- 2020-03-18FDA-DeviceMedicrea InternationalClass IIIliac Screws, multiple sizes, Article Nos. B36020500 B36020570 B36020580 B36020590 B36027500 B36027570 B36027580 B36027590 B36028500 B36028570 B36028580 B36028590 B36029500 B36029570 B36029580 B36029590 Screws manufactured in titanium alloy Ti-6Al-4V LI conforming to ISO5832-3 and ASTM. F136 specifications. These screws are part of an internal fixation system composed of hooks, rods, plates, cross links, connection and locking devices.
An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.
- 2020-03-18FDA-DeviceMedicrea InternationalClass IIPedicle Screws, multiple sizes, Article Nos. B36004525 B36004530 B36004535 B36004540 B36004545 B36004550 B36004555 B36005530 B36005535 B36005540 B36005545 B36005550 B36005555 B36005560 B36006530 B36006535 B36006540 B36006545 B36006550 B36006555 B36006560 B36007535 B36007540 B36007545 B36007550 B36007555 B36007560 B36008535 B36008540 B36008545 B36008550 B36008555 B36008560 B36009535 B36009540 B36009545 B36009550 B36009555 B36009560 Screws manufactured i
An internal component was assembled in the incorrect orientation, which may lead to the tulip head disengaging from the crew shank during or after screw insertion, or during the reduction step.
- 2020-03-18FDA-DeviceVerathon, Inc.Class IIGlideScope Video Laryngoscope, Operations & Maintenance Manual, 0900-4940 REV 03
A typographical error in the Operations & Maintenance Manual incorrectly identifies the use of low level disinfection for the reusable video laryngoscope, specifically for the Titanium Blades.
- 2020-03-18FDA-DeviceIntegra LifeSciences Corp.Class IIINeuraGen Nerve Guide, 7mm diameter, 2cm length; Part number PNG72 - Product Usage: NeuraGen nerve guide is indicated for the repair of peripheral nerve discontinuities where gap closure can be achieved by flexion of the extremity.
Nonconforming product which was dispositioned for rejection was mistakenly released for commercial use.
- 2020-03-18FDA-DeviceThe Binding Site Group, Ltd.Class IIVaccZyme Human Anti-Haemophilus influenza type b Enzyme Immunoassay Kit For in vitro diagnostic use, Product Code: MK016 - Product Usage: This assay is designed for the in vitro measurement of specific IgG antibodies against Haemophilus influenzae type b (Hib) capsular polysaccharide, present in human serum. Sufficient materials are supplied to allow a maximum of 41 samples to be tested in duplicate, with a calibration curve and two controls.
A deterioration of performance was identified with influenza type B Enzyme Immunoassay Kits with findings of a positive bias with kit controls. Patient sample results may be affected with falsely elevated results incorrectly indicating a protective level of anti-Hib antibody in the assessment of immunodeficiency.
- 2020-03-18FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation/RayPlan 9B, Build Number 9.1.0.933, stand-alone software treatment planning system
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
- 2020-03-18FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation/RayPlan 9A, Build Number 9.0.0.113, stand-alone software treatment planning system
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
- 2020-03-18FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation/RayPlan 8B Service Pack 2, Build Number 8.1.2.5, stand-alone software treatment planning system
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
- 2020-03-18FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation/RayPlan 8B Service Pack 1, Build Number 8.1.1.8, stand-alone software treatment planning system
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
- 2020-03-18FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation/RayPlan 8B, Build Number 8.1.0.47, stand-alone software treatment planning system
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
- 2020-03-18FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
- 2020-03-18FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation/RayPlan 8A, Build Number 8.0.0.61, stand-alone software treatment planning system
There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.
- 2020-03-18FDA-DeviceResponse Biomedical Corp.Class IIRandox Laboratories Ltd. Catalogue Number CQ5052, Lot Number 4244CK (relabeled as Response Biomedical Catalogue Number C2003-2, Lot #070318) IFU: "RAMP Cardiac Controls *** CAT. No.: C2003-2 *** Size: 3 x 3 ML". Product Usage: use in the quality control of cardiac marker assays run on the RAMP¿ platform.
The form received three complaints related to the user obtaining a result that fell below the target range for CK-MB provided in the RAMP Cardiac Controls Instructions for Use. Via in-house monitoring, determined that liquid Cardiac Control has seen a decrease in CK-MB Levels of up to 57% signal loss within 11 months. Shelf-life of product is 24 months.
- 2020-03-18FDA-DeviceBraemar Manufacturing, LLCClass IIBraemar Manufacturing LLC DL950 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only
An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.
- 2020-03-18FDA-DeviceBraemar Manufacturing, LLCClass IIBraemar Manufacturing LLC DL900 Holter Monitor 1285 Corporate Center Dr. Suite 150 Eagan, MN 55121 USA Rx Only
An issue with the DL900 and DL950 Series Holter Recorder was recognized on January 1, 2020. This issue can prevent the recorder from operating normally. Braemar confirmed that beginning on January 1, 2020 if a AAA battery is inserted in the recorder and a user attempts to start it, the recorder will display Error: 602 and fail to function for a new patient study. There are no actions that a clinical user can take to clear this error. Even if this error code is displayed, existing ECG recordings stored on the device can still be downloaded via the standard Holter Application Software. Error 602 may be able to be cleared by technical or engineering staff at the customer site.
- 2020-03-18FDA-DeviceNextremity SolutionsClass IINextra Hammertoe Correction System-4.5MM Middle, 3.2MM Proximal, SS, Sterile. REF: NX-4532K - Product Usage: The Nextra Hammertoe Correction System is a system of implants and associated instruments which is indicated for small bone reconstruction limited to inter-digital repair and fusion of the lesser toes.
Proximal end of the driver in the affected product may be oversized in varying degrees, potentially making it difficult to connect with and remove from the proximal Nextra implant
- 2020-03-18FDA-DeviceEHOB, Inc.Class IIEconocare Plus Overlay Item: 1025EC. Intended to aid in the prevention and treatment of pressure injuries and general comfort - Product Usage: The EHOB WAFFLE Overlay intended use is as first line of defense for pressure injury prevention. The design provides the caregiver a point-of-care solution with the versatility of patient handling and transferring, which encourages ease of use and compliance. The design provides airflow to aid in patient comfort and pressure injury prevention.
Due to a nonconformance of product contamination (human blood) of Econocare Plus Waffle Overlay products with the (master) lot # 932213A that may affect a subset of the lot- controlled product (# 932243A). This nonconformance has the potential to expose the end user to materials deemed to be a health risk. Firm is aware of zero instances of contamination in distributed products that have reached consignees.
- 2020-03-18FDA-DeviceCardinal Health 200, LLCClass IIMonoject Bluntfill with Filter, 18G x 1-1/2" Item Code 11811022F. Used for drawing fluid from a vial - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampules and parts of the body below the surface of the skin.
Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.
- 2020-03-18FDA-DeviceCardinal Health 200, LLCClass IIMonoject Hypodermic Safety Needle 25G x 5/8" Item Code 1182558 - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampules and parts of the body below the surface of the skin.
Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.
- 2020-03-18FDA-DeviceCardinal Health 200, LLCClass IIMonoject Hypodermic Safety Needle 30G X 1/2" Item Code 1183005 - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampules and parts of the body below the surface of the skin.
Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.
- 2020-03-18FDA-DeviceCardinal Health 200, LLCClass IIMonoject 3mL Syringe with Hypodermic Safety Needle, 22G x 1-1/2 . Item Code 11832215 - Product Usage: used for general purpose injection and aspiration of fluids from vials, ampules and parts of the body below the surface of the skin.
Non-sterile product was shipped to customers. Product package is labeled as both "STERILE" and "NON-STERILR" ; the product did not undergo sterilization process and is non-sterile. The usage of a needle that is not sterile could result in infection. Firm is not aware of any reports of patient harm.
- 2020-03-18FDA-DeviceARJOHUNTLEIGH POLSKA Sp. z.o.o.Class IIConcerto & Basic Shower Trolly, Model Numbers BAB1000-01 (UDI: (01)05055982788083), BAB1000CON6102 [(01)05056097364483], BAB1007-01 [(01)05055982796255], BAB1101-01 [(01)05055982787949], BAB1102-01 [(01)05055982796156], BAB1103-01 [UDI: (01)05055982796163], BAB1107-01 [(01)05055982796187], BAB2000-01 [UDI: (01)05055982788106], BAB2101-01 [(01)05055982788007], BAB2102-01 [01)05055982796194], BAB3000-01 [(01)05055982788120], BAB3101-01 [(01)05055982788021], BAB3107-01[ (01)05055982796248], BAB500
The firm has become aware that any of the 4 safety catches may exhibit a mechanical failure (break in one of the arms of safety catch), which in certain circumstances may contribute to the side support unintentionally opening and may lead to a potential risk of a patient fall. The patient falling could lead to various types of injuries.
- 2020-03-18FDA-DeviceBioFire Diagnostics, LLCClass IIFilmArray Gastronintestinal (GI) Panel, REF: RFIT-ASY-0104, 6-pack kit, UDI (01)00815381020116, CE IVD, Rx Only
Complaints have been received concerning elevated rates of false negative results while using the Gastrointestinal (GI) Panel. False negative results may lead to incorrect treatment being provided which could potentially lead to more serious health events.
- 2020-03-18FDA-DeviceMedtronic CoreValve LLCClass IIMedtronic Simulus 850 Semi Rigid Ring/Band Accessory Kit Used to size a patient's mitral or tricuspid valve in cardiac procedures.
Demonstration Simulus Reusable Sizer Accessory Kits were being incorrectly sourced to customers. The demonstration product (product used within a wet lab) could have come in contact with non-human (porcine) tissue.
- 2020-03-18FDA-DeviceMedtronic CoreValve LLCClass IIMedtronic Simulus 750 Flexible Ring/Band Accessory Kit - Product Usage: Used to size a patient's mitral or tricuspid valve in cardiac procedures.
Demonstration Simulus Reusable Sizer Accessory Kits were being incorrectly sourced to customers. The demonstration product (product used within a wet lab) could have come in contact with non-human (porcine) tissue.
- 2020-03-18FDA-DeviceCME America, LLCClass IBodyGuard Microset REF A120-003XYVA, Sterile EO, Tubing: 150 in. (380 cm), Prime Vol: approx. 8ml Product Usage: The BodyGuard Infusion Pump System is designed to transfer medication and fluids intravenously.
When using a specific tubing set with infusion pump system, may result in under deliver of fluids.
- 2020-03-18FDA-DeviceHeartware, Inc.Class IHeartWare HVAD System Battery Charger AC Adapter The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
- 2020-03-18FDA-DeviceHeartware, Inc.Class IHeartWare HVAD System Battery Charger The HeartWare HVAD System is indicated for hemodynamic support in patients with advanced, refractory left ventricular heart failure; either as a Bridge to Cardiac Transplantation (BTT), myocardial recovery, or as Destination Therapy (DT) in patients for whom subsequent transplantation is not planned.
Potential for an HVAD System user to mistakenly insert the Battery Charger AC Adapter into an HVAD Controller power port.
- 2020-03-11FDA-DeviceBiofire DefenseClass IIIFilmArray NGDS Warrior Panel, Ref: NGDS-ASY-0007, UDI: (01)00851458005136, IVD, Rx Only
The firm has identified that an incorrect instruction for use (IFU) revision was included in the qualitative, multiplexed, nucleic acid-based in vitro diagnostic test kit.
- 2020-03-11FDA-DeviceCovidien LlcClass IIKits containing Covidien Force TriVerse Electrosurgical Devices. Distributed in EMEA (Europe, Middle East, and Africa) only. No US distribution. Kits are manufactured in The Netherlands. Item codes BREASTKIT01 BREASTKIT01 BREASTKIT01 BREASTKIT01 BREASTKIT01 BREASTKIT01 BREASTKIT01 BREASTKIT01 BREASTKIT02 LAR28PLN MGBGLOCX1 SGGLOCX1 HEMICX1 LAPWL STOM25PL TGGLOCX1 TIROIDEKIT UROCYSPL VAGHCJ VATSZD2 VGYN VMED VMED3 ZGVATS1
Sterile packaging may potentially be compromised, and use of products may result in increased risk for infection.