Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2020-08-19FDA-DeviceWright Medical Technology, Inc.Class IIHardware removal kit, Part Number 1985KIT1 - Product Usage: The Wright Medical hardware removal kit is designed to facilitate removal of both plates and screws used in foot and ankle procedures. Along with a complete range of drivers, the kit also features driver handles and a range of k-wires to assist in removal of cannulated screws. This product is a multiple use, non-sterile instrument.
Hardware removal kit does not include drivers to remove Dart-Fire 2.5 and 3.0mm Headless screws.
- 2020-08-19FDA-DeviceBioseal CorporationClass IIBioseal Trachea Extender 1/pl 50pk/Cs, REF TRAX01/50, LOT XXXX, STERILE EO, Single Use Only, Date of Manufacture XXXX, Expiration Date XXXX, LF, UDI: (01) 00630094916019 - Product Usage: Airway connectors intended to connect a breathing gas source to a tracheal tube, tracheostomy tube, or mask.
The reported stress crack in the port of the swivel elbow where the flip cap is inserted could result in leakage from the breathing system resulting in the prescribed ventilation not being delivered to the patient. These cracks could possibly leak if they reach a significant size and pose a serious heath risk to patient.
- 2020-08-19FDA-DevicePhilips North America, LLCClass IIPhilips HeartStart MRx, Monitor/Defibrillators with model numbers M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9 - Product Usage: is intended for use in hospital and pre-hospital settings by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support or defibrillation.
The therapy switch may fail, resulting in the device exhibiting the following behaviors: the device may not perform the selected function, the therapy knob may not change to the energy setting selected, or the device may deliver a shock with an energy level different from the setting selected by the user. These device behaviors could result in a delay in therapy or a failure to deliver the intended therapy.
- 2020-08-19FDA-DeviceBrainlab AGClass IIBRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655
Incorrect manufacturer calibration
- 2020-08-19FDA-DeviceBrainlab AGClass IIBRAINLAB Curve; Model/catalogue (article) numbers for Curve: 19901; 19901B; 19900; 19905 UDI (GTIN) for Curve: 04056481138011
Incorrect manufacturer calibration
- 2020-08-19FDA-DeviceHF Acquisition Co., LLCClass IISTAT KIT Model 550AI Emergency Medical Kit. P/N: 1010830 containing various products to include Narcan Nasal spray, item number 1010300
Instead of 2 doses, packaged 1 dose of nasal spray in an emergency kits.
- 2020-08-19FDA-DeviceHF Acquisition Co., LLCClass IISTAT KIT Model 550 Emergency Medical Kit P/N: 1009580 containing various products to include Narcan Nasal spray, item number 1010300
Instead of 2 doses, packaged 1 dose of nasal spray in an emergency kits.
- 2020-08-19FDA-DeviceHF Acquisition Co., LLCClass IISTAT KIT, Model SM30 ; Emergency Medical Kit P/N: 1008670 containing various products to include Narcan Nasal spray, item number 1010300
Instead of 2 doses, packaged 1 dose of nasal spray in an emergency kits.
- 2020-08-19FDA-DeviceHF Acquisition Co., LLCClass IIHealthFirst Emergency Medical Kit, Model SM 7, P/N: 1008150; Manufactured by HF Acquisition Co. LLC, 11629 49th Place West/Mukilteo, WA 98275, HealthFirst.com; Store at Room Temperature. 9036-2v.5 08/2017 containing various products to include Narcan Nasal spray, item number 1010300
Instead of 2 doses, packaged 1 dose of nasal spray in an emergency kits.
- 2020-08-19FDA-DeviceFlower Orthopedics CorporationClass IICalcaneus Plate, Extensile, Large, Right, Catalog Number CPF 203 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.
The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
- 2020-08-19FDA-DeviceFlower Orthopedics CorporationClass IICalcaneus Plate, Extensile, Medium, Right, Catalog Number CPF 202 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.
The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
- 2020-08-19FDA-DeviceFlower Orthopedics CorporationClass IICalcaneus Plate, Extensile, Small, Right, Catalog Number CPF 201 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.
The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
- 2020-08-19FDA-DeviceFlower Orthopedics CorporationClass IICalcaneus Plate, Extensile, Large, Left, Catalog Number CPF 103 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.
The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
- 2020-08-19FDA-DeviceFlower Orthopedics CorporationClass IICalcaneus Plate, Extensile, Medium, Left, Catalog Number CPF 102 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.
The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
- 2020-08-19FDA-DeviceFlower Orthopedics CorporationClass IICalcaneus Plate, Extensile, Small, Left, Catalog Number CPF 101 - Product Usage: intended to be used for internal fixation of fractures and reconstruction of bones of the rearfoot, including the calcaneus.
The contract manufacturer of the calcaneus plate inadvertently used the incorrect material to manufacture the plates (a titanium alloy rather than pure titanium).
- 2020-08-19FDA-DeviceLinet AmericasClass IIIAVE 2 Birthing Bed - Product Usage: is designated to be used by mothers ready for spontaneous vaginal delivery and also as aid to search for individual positions providing relief in all delivery stages including postpartum period. Birthing bed is not intended for use as a general hospital bed.
The UDI on the device label indicates incorrect manufacturer.
- 2020-08-19FDA-DeviceMaquet Cardiovascular Us Sales, LlcClass IIHolder HKHZ 19 for use with Quadrox-i and Quadrox-iD Neonatal / Pediatric Oxygenators, Model Number: 701047495 - Product Usage: The Holder HKHZ 19 is an accessory to QUADROX i and QUADROX iD Neonatal / Pediatric oxygenator and is being used as a single holder for QUADROX i and QUADROX iD Neonatal / Pediatric Oxygenator and can also be used in combination with the holder's arm (016061) to use in cardiopulmonary bypass procedures (CPB) in a pediatric patient population.
Removal for a specific lot, lot number 70128207, of Holder HKHZ 19 for use with QUADROX iD Pediatric Oxygenator. The Holder may not always snap and securely hold the oxygenator as intended.
- 2020-08-19FDA-DeviceHF Acquisition Co. LLCClass IISTAT KIT Z-1000 emergency medical kit HealthFirst - Product Usage: is intended to prepare physicians with advanced cardiac life support devices and medications to respond to cardiopulmonary medical emergencies until EMS arrives.
A missing component (hand suction pump) in the emergency medical kits.
- 2020-08-19FDA-DeviceAnjon HoldingsClass IIAnjon Bremer Molded Crown. bone fixation fastener for spinal use. REF number 100048 - Product Usage: Smooth or threaded metallic bone fixation fastener. Provides cervical spine immobilization and therapeutic traction for treatment of patients with cervical trauma or other neck conditions.
High rate of galling of one threaded component.
- 2020-08-19FDA-DeviceSmith & Nephew, Inc.Class IIPERI-LOC TARGETER 4.5MM DISTAL FEMUR, HANDLE, LEFT, NON-STERILE - Product Usage: used for adult and pediatric patients, as well as patients with osteopenic bone. PERI-LOC plates and screws are indicated for fixation of pelvic, small and long bone fractures, including those of the tibia, fibula, femur, pelvis, acetabulum, metacarpals, metatarsals, humerus, ulna, radius, calcaneus and clavicle.
The 4.5MM PeriLoc Targeter right hand part was anodized green and was marked as a left-hand part.
- 2020-08-19FDA-DeviceLuminex CorporationClass IIVerigene Processor SP with assay Luminex, VERIGENE Gram-Negative Blood Culture Nucleic Acid Test, BC-GN, IVD - Product Usage: utilizes a single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested.
An instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.
- 2020-08-19FDA-DeviceLuminex CorporationClass IIVerigene Processor SP with assay: Luminex, VERIGENE Gram-Positive Blood Culture Nucleic Acid Test,BC-GP, IVD - Product Usage: utilizes a single-use disposable test consumables and a self-contained Verigene Test Cartridge for each sample tested.
An instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.
- 2020-08-19FDA-DeviceSmith & Nephew, Inc.Class IITibial knee prosthesis
Due to an inconsistency in the raw material process, specific lots may contain units with internal non-homogenous material defects.
- 2020-08-19FDA-DeviceIon Beam Applications S.A.Class IIProteus 235 The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.
Signature from the user is necessary to proceed with specific actions in the Proton Therapy System (PTS). IBA became aware that the PTS does not accept user names with more than ten characters. It is an issue when the user has no other choice than resuming an aborted treatment field based on the overall delivered dose displayed on the Dose Counter Electronic Unit (DCEU). User s signature is required to perform this action. If the signature contains more than ten characters, the user will not be able to complete the aborted treatment field. An additional issue applies to Electronic Medical Record (EMR) centric sites and may increase the probability of not being able to complete an aborted treatment field. It is not possible, in a new session, to resume from a local partial archive if the Patient Positioning System (PPS) position has changed. If the user captures the PPS position at every session in the Oncology Information System (OIS) just after the setup process, the prescribed PPS position is changed for the next session in the OIS. This includes the partial continuation session. Therefore, when comparing the prescribed PPS position between OIS and local database, the PTS sees a difference and rejects the local partial archive. This problem forces the user to resume the interrupted irradiation based on the overall delivered dose displayed on the DCEU instead of resuming from the full details of the interrupted beam.
- 2020-08-19FDA-DeviceNeuroLogica CorporationClass IIOmniTom Computed tomography x-ray system, Model NL5000, all NL5000 systems with software version before 05.01.01 - Product Usage: is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 40 cm aperture, primarily head and neck.
There are potential issues related to the operational state of the scanner, specifically for transport and inadvertent system shutdowns. There could be a possibility of potential delay in patient treatment if proper use instructions are not followed.
- 2020-08-19FDA-DeviceOrtho Clinical DiagnosticsClass IIVITROS Anti-SARS-CoV-2 IgG Reagent Pack, COV2IgG, VITROS CoV2G. Catalog number 6199919, UDI 10758750033355 - Product Usage: Rx ONLY. For in vitro diagnostic and Laboratory Professional use. For emergency authorization use only.
The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality control failures.
- 2020-08-19FDA-DeviceOrtho Clinical DiagnosticsClass IIVITROS Anti-SARS-CoV-2 Total Reagent, COV2TOT, VITROS CoV2T. Catalog number 6199922, UDI 10758750033386 - Product Usage: Rx ONLY. For in vitro diagnostic and Laboratory Professional use. For emergency authorization use only.
The reagent bottles may be swapped within the reagent pack, which may lead to false negative results or delayed results due to calibration failures, or quality control failures.
- 2020-08-12FDA-DeviceCovidien LlcClass IIBeacon EUS Access System Preloaded Access 135degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.
This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.
- 2020-08-12FDA-DeviceCovidien LlcClass IIBeacon EUS Access System Preloaded Access 90degree Needle sterile, single-use, endoscopic ultrasound device consisting of a sharp stylet and an 18.5-gauge, echo-enhanced, tip-shaped, nitinol cannula intended to pierce tissue and then act as an access channel to facilitate guidewire placement to locations of the gastrointestinal tract.
This voluntary recall is being conducted following three customer reports of the device cannula detaching during a procedure. In one case, the detached cannula was not retrieved during the procedure. Potential risks associated with a detached cannula include delay in treatment, biliary leak, pain and retained device fragments.
- 2020-08-12FDA-DeviceBoss Instruments, Ltd.Class IIBOSS DeBakey "Gator-Grip" Needle Holder 2 mm tip, 9", Part Numbers 50-2503 and 51-2503 - Product Usage: Endoscopic cleaning brushes are single use devices intended to clean the interior of endoscopes after endoscopy procedures. The brush is threaded through the biopsy valve and pulled through the interior of the endoscope to remove debris.
The tungsten carbide plate, which is intended to hold the sewing needle securely, can break when holding the sewing needle under pressure.