- 2020-12-02FDA-DeviceFlexicare Medical Ltd.Class IProVu Single Use Video Stylet with ET Tube, Part/Size: 038-990-080U/ 8.0, 038-990-075U/ 7.5, 038-990-070U/ 7.0, 038-990-065U/ 6.5 *Note: product list updated per 12/23/2020 letter to customers
When the video stylet cuff is over-inflated there is potential for occlusion of the corrugated section of the ETT, and increased resistance to gas flow and difficulty in ventilating the patient.
- 2020-12-02FDA-DeviceMEDTRONIC ATS MEDICAL, INC.Class IICardioblate CryoFlex, Surgical Ablation Probe, Model Number 60CM1, CLAMP 60CM1 CRYOFLEX PROBE 26L - is a single use, disposable cryoprobe that is designed for use with the CryoFlex Surgical Ablation System. The probe has an integrated thermocouple for monitoring temperature at its ablation segment.
There is a potential defect in the seals of the sterile barrier pouch used to package the CryoFlex disposables devices
- 2020-12-02FDA-DeviceMEDTRONIC ATS MEDICAL, INC.Class IICardioblate CryoFlex, Surgical Ablation Probe, Model Number 60SF7, PROBE 60SF7 CRYOFLEX 7CM 26L - is a single use, disposable cryoprobe that is designed for use with the CryoFlex Surgical Ablation System. The probe has an integrated thermocouple for monitoring temperature at its ablation segment.
There is a potential defect in the seals of the sterile barrier pouch used to package the CryoFlex disposables devices
- 2020-12-02FDA-DeviceMEDTRONIC ATS MEDICAL, INC.Class IICardioblate CryoFlex, Surgical Ablation Probe, Model Number 60SF3, PROBE 60SF3 CRYOFLEX 10-S 26L - is a single use, disposable cryoprobe that is designed for use with the CryoFlex Surgical Ablation System. The probe has an integrated thermocouple for monitoring temperature at its ablation segment.
There is a potential defect in the seals of the sterile barrier pouch used to package the CryoFlex disposables devices
- 2020-12-02FDA-DeviceMEDTRONIC ATS MEDICAL, INC.Class IICardioblate CryoFlex, Surgical Ablation Probe, Model Number 60SF2, PROBE 60SF2 CRYOFLEX 10CM 26L - Product Usage: is a single use, disposable cryoprobe that is designed for use with the CryoFlex Surgical Ablation System. The probe has an integrated thermocouple for monitoring temperature at its ablation segment.
There is a potential defect in the seals of the sterile barrier pouch used to package the CryoFlex disposables devices
- 2020-12-02FDA-DeviceGeistlich Pharma North America, Inc.Class IIGeistlich Bio-Oss Pen 1-2mm 0.5g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20126
Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane
- 2020-12-02FDA-DeviceGeistlich Pharma North America, Inc.Class IIGeistlich Bio-Oss Pen 0.25-1mm 0.5g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20116
Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane
- 2020-12-02FDA-DeviceGeistlich Pharma North America, Inc.Class IIGeistlich Bio-Oss Pen 0.25-1mm 0.25g - Product Usage: natural bone mineral for filling of bone defects in maxillofacial surgery, implantology, and periodontology. Product Number: 20115
Difficult to expel Geistlich Bio-Oss granules from the pen, resistance could suddenly drop, and the granules may be expelled all at once, causing surrounding tissue, particularly in use for the sinus lift indication and while unlikely, could pose a risk for patient harm, including irritation of the oral mucosa or even a perforation of the Schneiderian membrane
- 2020-12-02FDA-DeviceIntuitive Surgical, Inc.Class IIBox Label: ION Fully Articulating Catheter, REF 490105, VER 08 - Product Usage: used to navigate the Ion Fully Articulating Catheter within the lung to a planned path in order to sample a target. User Manual: ION Endoluminal System, System, Instruments, and Accessories User Manual, PN 553990-05 REV. A 2019.12
The distal tip ring of the fully articulating catheter may become dislodged during the procedure.
- 2020-12-02FDA-DeviceColoplast Manufacturing US, LLCClass IITitan Pump - Product Usage: indicated for male patients suffering from erectile dysfunction (impotence) who are considered to be candidates for implantation of a penile prosthesis.
Potential increased occurrence of pump fracture over the lifetime of the device.
- 2020-12-02FDA-DeviceAdvanced Bionics, LLCClass IIIAB PowerCel 170 (velvet black), REF: CI-5517-150, UDI: (01) 07630016833113 used to power the following devices: Na¿da CI Q90 Sound Processor (CI-5280), Na¿da CI Q70 Sound Processor (CI-5245) and Na¿da CI Q30 Sound Processor (CI-5260). For cochlear implant.
Mislabeling; It has been determined that some batteries were improperly labeled and may contain a different battery life than indicated.
- 2020-12-02FDA-DeviceAdvanced Bionics, LLCClass IIIAB PowerCel 230 (velvet black), REF: CI-5523-150, UDI: (01) 07630016832956 used to power the following devices: Na¿da CI Q90 Sound Processor (CI-5280), Na¿da CI Q70 Sound Processor (CI-5245) and Na¿da CI Q30 Sound Processor (CI-5260) for cochlear implant.
Mislabeling; It has been determined that some batteries were improperly labeled and may contain a different battery life than indicated.
- 2020-12-02FDA-DeviceMevion Medical Systems, Inc.Class IIMEVION S250i, MEVION S250; Proton Radiation Treatment System - Product Usage: intended to deliver proton radiation treatment to patients with localized tumors or any other conditions susceptible to treatment by radiation
Treatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occur
- 2020-12-02FDA-DeviceHill-Rom Manufacturing, Inc.Class IILife2000 Ventilation System, including Life2000 Ventilator and Life2000 Compressor, Model MS-01-0118 - Product Usage: use in home and institutional settings and is not intended for ambulance or air transport.
Under certain conditions, there is a risk for fire.
- 2020-12-02FDA-DeviceBayer Medical Care, Inc.Class IIBrand Name: MEDRAD Stellant FLEX Syringe Kits Common Name: Angiographic Syringe Catalog/REF Number: FLEXD-150-SCS Description: 2 - 150 mL FLEX syringes, 1 large saline spike, 1 small contrast spike, patient t-connector tubing & prime tube. Catalog/REF Number: FLEXD-150-SPK Description: 2 - 150 mL FLEX syringes, 2 small spikes, patient t-connector tubing & prime tube.
Specific lots of MEDRAD Stellant FLEX 150 mL syringe kits may have holes and/or cracks that have the potential to allow fluid to leak out of the syringe.
- 2020-12-02FDA-DeviceViewray, Inc.Class IIViewRay MRIdian Linac System: Model No. 20000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A, UDI. The MRIdian system and the MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.
- 2020-12-02FDA-DeviceViewray, Inc.Class IIViewRay MRIdian System: Model No. 10000, CE, Rated Supply Voltage - 380/480 VAC, 3 - Frequency Range (Hertz) - 50/60 Hz, Rated Input in Amperes - 210 A, UDI. The MRIdian system and the MRIdian Linac system, with magnetic resonance imaging capabilities, is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors, and conditions anywhere in the body where radiation treatment is indicated.
There is a potential that components of the receive coil can reach elevated temperatures which has the potential to cause a burn injury to the patient or user.
- 2020-12-02FDA-DeviceEncore Medical, LPClass IIEMPOWR Partial Knee Peg Drill; Model: 800-06-008.
There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
- 2020-12-02FDA-DeviceEncore Medical, LPClass IIEMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.
There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
- 2020-12-02FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation standalone software treatment planning system, Model No. 4.0, 4.5, 4.7, 4.9, 5.0, 6.0, 6.1, 6.2, 6.3, 7.0, 8.0, 8.1, 9.0, 9.1, 9.2, 10.0 - Product Usage: RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments. The system functionality can be configured based on user nee
If a region of interest (ROI) or point of interest (POI) that is referenced from an imported plan is missing in the imported RT Structure Set, the reference may become linked to the wrong ROI or POI.
- 2020-12-02FDA-DeviceLuminex CorporationClass IIVerigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge for each sample tested.
Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENE SP devices.
- 2020-12-02FDA-DeviceLuminex CorporationClass IIVerigene Processor SP, Model Numbers 10-0000-07 and 10-0000-07R. Used with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: IVD sample processor, which utilizes a single-use disposable test consumable and a self-contained Verigene Test Cartridge for each sample tested.
Luminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in some cases, were manufactured using unvalidated processes. These text fixtures were used to calibrate and verify VERIGENESP devices.
- 2020-12-02FDA-DeviceConformis, Inc.Class IIiTotal Hip Replacement System, Model number HAA-050-D020-020102 The Conformis Hip System includes standard hip replacement components as well as the following patient specific components: femoral stem and single use instrumentation.
Incorrect hip components were provided in kits.
- 2020-11-25FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica IM 1600 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066000
Test Definition scanning may reset custom settings to defaults. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values
- 2020-11-25FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIAtellica IM 1300 Analyzer - Product Usage: automated, immunoassay analyzers designed to perform in vitro diagnostic tests on clinical specimens. Siemens Material Number (SMN): 11066001
Test Definition scanning may reset custom settings to defaults causing falsely low or high results. When a 2D Master Curve and TDef barcode for a new kit lot of reagent is scanned and the TDef version is a newer version than the version that is currently on the system, some of the customer defined settings for that assay may reset to default values
- 2020-11-25FDA-DeviceMako Surgical CorporationClass IIMako Integrated Cutting System (MICS) Handpiece, Catalog Number 209063, Software Version#: THA 3.0 and later, PKA 3.0 and later, and TKA 1.0 - Product Usage: The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO (Mako robot) and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later).
Stryker has discovered a characterization issue associated with the Mako Integrated Cutting System (MICS) Handpiece. The MICS Handpiece is attached to the distal end of the robotic arm and is used to register the RIO and to execute bone preparation during a Mako Total Knee or Mako Partial Knee procedure (PKA Software Application 3.0 or later). This issue may result in unsuccessful checkpoint verification when the Mako System attempts to verify the location of the saw blade prior to bone preparation. This issue may also cause a discrepancy during bone preparation in a Mako Total Knee or a Mako Partial Knee procedure in which the surgeon uses a straight or angled saw attachment with the MICS Handpiece. The surgeon may make cuts during bone preparation that differ from the surgeon s preoperative plan.
- 2020-11-25FDA-DeviceOrtho Clinical Diagnostics IncClass IIIVITROS Chemistry Products Performance Verifier II-For in vitro diagnostic use only.VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated System - Product Code: 8231474
Product instability; biased ALKP values obtained from VITROS Chemistry Products Performance Verifier II Lot Q7692 when stored unopened at refrigerated temperatures
- 2020-11-25FDA-DeviceMicro-Tech UsaClass IIElastic Traction Device, 11mm Lower. UPN: ER56041 Reference Number: ETD-D-26-235-C-N - Product Usage: indicated for use in flexible Endoscopy to provide retraction to assist in tissue resection, exposure, and removal of tissue within the gastrointestinal tract.
FDA has informed Micro-Tech that the Elastic Traction Device is not appropriately classified as a Class I device and requires 510(k) clearance to be marketed in the US.
- 2020-11-25FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis zee floor, Artis Q ceiling, Artis Q.zen ceiling. Image-Intensified Fluoroscopic X-Ray System. Material numbers 10848281, 10848354, 10094135, 10848354 - Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Some electronic components within the generator were equipped with an incorrect protection diode. As a result, protection against overvoltage may be inadequate, and the system may not function properly in the event an overvoltage situation occurs.
- 2020-11-25FDA-DeviceColtene Whaledent IncClass IIStrauss 25 Carbide Dental Burs (100057XXV)
The FG 557 bur was incorrectly packaged as the FG 57 bur. The FG 557 burs will remove more material than the FG 57 burs.