Medical device recalls
39180 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2019-07-17FDA-DeviceHelena Laboratories, Corp.Class IIColoCARE, Cat. No. 5650, packaged 250 envelopes/box, For In-Vitro Diagnostic Use. The firm name on the label is Helena Laboratories, Beaumont, TX.
The positive control on some tests of the lots did not appropriately react within the specified timeframe to verify the test was functioning correctly.
- 2019-07-17FDA-DevicePhilips North America, LLCClass IIPhilips HeartStart MRx devices in use with the M3539A AC Power Module: M3535A, M3536A, M3536M, M3536MC, M3536M2, M3536M3, M3536M4, M3536M5, M3536M6, M3536M7, M3536M8, M3536M9
The AC power module may fail at a higher than expected rate. If a fully charged battery is not properly installed, the HeartStart MRx may lose all power and fail to operate. Interruption of monitoring and failure or delay in the delivery of a shock or pacing can occur. A faulty AC Power Module may also fail to charge the battery.
- 2019-07-17FDA-DeviceHamilton Medical AGClass IHAMILTON-G5, with software versions less than or equal to 2.60 The HAMILTON-G5/S1 ventilators are designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients. The device is intended for use in the hospital and institutional environment where health care professionals provide patient care.
New software version for affected ventilators reduces the probability of the ventilator entering an ambient state, in which the inspiratory channel and expiratory valves are opened, letting the patient breathe room air unassisted. When the ventilator enters the ambient state, alternative ventilation must be provided immediately.
- 2019-07-17FDA-DeviceGE Healthcare, LLCClass IPanda Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
- 2019-07-17FDA-DeviceGE Healthcare, LLCClass IGiraffe Infant Radiant Warmer Infant radiant warmers provide infrared heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Infant radiant warmers may be used to facilitate the neonate's transition to the external environment or to provide a controlled open environment.
The bedside panels and latch areas on the Giraffe Bedded Warmer and Panda iRes Warmer can crack or break if the unit is moved using the bedside panels instead of the front handle or the maneuvering handle.
- 2019-07-17FDA-DeviceBecton Dickinson & CompanyClass IAlaris Pump Infusion Set, Model Codes 10010453, 10010454, 10012144, 10015294, 10015414, 10933805, 11171447, 11404930, 11522558, 11607704, 10013361T, 2110-0500, 2126-0500, 2200-0500, 2411-0500, 2419-0007, 2420-0007, 2420-0500, 2421-0500, 2426-0007, 2426-0500, 2429-0500, 2433-0007, 2477-0000, 2477-0007 2478-0000 C24116E Infusion sets allow for the intravenous administration of fluids, medications, nutritional support and transfusion therapy through a needle or catheter inserted into the patient
An incomplete occlusion can be caused by a variation in the wall thickness of the pumping segment of the affected infusion sets used with the. The issue has the potential to lead to unintended delivery of medication when the pump module is not in running status or can result in faster than expected drug delivery flow when the pump is infusing.
- 2019-07-10FDA-DeviceZimmer Biomet, Inc.Class IIK-Wire Trochar Tip, Model Number 212000008 Product Usage: These DVR Express Disposable Instruments are used to facilitate implantation of orthopedic medical devices in trauma.
There is a potential for weak seals of the sterile packaging.
- 2019-07-10FDA-DeviceZimmer Biomet, Inc.Class IIDVR Crosslock ePAK Depth Gauge, Model Number 212000003 Product Usage: These DVR Express Disposable Instruments are used to facilitate implantation of orthopedic medical devices in trauma.
There is a potential for weak seals of the sterile packaging.
- 2019-07-10FDA-DeviceZimmer Biomet, Inc.Class IIDVR Crosslock ePAK Screw Driver, Model Number 212000002 Product Usage: These DVR Express Disposable Instruments are used to facilitate implantation of orthopedic medical devices in trauma.
There is a potential for weak seals of the sterile packaging.
- 2019-07-10FDA-DeviceBio-Detek, Inc.Class IIZOLL Pro-padz Liquid Gel Radiolucent, Part numbers: 8900-2105-01 (12/case) and 8900-2106-01 (single)
Some of the electrodes may have been assembled incorrectly, preventing delivery of defibrillation or pacing therapy.
- 2019-07-10FDA-DeviceCardiac Science CorporationClass IIPowerheart¿ G5 Automatic AED
Sub-supplier s documentation for specific Printed Circuit Boards (PCB) is insufficient to determine if the devices conform to specifications.
- 2019-07-10FDA-DeviceBeckman Coulter Inc.Class IIiChem VELOCITY Urine Chemistry Strips, REF 800-7212, 100 test strips/vial. The firm name on the label is Beckman Coulter Ireland, Inc., Co. Clare, Ireland, Made in Germany.
Falsely elevated (false positive) urobilinogen patient results due to an increase in the sensitivity of the urobilinogen reaction.
- 2019-07-10FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS 350 Chemistry System, Product Code 6802153, (UDI # 10758750002054) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
- 2019-07-10FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS 250 Chemistry System Refurbished, Product Code 6801759, (UDI # 10758750001330) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
- 2019-07-10FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS 250AT Chemistry System, Product Code 1758143, (UDI # 10758750000036) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
- 2019-07-10FDA-DeviceOrtho-Clinical DiagnosticsClass IIVITROS 250 Chemistry System, Product Code 8132086, (UDI # 10758750004409) Product Usage: Performs discrete clinical tests on serum, urine, and cerebral spinal fluid specimens. Methodologies include colorimetric (CM), potentiometric (PM), rate (RT), and immuno-rate (IR) tests using multi-layered VITROS Slides.
Potential negative drift in Creatine Kinase (CK) results when used on the VITROS 250 and 350 Systems.
- 2019-07-10FDA-DeviceMedtronic Navigation, Inc.-LittletonClass IIMedtronic 0-arm TM 1000 Imaging Systems: Bl-700-00027-XXX and B1-700-00028-XXX (where XXX indicates the system voltage): Product Name Model#/ CFN OARM ASSY 8170000027 SYSTEM PRODUCT BI70000027 OARM ASSY BI70000027R SYS PRODUCT RWK BI70000027R BASE OARM 8170000027100 SYSTEM !DOV BI70000027100 BASE OARM Bl70000027100R SYSTEM lOOV RWK Bl70000027100R BASE OARM 8170000027120 SYSTEM 120V 8170000027120 BASE OARM Bl70000027120R SYSTEM 120V RWK BI70000027120R BASE OARM Bl70000027230 SYSTEM 230V 8
The 0-arm 1000 Imaging System uses energy from the batteries to generate X-rays and move the system. If a charger board is not functioning properly, the batteries will not receive a full recharge of the potential capacity and results in the inability of the system to take X-rays, open the gantry, and move the system from one place to another, which may affect the ability to continue use within the operating room until the charger board and/or the batteries are replaced
- 2019-07-10FDA-DeviceZOLL Medical CorporationClass IIZOLL AED PRO Automated External Defibrillator Product Usage: The AED PRO unit is intended to defibrillate victims of ventricular fibrillation or pulseless ventricular tachycardia, for ECG monitoring, and for CPR monitoring of patients. The CPR monitoring function provides a metronome designed to encourage rescuers to perform chest compressions at the AHA/ERC recommended rate of 100 compressions per minute. Voice and visual prompts encourage a compression depth of at least 2 inches (5 cm) for
There is a potential for the device to fail to deliver energy to the patient.
- 2019-07-10FDA-DeviceRAYSEARCH LABORATORIES ABClass IIRayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6) . Intended Use: software system designed for treatment planning and analysis of radiation therapy
When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong
- 2019-07-10FDA-DevicePhilips Medical Systems (Cleveland) IncClass IIVereos PET/CT, Model No. 882446 Product Usage: Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem images anatomical cross-sections by computer reconstruction of X-ray transmission data. The PET subsystem images the distribution of PET anatomy-specific radiopharmaceuticals in the patient. The PET/CT system is used for the purposes of detecting, localizing, diagnosing, staging, re-staging, and follow-up for monitoring therapy response of various dis
The Vereos PET/CT may have misaligned front and rear covers, which may result in gaps between cover parts and/or protrusion of the cover into the bore diameter.
- 2019-07-10FDA-DeviceRoche Diagnostics CorporationClass IIIcobas p 701 post-analytical units, Part Number 05349966001. For diagnostic sample storage and retrieval.
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
- 2019-07-10FDA-DeviceRoche Diagnostics CorporationClass IIIcobas p 501 post-analytical units, Part Number 05158494001. For diagnostic sample storage and retrieval.
There is a potential for the tray input flap on the post analytical units to become loose, potentially detaching from the instrument.
- 2019-07-10FDA-DeviceInfinium Medical, IncClass IICLEO Patient Monitor /Tranquility VS Patient Monitor
The CLEO & Tranquility VS Patient Monitors 510(k) indications for use in the United States did not include the temperature monitor accessory option.
- 2019-07-10FDA-DeviceMadison Polymeric EngineeringClass IIFirst Step Draco Pad, Product Number EP-4D
This product has been contaminated with Stenotrophomonas lactiubi.
- 2019-07-10FDA-DeviceMadison Polymeric EngineeringClass IIDraco Enzymatic Deep-Cleaning Pad with Ready-to-Use Enzymatic Detergent, Product Number HY0305Z
This product has been contaminated with Stenotrophomonas lactiubi.
- 2019-07-10FDA-DeviceSiemens Medical Solutions USA, IncClass IISOMATOM go.Top, Model No. 11061648 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy pl
The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.
- 2019-07-10FDA-DeviceSiemens Medical Solutions USA, IncClass IISOMATOM go.All, Model No. 11061638 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy pl
The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.
- 2019-07-10FDA-DeviceSiemens Medical Solutions USA, IncClass IISOMATOM go.Now, Model No. 11061618 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy pl
The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.
- 2019-07-10FDA-DeviceSiemens Medical Solutions USA, IncClass IISOMATOM go.Up, Model No. 11061628 with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy pla
The potential sporadic performance problems may cause scanning workflow interruptions and unexpected user notifications resulting in diagnostic delay or need for patient rescan.
- 2019-07-10FDA-DeviceSiemens Medical Solutions USA, IncClass IISOMATOM go.Top (Model #1161640), with syngo.CT software versions VA20A, VA20A_SP0, VA20_SP1. Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data. The images delivered by the system can be used by a trained physician as an aid in diagnosis. The images delivered by the system can be used by trained staff as an aid in diagnosis, treatment preparation and radiation therapy plannin
The potential sporadic performance problems may result in scanning workflow interruptions and unexpected user notifications. This may cause a delay in diagnosis or patient scans.