- 2020-03-25FDA-DeviceTeDan Surgical Innovations LLCClass IID-0012 3COR 12MM DISTRACTION SCREWS, 5/BOX, STERILE - Product Usage: The 3COR Distraction Screws and the TSI Quick Start Distraction Screws products are intended for the distraction in the anterior approach to the cervical spine from C2 thru C7. The products are to be used by trained surgeons and are designed for use with TSI distractors and screwdrivers.
The items identified for recall are labeled as sterile. However, the manufacturer has found that product packaging for the lots listed above may be damaged, potentially exposing the product to non sterile conditions. This situation presents a risk of infection to patients.
- 2020-03-25FDA-DeviceAgfa N.V.Class IIAGFA Digital Radiography X- Ray system DR 800 with TomoSynthesis- a Tomographic X-ray system - Product Usage: TomoSynthesis is used to synthesize tomographic slices from a single tomographic sweep. The DR 800 is not intended for mammography applications
Tomosynthesis image acquisition sequence did not stop automatically after expected number of exposures.
- 2020-03-25FDA-DeviceSiemens Medical Solutions USA, IncClass IIYsio, model no. 10281013 Product Usage: The product is a stationary X-Ray system.
The Ysio system may process incorrect internal values for the position of its ceiling tube stand. If in this situation the user activates longitudinal or transversal stand movement (by using the buttons at the tube head or the remote console), the systems controls may (depending on the values being processed) initiate a correction of the tube stand position, resulting in an unexpected movement.
- 2020-03-25FDA-DeviceThe Binding Site Group, Ltd.Class IIRheumatoid Factor (RF) Kit for use on SPAPLUS, for in-vitro diagnostic use, Product code: LK151.S.A. , UDI: 05051700018319
It has been identified that the rheumatoid factor (RF) calibrator values over recover in comparison to the reference material RF serum. This could result in reporting falsely abnormal results which may lead to unnecessary additional serological testing.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIHeartStart MRx Monitor/Defibrillators with model numbers M3535A M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, and M3536M9.
Philips has received a number of reports of HeartStart MRx Monitor/Defibrillators that have suffered internal damage and were not able to deliver therapy after having been dropped or subjected to a severe mechanical shock, even though the device did not have visible external damage or the Ready for Use ( RFU ) indicator on the unit did not immediately indicate a problem. One of these reports involved a death of a patient who could not be resuscitated.
- 2020-03-25FDA-DeviceLivaNova USA Inc.Class IILivaNova Heater-Cooler System 3T - Product Usage: The Heater-Cooler 3T System is used to circulate water through heat exchangers to warm or cool a patient during cardiopulmonary bypass procedures lasting 6 hours or less.
The firm is providing validated cleaning and disinfection instructions in their revised operating instructions to reduce the risk of NTM contamination.
- 2020-03-25FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis zee Biplane, Model No. 10094141; Artis zee Ceiling, Model No. 10094137; Artis zee Floor, Model No. 10094135; Artis zeego, Model No. 10280959 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
A software issue could potentially cause the stand and table movements to be blocked.
- 2020-03-25FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis Q.zen Biplane, Model No. 10848355 Artis Q.zen Ceiling, Model No. 10848354 Artis Q.zen Floor, Model No. 10848353 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
A software issue could potentially cause the stand and table movements to be blocked.
- 2020-03-25FDA-DeviceSiemens Medical Solutions USA, IncClass IIArtis Q Biplane, Model No. 10848282; Artis Q Ceiling, Model No.10848281; Artis Q Floor, Model No. 10848280; Artis Q zeego, Model No. 10848283 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
A software issue could potentially cause the stand and table movements to be blocked.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IICardio Vascular-Allura Centron, system code 722400
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIAllura CV20, system code 722031
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIUroDiagnost Eleva, system code 708033
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIPoly G Integris H5000, System code 72246
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIOmniDiagnost Eleva System codes 708028 708027
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIMultiDiagnost Eleva, System codes 722030 722016 722017 708032 708036 708034 708037 708038
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIIntegris Allura 9, system codes 722018 722021
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIIntegris Allura 15 & 12, monoplane (system code 722043) and biplane (system code 722044)
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DevicePhilips North America, LLCClass IIAllura Exper FD Series. System codes 722003 722010 722022 722005 722011 722019 722001 722002 722006 722012 722008 722013 722020 722025 722015 722023 722038 889006 889016 889009 722005 The device is an X-Ray Generator.
A capacitor inside the converter of the Velara X-ray generator may fail after a large number of surges in a short period of time. When this occurs, no image acquisition is possible.
- 2020-03-25FDA-DeviceMaterialise N.V.Class IITRUMATCH Orthognathics - Genioplasty Surgical Kit, Add-on, Catalog Number: SD980.013
Custom surgical kits contain a plate different than indicated by the package labeling.
- 2020-03-25FDA-DeviceMaterialise N.V.Class IITRUMATCH Orthognathic - Full Bimaxillary Surgical Kit, Catalog Number SD980.001
Custom surgical kits contain a plate different than indicated by the package labeling.
- 2020-03-25FDA-DeviceDeroyal Industries, Inc. LafolletteClass IITeleflex Pilling Aortic Punch, 4.0mm. Sterile, Rx Only, disposable. Included as part of Intl-Cardiac Surgery Kit - Product Usage: This device is used for cardiovascular surgical procedures.
The sterile packaging has the potential to be compromised.
- 2020-03-25FDA-DeviceAesculap Implant Systems LLCClass IICOLUMBUS REV F TIBIA OFFSET CEMENTED, Knee implant components, various sizes, model nos. NR068K NR068Z NR070K NR070Z NR071K NR071Z NR072K NR072Z NR073K NR073Z NR074K NR074Z NR075K NR075Z NR076K NR076Z NR077K NR077Z NR078K NR078Z NR079K NR079Z Product Usage: Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
- 2020-03-25FDA-DeviceAesculap Implant Systems LLCClass IIAS COLUMBUS CR/PS TIB.PLAT.CEMENTED,Knee implant components, various sizes, model nos. NN058Z NN070Z NN071Z NN072Z NN073Z NN074Z NN075Z NN076Z NN077Z NN078Z NN079Z NN470Z NN471Z NN472Z NN473Z NN474Z NN475Z NN476Z NN477Z NN478Z NN479Z Product Usage: Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
- 2020-03-25FDA-DeviceAesculap Implant Systems LLCClass IICOLUMBUS CRA/PSA TIB.PLAT.CEMENTED, Knee implant components, various sizes, model nos. NN469K NN470K NN471K NN472K NN473K NN474K NN475K NN476K NN477K NN478K NN479K Product Usage: Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
- 2020-03-25FDA-DeviceAesculap Implant Systems LLCClass IICOLUMBUS CR/PS TIB.PLAT.PLASMAPORE, Knee implant components, various sizes, model nos. NN081K NN082K NN083K NN084K NN085K NN086K NN087K NN088K NN089K Product Usage: Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
- 2020-03-25FDA-DeviceAesculap Implant Systems LLCClass IIColumbus R/PS TIB.PLATEAU Cemented, various sizes, Knee implant components, model nos. NN058K NN070K NN071K NN072K NN073K NN074K NN075K NN076K NN077K NN078K NN079K Product Usage:Tibia-Plateaus are part of implants in the group of knee endoprosthetics. For these implant parts there are cemented and cement-free versions as well as versions with zirconium nitride multi layer.
Low Density Polyethylene (LDPE) particles may be present on the surface of the implant upon the opening of the inner packaging.
- 2020-03-25FDA-DeviceCME America, LLCClass ICME America BodyGuard Infusion Pump System, All models of BG 323, BG 121, BG 545, BG CV545, BG 575, BG CV575, and CMExpress
Infusion Pump Systems may have a delivery inaccuracy of up to 13%, which may result in 1) faster than expected drug delivery when infusing at a very low rate (0.1 mL/h), or 2) slower than expected drug delivery when infusing at high flow rates (greater-than 500 mL/h)
- 2020-03-18FDA-DeviceCordis CorporationClass IICordis POWERFLEX P3, PTA Dilation Catheter, REF 420-6020S
Cordis is recalling one (1) lot of POWERFLEX P3 PTA Dilatation Catheter due to the potential for body/shaft voids in the proximal seal area.
- 2020-03-18FDA-DeviceQiagen Sciences LLCClass IIQIAGEN Rotor Gene Q Software version 2.3.4. Real-time PCR analyzer designed for rapid thermalcycling and real-time detection of PCR assays.
When using the Rotor-Gene Q with Software version 2.3.4 completing a LIMs export, the .csv file reports the calculated concentration result value as a logarithmic value, and could lead to a false negative result, which could lead to serious medical consequences such as the suspension or non-initiation of treatment
- 2020-03-18FDA-DeviceMaquet Cardiovascular Us Sales, LlcClass IIGetinge 88-Series-washer disinfectors intended for cleaning, disinfection and drying Model Number: 88-5 Part Number: 88-103; 88-203
Potential leakage that may occur due to ruptured spot welds in the chamber, could generate a situation for the user such as risk for slipping or infection from contaminated water