer reports of a false display of the message, "Temperature Exceeds Limits" followed by automatic shutdown of the monitor with no accompanying alarm.
- 2020-12-09FDA-DeviceInvacare CorporationClass IIInvacare PerfectO2 V Oxygen Concentrator- intended for patients with respiratory disorders requiring supplemental oxygen Model Numbers: IRC5PO2V (US), IRC5PO2VC (Canada), All World: IRC5PO2VAW, IRC5PO2VAWL2, IRC5PO2VAWL3, IRC5PO2VAWL4, IRC5PO2VAWL5
Breakdown of the sound abatement washer, and metal on metal wear inside the Pressure Equalization (P.E.) valve assembly in the concentrator may lead to failure that can result in self-extinguishing fires or cap explosions.
- 2020-12-09FDA-DeviceBiomeme, Inc.Class IIThe Biomeme SARS-CoV-2 Real-Time RT-PCR Test is a molecular in vitro diagnostic test that is based on widely used nucleic acid amplification technology. The Biomeme SARS-CoV-2 Real-Time RT-PCR Test contains primers and probes and internal controls used in RT-PCR for the in vitro qualitative detection of SARS-CoV-2 RNA in upper respiratory specimens. The Biomeme SARS-CoV-2 Real-Time RT-PCR Test utilizes Biomemes M1 Sample Prep Cartridge for RNA extraction, Biomemes SARS-CoV-2 Go-Strips assay, a
The firm has become aware of nine reports by a single customer that the software made false positive calls. The investigation determined that the root cause is the higher than expected amber-to-red co-excitation in the assay.
- 2020-12-09FDA-DeviceTalladium IncClass IIILuminesse ML PMMA, 98x20mm, REF: 5336-B1 - Product Usage: used as a milling blank in the fabrication of provisional crowns and bridges. All blanks are solely used by a Trained Dental Technician or on the order of a dental professional.
Heat-cured polymethylmethacrylate (PMMA) milling blank product used to make long-term temporary crowns and bridge restorations has incorrect label. label states Part no. 5336-B1 (98x20mm). However, the item inside the box was Part no. 4302-B1 (95x20mm). An incorrect label may result in compatibility issues when setting up the disc with milling machine fixtures due to size variation.
- 2020-12-09FDA-DeviceICU Medical, Inc.Class IIREF MC330209 6' (15 cm) Appx 0.09 ml, Smallbore Trifuse Ext Set w/3 MicroClave Clear, 0.2 Micron Filter, 1.2 Micron Filter, 3 Clamps (Blue, Yellow, White), Rotating Luer - Product Usage: The intravascular administration set is a single use, sterile, non-pyrogenic device, which provides access for the administration of fluids from a container to a patient s vascular system through the administration set s needle catheter (which is inserted into a vein). Directions for Use: icumedical IV Exten
Incorrect filter assemblies in IV administration set.
- 2020-12-09FDA-DeviceVEO DIAGNOSTICS, LLCClass IIACCEL ELISA COVID-19 for the detection of total antibodies to SARS-CoV-2 Catalog AE301U Kit lots: PXCOV073020, PXCOV081220 Expiration dates: January 18, 2021 and January 26, 2021
Emergency Use Authorization (EUA) removed from list of COVID-19 antibody tests
- 2020-12-09FDA-DeviceJanus Trade GroupClass IIB. Braun Introcan Safety IV Catheters, Product codes REF 4252500-01 4252519-01 4252535-01 4252560-01 4252586-01 4251890-01 4252535-01 4252586-01 4251644-01
Distribution of B.Braun Introcan Safety IV catheters without a 510(k) within the United States by JTG.
- 2020-12-09FDA-DeviceBecton Dickinson & CompanyClass IIBD Ultra-Fine Pen Needles 32G x 4mm (5/32' )-intended for parenteral administration of a drug from a cartridge contained in a drug pen injector Catalog Number: 320122
Product shelf cartons incorrectly labeled as products intended for the Latin American market. Although the product description on the label 32G x 4mm pen needles is correct, information pertinent to users in the US regarding compatibility with specific pen needles is missing.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number: GF-UE160-AL5 - Product Usage: used for endoscopic ultrasonic imaging of the gastrointestinal wall, bile and pancreatic ducts and surrounding organs.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number: GF-UCT160-OL5 - Product Usage: used for diagnostic ultrasound imaging or fluid flow analysis of the human body.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number: GF-UC160P-OL5 - Product Usage: used for diagnostic ultrasound imaging or fluid flow analysis of the human body.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number: GF-UM160 - Product Usage: used for diagnostic ultrasound imaging or fluid flow analysis of the human body.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number: GF-UMQ130 - Product Usage: used with an Olympus Endoscopic Ultrasound Center, EVIS Video System Center, light source, documentation equipment, video monitor, Endo-Therapy accessories and other ancillary equipment for endoscopic ultrasound imaging of the gastrointestinal wall, biliary and pancreatic duct, and surrounding organs.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number: GF-UM130 - Product Usage: used with an Olympus Endoscopic Ultrasound Center, EVIS Video System Center, light source, documentation equipment, video monitor, Endo-Therapy accessories and other ancillary equipment for endoscopic ultrasound imaging of the gastrointestinal wall, biliary and pancreatic duct, and surrounding organs.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number :GF-UM20 - Product Usage: used with other supporting equipment for endoscopic real-time ultrasound imaging and endoscopic surgery within the gastrointestinal tract.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number: GF-UCT180 - Product Usage: used with an Olympus universal endoscopic ultrasound center or a diagnostic system (ALOKA CO. LTD), video system center, light source, documentation equipment, monitor, EndoTherapy accessories and other ancillary equipment.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model Number: GF-UCT140-AL5 - Product Usage: used with other supporting equipment for endoscopic real-time ultrasound imaging and endoscopic surgery within the gastrointestinal tract.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.
- 2020-12-09FDA-DeviceOlympus Corporation of the AmericasClass IIOlympus Ultrasound Endoscope, Model number: GF-UCT140P-AL5 - Product Usage: used with other supporting equipment for endoscopic real-time ultrasound imaging and endoscopic surgery within the gastrointestinal tract.
A new inspection step was added to the cleaning and disinfection section of the Instructions For Use (IFU) for this device in order to reduce the risk of causing patient infections.