Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2019-09-18FDA-DeviceRandox Laboratories Ltd.Class IIIThis is an in vitro diagnostic product intended for use as a calibration serum in clinical chemistry assays. Randox calibration sera contain 42 analytes and are based on lyophilized human serum. The concentrations and activities are suitable for calibration of clinical chemistry assays both manually and on a wide range of automatic analyzers. Constituent concentrations are available at 2 levels.
Randox Laboratories realigned RX Analyser Series calibration targets by 10% for AST and 9% for ALT in line with the Mean of All Instrument method target for the Randox Calibration Serum Level 3 (for calibrator lot 961UE onwards).
- 2019-09-18FDA-DevicePhilips North America, LLCClass IIHeartStart MRx Monitor/Defibrillator Model # M3535A - Product Usage: The HeartStart MRx is a lightweight, portable external defibrillator, offering two modes of operation for defibrillation: manual mode and semi-automatic mode (AED). In manual mode, the HeartStart MRx is a full featured manual defibrillator, designed for use by clinicians trained in Advanced Cardiac Life Support (ACLS). Manual operation allows users to select energy levels for external and internal defibrillation. Users may als
Non-conforming devices are identified, which may not have been included in prior field actions and may not have had a number of corrections applicable to these devices.
- 2019-09-18FDA-DeviceInstrumentation Laboratory Co.Class IIROTEM ex-tem Rest of the world part No: 503-05 (ROW)
Prolonged clotting times
- 2019-09-18FDA-DeviceInstrumentation Laboratory Co.Class IIROTEM ex-tem Canada Part No: 503-05-CA
Prolonged clotting times
- 2019-09-18FDA-DeviceInstrumentation Laboratory Co.Class IIROTEM ex-tem; US Part No: 503-05-US
Prolonged clotting times
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle 150 Full Height and Short Height Implants, BIOCELL, INTRASHIEL, BioDIMENSIONAL Breast Implant, Styles 150 FH and 150 SH. Product Usage: Breast augmentation and Breast reconstruction
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass IRITZ PRINCESS, BIOCELL, INTRASHIEL, BioDIMENSIONAL Breast Implant, Styles RML, RMM, RFL, RFM Product Usage: Breast augmentation and Breast reconstruction
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass IKomuro, Soft Touch, BIOCELL, INTRASHIEL, BioDIMENSIONAL breast implants, Styles KML, KMM, KLL, KLM Product Usage: Breast augmentation and Breast reconstruction
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle and McGhan Silicone-Filled Breast Implants, TruForm 2 (Soft Touch), BIOCELL, Styles 110 Soft Touch and 120 Soft Touch Product Usage : Breast augmentation and Breast reconstruction
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle 510, TruForm, Dual-Gel, BIOCELL, Styles LX, MX, FX Product Usage: Breast augmentation and Breast reconstruction
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle and McGhan 410 TruForm 2 (Soft Touch) BIOCELL breast implant, Styles LL, LM, LF, ML, MM, MF, FL, FM, FF Product Usage: Breast augmentation and Breast reconstruction
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass IMcGhan BioDIMENSIONAL silicone-Filled Breast Implants, style 153 Product Usage: Breast augmentation and Breast reconstruction
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass IMcGhan Croissant-Shaped Tissue Expanders, Style 134 Product Usage: intended for temporary subcutaneous implantation and require periodic, incremental inflation with sterile saline for injection until the desired amount of tissue is developed.
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle 133 Tissue Expanders w/Suture Tabs BIOCELL Full Height Variable Projection with MAGNA-FINDEr Xact & 21G Needle Infusion Set STYLE 133FV-T: 133FV-11-T, 133FV-12-T, 133FV-13-T, 133FV-14-T, 133FV-15-T, 133FV-16-T; STYLE 133MV-T: 133MV-11-T, 133MV-12-T, 133MV-13-T, 133MV-14-T, 133MV-15-T, 133MV-16-T; STYLE 133LV-T: 133LV-11-T, 133LV-12-T, 133LV-13-T, 133LV-14-T, 133LV-15-T, 133LV-16-T; STYLE 133SV-T: 133SV-11-T, 133SV-12-T, 133SV-13-T, 133SV-14-T, 133SV-15-T,
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle133 Plus Tissue Expander With/Without Suture Tabs, BioCell Full Height Variable Projection with Magna-Fider Xact & 21G Needle Infusion Set, Sterile, for the following style numbers: STYLE 133P-FV: 133P-FV-11, 133P-FV-12, 133P-FV-13, 133P-FV-14, 133P-FV-15, 133P-FV-16; STYLE 133P-MV: 133P-MV-11, 133P-MV-12, 133P-MV-13, 133P-MV-14, 133P-MV-15, 133P-MV-16; STYLE 133P-LV: 133P-LV-11, 133P-LV-12, 133P-LV-13, 133P-LV-14, 133P-LV-15, 133P-LV-16; STYLE 133P-FX: 133P-FX-11,
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle 133 Tissue Expanders (also named style 133 BIOSPAN Tissue Expander and Mcghan Magna-Site Tissue Expander) BioCell Full Height Variable Projection with Magna-Fider Xact & 21G Needle Infusion Set, Sterile, for the following style numbers: STYLE 133FV: 133FV-11, 133FV-12, 133FV-13, 133FV-14, 133FV-15, 133FV-16; STYLE 133MV: 133MV-11, 133MV-12, 133MV-13, 133MV-14, 133MV-15, 133MV-16; STYLE 133LV: 133LV-11, 133LV-12, 133LV-13, 133LV-14, 133LV-15, 133LV-16; S
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle and McGhan Silicone-Filled Breast Implant BioCell Round Moderate Profile, Sterile, for the following style numbers: Style 110:110-090, 110-120, 110-150, 110-180, 110-210, 110-240, 110-270, 110-300, 110-330, 110-360, 110-390, 110-420, 110-450, 110-480, 110-510; Style 115 :115-150, 115-167, 115-185, 115-203, 115-222, 115-253, 115-272, 115-290, 115-322, 115-354, 115-378, 115-401, 115-435, 115-469, 115-507, 115-547, 115-586, 115-627, 115-666, 115-716;
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle Inspira, Silicone-Filled Breast Implants, Sterile, for the following style numbers: Style TRL: TRL-110TRL-125,TRL-140,TRL-170,TRL-200,TRL-230,TRL-260,TRL-290,TRL-320,TRL-350,TRL-380,TRL-410,TRL-460,TRL-510,TRL-560,TRL-610; Style TRLP: TRLP-125TRLP-145,TRLP-165,TRLP-190,TRLP-205,TRLP-220,TRLP-235,TRLP-250, TRLP-265,TRLP-280,TRLP-300,TRLP-320,TRLP-340,TRLP-360,TRLP-400,TRLP-440,TRLP-490,TRLP-540,TRLP-590,TRLP-640; Style TRM: TRM-140,TRM-175,TRM-195,TRM-210,TRM-240,TRM-255,TRM-275,
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle 410, Highly Cohesive Anatomically Shaped Silicone-Filled Breast Implant Biocell Low Height Low Projection, Sterile, for the following style numbers: Syle LL: LL-410135,LL-410180,LL-410210,LL-410240,LL-410300; Style LM: LM-410140, LM-410190, LM-410220, LM-410250, LM-410320; Style LF: LF-410125, LF-410150,LF-410175,LF-410205,LF-410240,LF-410270,LF-410310,LF-410390,LF-410440,LF-410490,LF-410540,LF-410595,LL-410135,LL-410180,LL-410210,LL-410240,LL-410300; style LX: LX-410145,LX-
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceAllergan PLCClass INatrelle, Saline-Filled Breast Implant Biocell Full Height with Fill Tube, Sterile, for the following style numbers: Style 163:163-360CC, 163-440CC, 163-530CC, 163-655CC, 163-780CC; Style 168:168-120,168-150,168-180,168-210,168-240,168-270,168-300,168-330,168-360,168-390,168-420,168-450,168-480,168-510,168-550,168-600, 168-650,168-700,168-750, 168-800; Style 363: 363LF-170, 363LF-200, 363LF-230, 363LF-260, 363LF-300, 363LF-330, 363LF-370, 363LF-410, 363LF-450, 363LF-510, 363LF-560, 363
The action was initiated following notification by the U.S. FDA of their recently updated, global safety information concerning the higher incidence of anaplastic large cell lymphoma (BIA-ALCL) in patients who have textured breast implants.
- 2019-09-18FDA-DeviceTeleflex MedicalClass ISheridan/HVT Novaplus: a) 7.0 mm, REF V5-10314 b) 7.5 mm, REF V5-10315 c) 8.0 mm, REF V5-10316 d) 8.5 mm, REF V5-10317 Product Usage: Tracheal tube/airway management
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- 2019-09-18FDA-DeviceTeleflex MedicalClass ISheridan/CF Novaplus 7.0 mm, REF V5-10114 Product Usage: Tracheal tube/airway management
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- 2019-09-18FDA-DeviceTeleflex MedicalClass IHudson RCI Sheridan/HVT: a) 5.0 mm, REF 5-10310 b) 5.5 mm, REF 5-10311 c) 6.0 mm, REF 5-10312 d) 6.5 mm, REF 5-10313 e) 7.0 mm, REF 5-10314 f) 7.5 mm, REF 5-10315 g) 8.0 mm, REF 5-10316 h) 8.5 mm, REF 5-10317 i) 9.0 mm, REF 5-10318 j) 10.0 mm, REF 5-10320 Product Usage: Tracheal tube/airway management
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- 2019-09-18FDA-DeviceTeleflex MedicalClass IHudson RCI Sheridan Uncuffed: a) 2.0 mm, REF 5-10404 b) 2.5 mm, REF 5-10405 c) 3.0 mm, REF 5-10406 d) 3.5 mm, REF 5-10407 e) 4.0 mm, REF 5-10408 f) 4.5 mm, REF 5-10409 g) 5.0 mm, REF 5-10410 h) 5.5 mm, REF 5-10411 i) 6.0 mm, REF 5-10412 j) 6.5 mm, REF 5-10413 k) 7.0 mm, REF 5-10414 Product Usage: Tracheal tube/airway management
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- 2019-09-18FDA-DeviceTeleflex MedicalClass IHudson RCI Sheridan Preformed: a) 3.0 mm, REF 5-22006, REF 5-22106 b) 3.5 mm, REF 5-22007, REF 5-22107 c) 4.0 mm, REF 5-22008, REF 5-22108, REF 5-22208 d) 4.5 mm, REF 5-22009, REF 5-22109, REF 5-22209 e) 5.0 mm, REF 5-22010, REF 5-22110, REF 5-22210 f) 5.5 mm, REF 5-22011, REF 5-22111, REF 5-22211 g) 6.0 mm, REF 5-22212, REF 5-22312, REF 5-22112 h) 6.5 mm, REF 5-22313, REF 5-22213, REF 5-22013, REF 5-22113 i) 7.0 mm, REF 5-22214, REF 5-22314, REF 5-22014, REF 5-22114 j) 7.5 mm, REF 5-2
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- 2019-09-18FDA-DeviceTeleflex MedicalClass IHudson RCI Sheridan EZ-ENDO: a) 5.0 mm, REF 5-22510 b) 5.5 mm, REF 5-22511 c) 7.0 mm, REF 5-22514 d) 7.5 mm, REF 5-22515 e) 8.0 mm, REF 5-22516 f) 8.5 mm, REF 5-22517 g) 9.0 mm, REF 5-22518 h) 6.0 mm, REF 5-22512 i) 6.5 mm, REF 5-22513 Product Usage: Tracheal tube/airway management
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- 2019-09-18FDA-DeviceTeleflex MedicalClass IHudson RCI Sheridan/CF: a) 3.0 mm, REF 5-10106, REF 5-10206 b) 3.5 mm, REF 5-10107 c) 4.0 mm, REF 5-10108, REF 5-10208 d) 4.5 mm, REF 5-10109 e) 5.0 mm, REF 5-10110, REF 5-10210 f) 5.5 mm, REF 5-10111 g) 6.0 mm, REF 5-10112, REF 5-10212 h) 6.5 mm, REF 5-10113, REF 5-10213 I) 7.0 mm, REF 5-10114, REF 5-10214 j) 7.5 mm, REF V5-10115, REF 5-10115, REF 5-10215 k) 8.0 mm, REF 5-10116, REF 5-10216, REF V5-10116 l) 8.5 mm, REF 5-10117, REF 5-10217 m) 9.0 mm, REF 5-10118, REF5-10218 n) 9.
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- 2019-09-18FDA-DeviceTeleflex MedicalClass IHudson RCI Sheridan LTS: a) 4.0 mm, REF 5-11108 b) 5.0 mm, REF 5-11110 c) 6.0 mm, REF 5-11112 Product Usage: Tracheal tube/airway management
Reported complaints indicate an increased incidence of specific lots of the 15 mm Sheridan connector becoming disconnected from the endotracheal tube (ETT). Disconnection from the breathing circuit may result in insufficient oxygenation, requiring medical intervention.
- 2019-09-11FDA-DeviceHaag-Streit USA IncClass IIHS Hi-R NEO 900A NIR Reference # 657 822 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of opera
software error in the central control unit of the floor stand.
- 2019-09-11FDA-DeviceHaag-Streit USA IncClass IIHS Hi-R NEO 900A Reference # 657 821 Serial # from 101 Product Usage: A wide range of surgical interventions are made possible through the modular construction and the diversity of the equipment and accessories. The operating microscope HS Hi-R NEO 900/HS Hi-R NEO 900A/HS Hi-R NEO 900A NIR and HS ALLEGRA 900/590/90 are designed for use in Ophthalmology and in other disciplines where a vertical line of viewing is required. Due to their high maneuverability, the HS ALLEGRA range of operating
software error in the central control unit of the floor stand.