Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2020-08-12FDA-DeviceMedline Industries IncClass IIEndoscopic Cleaning Brush - Product Usage: Endoscopic cleaning brushes are single use devices intended to clean the interior of endoscopes after endoscopy procedures. The brush is threaded through the biopsy valve and pulled through the interior of the endoscope to remove debris.
Potential risk of brush detachment during cleaning.
- 2020-08-12FDA-DeviceBoston Scientific CorporationClass IICaptivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loo
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
- 2020-08-12FDA-DeviceBoston Scientific CorporationClass IICaptiflex Snares, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snares flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loop.
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
- 2020-08-12FDA-DeviceBoston Scientific CorporationClass IICaptivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loo
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
- 2020-08-12FDA-DeviceBoston Scientific CorporationClass IICaptivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loo
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
- 2020-08-12FDA-DeviceBoston Scientific CorporationClass IICaptivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loo
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
- 2020-08-12FDA-DeviceBoston Scientific CorporationClass IICaptivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loo
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
- 2020-08-12FDA-DeviceBoston Scientific CorporationClass IICaptivator Polypectomy Snare, Single-Use Polypectomy Snare consists of a flexible wire cable and loop which can be extended and retracted from the snare s flexible outer sheath using a three-ring handle. The inner diameter of the sheath is PolyGlide" Lubricant coated to provide minimal friction during extension and retraction of the loop from the sheath. When passed through an endoscope the snare can be activated to deliver a monopolar electrical current to cut and cauterize tissue with the loo
BSC has noted a potential inability to cut and remove polyps for these lots/batches. potential consequences related to the snares not performing as intended would be a prolonged procedure duration, tissue damage and/or self-limited bleeding. The most severe potential injury would include immediate or delayed hemorrhage and/or perforation requiring intervention up to and including open surgery; the likelihood of these severe injuries occurring is remote and, to date, have not been reported as a complaint. The potential inability to cut may also lead to inadequate removal of target tissue.
- 2020-08-12FDA-DeviceSiemens Medical Solutions USA, IncClass IISIEMENS Healthineers Ysio Max - Interventional Fluoroscopic X-Ray System, Model: 10762470 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis
- 2020-08-12FDA-DeviceSiemens Medical Solutions USA, IncClass IISIEMENS Healthineers Uroskop Omnia max Model: 10762473 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis
- 2020-08-12FDA-DeviceSiemens Medical Solutions USA, IncClass IISIEMENS Healthineers Luminos Agile Max Model: 10762472 - Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis
- 2020-08-12FDA-DeviceSiemens Medical Solutions USA, IncClass IISIEMENS Healthineers Luminos dRF Max Model: 10762471- Product Usage: is a device intended to visualize anatomical structures by converting an X-ray pattern into a visible image.
Two software issues (1) Using the override function in case of blocked system movements affecting Luminos dRF Max, Luminos Agile Max and Uroskop Omnia Max systems may cause collision of the system under operator control with obstacles or persons .(2) Incorrectly assigned image affecting Ysio Max, Luminos dRF Max and Luminos Agile Max systems with detector MAX Static causing incorrect base for diagnosis
- 2020-08-12FDA-DeviceBecton Dickinson & Co.Class IIBD MAX SARS-CoV-2 Reagents, Cat. No. 445003 (UDI 00382904450038) and 445003-01 (UDI (01)603829044500301) - Product Usage: intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.
BD is experiencing elevated rates of potential false positive results from certain customers when using the BD SARS-CoV-2 Reagents for the BD MAX System. False positive test results could lead to a delay in diagnosis and potentially cause the patient to be exposed to COVID-19 if they are isolated with an infected individual.
- 2020-08-12FDA-DeviceRANDOX LABORATORIES, LTD.Class IIRandox Lipase Assay: Lipase Colorimetric R2 Substrate - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.
Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase R2 Substrate batches.
- 2020-08-12FDA-DeviceRANDOX LABORATORIES, LTD.Class IIRandox Lipase Assay: Lipase Colorimetric R1 Buffer - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.
Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase buffer batches.
- 2020-08-12FDA-DeviceRANDOX LABORATORIES, LTD.Class IIRandox Lipase Assay: Lipase Colorimetric Reagent - Product Usage: A Lipase test system is a device intended for the quantitative in vitro determination of Lipase in human serum and plasma.
Randox have confirmed imprecision of quality control and patient samples when using specific Randox Lipase reagent batches.
- 2020-08-12FDA-DeviceGE Healthcare, LLCClass IICARESCAPE ONE, Physiological Patient Monitor - Product Usage: is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility.
CARESCAPE ONE may not provide visual and audible alarms for Ventricular Fibrillation (V Fib), if V Fib occurs at the time CARESCAPE ONE is docked to a CARESCAPE B450/B650/B850 host monitor.
- 2020-08-12FDA-DeviceTELEFLEX MEDICAL INCClass IIPleur-Evac Adult-Ped Wet, LF Eng/Spanish, product code: A-8000I-06 - Product Usage: CHEST DRAINAGE: To evacuate air and/or fluid from the chest cavity or mediastinum. To help prevent air and/or fluid from re-accumulating in the chest cavity or mediastinum. To help re-establish and maintain normal intra-thoracic pressure gradients. To facilitate complete lung re-expansion to restore normal breathing dynamics.
Potential for sterile packaging to be compromised¿
- 2020-08-12FDA-DeviceTELEFLEX MEDICAL INCClass IIPleur-Evac Adult-Ped Wet, LF Intl Eng, product code: A-8000I - Product Usage: CHEST DRAINAGE: To evacuate air and/or fluid from the chest cavity or mediastinum. To help prevent air and/or fluid from re-accumulating in the chest cavity or mediastinum. To help re-establish and maintain normal intra-thoracic pressure gradients. To facilitate complete lung re-expansion to restore normal breathing dynamics.
Potential for sterile packaging to be compromised¿
- 2020-08-12FDA-DeviceSiemens Healthcare Diagnostics, Inc.Class IIDCA Vantage Handheld Barcode Scanner - Zebra Model - Model # DS4308 - HC0015BZZWW external accessory to DCA Vantage Analyzer US
If DCA Vantage Analyzer is configured to run Code 39 with check digit, the scanner is not confirming the integrity of read barcode data using the check digit. Also, if DCA Vantage Analyzer is configured to run Code 39 with check digit and an operator erroneously scans a Code 39 barcode without a check digit, the scanner incorrectly sends the code and the last character is removed.
- 2020-08-12FDA-DeviceSmith & Nephew, Inc.Class IISUTUREFIX ULTRA Suture, Product number 72203855
A packaging error resulted in a XL drill being presented for use instead of the S drill. The XL drill is one inch longer than the S drill.
- 2020-08-12FDA-DeviceMako Surgical CorporationClass IIMako Hip End Effector, Variable Angle Catalog Number: 206967 GTIN:00848486022167
Characterization issue associated with the product. This issue may cause a discrepancy during bone preparation in certain Mako Total Hip procedures.
- 2020-08-12FDA-DeviceSutter Medizintechnik GmbHClass IISutter Swyng non-stick bipolar forceps - Product Usage: single-use re electrosurgical instruments. The bipolar for-ceps are provided with bayonet-style handle design with straight or angled tips and different total lengths. They are to be connected through an appropriate bipolar cable with the bipolar output of an electrosurgical generator. The electrodes are provided sterile and are single-use instruments.
The label on the cardbox and the blister show the correct symbols for sterilization using irradiation and storage at 5 C to 40 C, the package insert (IFU) is labelled incorrectly. It shows the symbol for sterilization using Ethylene Oxide with the description for sterilization using irradiation.
- 2020-08-12FDA-DeviceNormand-Info S.A.S.U.Class IIRemisol Advance running Software Version 1.12 The Normand Info Remisol Advance Data Manager collects and manages data and workflow for 128 connections (Beckman Coulter Instruments, automations, LIS &). It can also be connected to a host system.
Results from repeated run for WBC (White Bloodcell Count), UWBC (Urine Whit Bloodcell Count) and PLT (Platelets) are deleted in RADV. Use may lead to erroneous results.
- 2020-08-12FDA-DeviceCareFusion 303, Inc.Class IAlaris System PC Unit Model 8015. modular infusion pump and monitoring system
If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.
- 2020-08-12FDA-DeviceCareFusion 303, Inc.Class IAlaris System PC Unit Model 8000 modular infusion pump and monitoring system
If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.
- 2020-08-12FDA-DeviceCareFusion 303, Inc.Class IAlaris Pump Module Model 8100
Broken elements on pump module platen such as broken upper hinge post, lower hinge, and membrane frame can occur when the door is forcefully closed on an object, the device is dropped, or if improper cleaners are used on the membrane frame assembly. These may prevent devices from delivering accurate amounts of fluid, which may result in an over infusion, free-flow conditions, or under infusion.
- 2020-08-12FDA-DeviceSmiths Medical ASD Inc.Class IMedfusion Syringe Pump Model 3500, Software Version V6.0. Infusion pump.
Inaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
- 2020-08-12FDA-DeviceSmiths Medical ASD Inc.Class IMedfusion Syringe Pump Model 4000; Software Versions V1.5.0, V1.5.1, V1.6.0, and V1.6.1
Inaccurate delivery can occur following an interrupted Bolus or Loading dose, if a specific sequence of events occurs.
- 2020-08-12FDA-DeviceCareFusion 303, Inc.Class IAlaris Auto ID Module Model 8600, a modular infusion pump and monitoring system
Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.