- 2019-04-17FDA-DeviceRoche Diagnostics CorporationClass IICOBAS INTEGRA 400 plus analyzer used in the following assays: a)Tina-quant Apolipoprotein A-1 ver.2 Catalog Number: 03032566122 b)Tina-quant Albumin Gen.2 Catalog Number 04469658190 c)Creatinine plus ver.2 Catalog Number 03263991190 d) riglycerides Catalog Number :20767107322 - Product Usage: Tina-quant Apolipoprotein A-1 ver.2 A lipoprotein test system is a device intended to measure lipoprotein in serum and plasma. Lipoprotein measurements are used in the diagnosis and treatmen
COBAS INTEGRA¿ 400 plus and cobas c 111 analyzers Additional Extra Wash Cycle Configurations and/or Batch Mode Needed for Specific Tests Due to Carry-over
- 2019-04-17FDA-DeviceEncore Medical, LpClass IIExprt Precision System: Revision Hip, Proximal Body with Bolt, Lateral Offset, Sterile R, djo surgical, REF: 495-00-065, 495-00-075, 495-00-085, 495-01-065, 495-01-075, 495-01-085
Complaints regarding loose proximal body bolts or bolts that were found through x-ray as sitting above the proximal body and not properly seated. Patient risks may include: pain, dissociation, and possible revision surgery
- 2019-04-17FDA-DeviceBecton Dickinson & CompanyClass IIBD Blood Collection Assembly with Male Luer Lock, Catalog Number MBC6010
The luer could break and cause the device to leak or break off and get stuck in the fistula needle port rendering the port inaccessible for dialysis. As a result, the patient would need to be re-cannulated with a new fistula needle to obtain their dialysis treatment.
- 2019-04-17FDA-DeviceBecton Dickinson & CompanyClass IIGravity IV Set with 3-port Closed Stopcock Manifold Model CM42500E-07
The firm has initiated a recall due to a confirmed complaint trend for kinked tubing on the specified lots of the product.
- 2019-04-17FDA-DeviceBecton Dickinson & CompanyClass IIGravity IV Set with 3-port Closed Stopcock Manifold Model CS42522E-07
The firm has initiated a recall due to a confirmed complaint trend for kinked tubing on the specified lots of the product.
- 2019-04-17FDA-DeviceArrow International IncClass IIArrow¿ Two-Lumen Central Venous Catheterization Set with Blue FlexTip¿ Catheter Product Code: CD-10802 GTN: 00801902015396
The product lidstock contains a labelling error. The lidstock states the incorrect priming volume. The priming volumes printed on the extension lines are correct
- 2019-04-17FDA-DeviceArrow International IncClass IIArrow¿ Two-Lumen Central Venous Catheterization Kit with Blue FlexTip¿ Catheter Product Code: AH-11802 (OUS) GTN: 00801902026132
The product lidstock contains a labelling error. The lidstock states the incorrect priming volume. The priming volumes printed on the extension lines are correct
- 2019-04-17FDA-DeviceVygon U.S.A.Class IIVyon DUALSTOP White Cap used for intravascular lines Product Code: 9888.002- Product Usage: Cap used for intravascular lines.
DualStop Caps defect causing leaks in affected product
- 2019-04-17FDA-DeviceVygon U.S.A.Class IIVygon DUALSTOP Red Cap used for intravascular lines Product Code: 9888.00 - Product Usage: Cap used for intravascular lines.
DualStop Caps defect causing leaks in affected product
- 2019-04-17FDA-DeviceSiemens Medical Solutions USA, IncClass IIsyngo.via syngo.CT Cardiac Planning, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software packages for evaluating cardiac CT angiography (CTA) volume sets. The software packages are designed to support the physician in determining the functional parameters of the left and right ventricles, confirming the presence or absence of physicianidentified myocardial enhancement defects and evaluation documentation and follow-up of a
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
- 2019-04-17FDA-DeviceSiemens Medical Solutions USA, IncClass IIsyngo.via syngo.CT Cardiac Function, Model Number 10496180 Product Usage: Syngo.CT Cardiac Planning and syngo CT. Cardiac Function are image analysis software packages for evaluating cardiac CT angiography (CTA) volume sets. The software packages are designed to support the physician in determining the functional parameters of the left and right ventricles, confirming the presence or absence of physicianidentified myocardial enhancement defects and evaluation documentation and follow-up of a
There is a potential risk of a wrong measurement in the annulus plane during a TAVI planning procedure. This risk is due to a software issue found in the TAVI algorithm.
- 2019-04-17FDA-DeviceBeaver VisitecClass IIBeaver Xstar Slit Knife 2.4 mm, 45degree Double Bevel, Ophthalmic-utilized in Cataract and Glaucoma procedures REF # 370824 Product Usage: Ophthalmic techniques-utilized in Cataract and Glaucoma procedures
Mislabeled. The package labeling indicates a size 2.4mm, 45 degree Double Bevel knife however, the marking on device shows a 2.75mm size device
- 2019-04-17FDA-DevicePROCEPT BIOROBOTICS CORPORATIONClass IIAquaBeam Handpiece, Sterile EO, Rx Only, REF: 310301, and 320301, a component of AquaBeam Robotic System. 320301 distributed outside the U.S. Device Description The AQUABEAM Robotic System is intended for use in patients suffering from lower urinary tract symptoms resulting from benign prostatic hyperplasia (BPH).
Certain lots may develop an internal leak due to a micro-crack in a specific weld joint within the device. When the Robot detects this failure, it immediately renders the device inoperable and notifies the user via an error message, which may cause delay and/or inability to complete the procedure.
- 2019-04-17FDA-DeviceElekta LimitedClass IIElekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.
Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
- 2019-04-17FDA-DeviceConformis, Inc.Class IIiTotal CR Total Knee Replacement System, TCR-121-1111 (Catalog No.) / TCR-121-1111-010102 (Model No.) M5725INT0600010 (Model No.) Product Usage: The iTotal CR Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral, or bicompartmental prosthesis.
The kits were prepared with incorrect patient ID labels and may contain incorrect components.
- 2019-04-17FDA-DeviceRandox Laboratories Ltd.Class IIILiquid Immunoassay Premium Controls, Model Number LIA3105 Product Usage: This product is intended for in vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
- 2019-04-17FDA-DeviceRandox Laboratories Ltd.Class IIIImmunoassay Premium Plus Controls, Model Numbers IA3109 and IA3112 Product Usage: This product is intended for in vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
- 2019-04-17FDA-DeviceRandox Laboratories Ltd.Class IIIImmunoassay Premium Controls, Model Numbers IA2633 and IA2638 Product Usage: This product is intended for in vitro diagnostic use, as assayed quality control material to monitor the accuracy and reproducibility of analytes listed in the package insert
The target value for uptake is incorrect on the value sheet for an analyzer. The target value for this analyzer is listed as 43.8%U and should be 39.4%U.
- 2019-04-17FDA-DeviceAbbott Laboratories, IncClass IIAbbott ARCHITECT c16000 Processing Module, List No. 03L77 - Product Usage: The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The ARCHITECT c4000, c8000, c16000 sample pipettor assemblies include a fluid sense/pressure monitoring system that helps identify errors in aspiration.
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
- 2019-04-17FDA-DeviceAbbott Laboratories, IncClass IIAbbott ARCHITECT c8000 Processing Module, List No. 01G06 - Product Usage: The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The ARCHITECT c4000, c8000, c16000 sample pipettor assemblies include a fluid sense/pressure monitoring system that helps identify errors in aspiration.
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
- 2019-04-17FDA-DeviceAbbott Laboratories, IncClass IIAbbott ARCHITECT c4000 Processing Module, List No. 02P24 - Product Usage: The Abbott ARCHITECT System is intended for In Vitro diagnostic use only. The Abbott ARCHITECT Clinical Chemistry Systems are designed to perform automated chemistry tests, utilizing photometry and potentiometric technology. The ARCHITECT c4000, c8000, c16000 sample pipettor assemblies include a fluid sense/pressure monitoring system that helps identify errors in aspiration.
There is a potential to generate incorrect results on the instrument if particular error codes are not resolved before transitioning to the Running status.
- 2019-04-17FDA-DeviceEdwards Lifesciences, LLCClass IIEdwards Lifesciences(R) IntraClude(TM) Intra-Aortic Occlusion Device is a 10.5 Fr (3.5mm), triple lumen, 100cm long catheter with an elastomeric balloon near its distal tip designed to occlude the ascending aorta in order to partition the aortic root from arterial circulation. The balloon expands to occlude a range of aorta sizes from 20 to 40 mm. The large central lumen of the IntraClude Device serves three functions: accommodating the guidewire, delivering cardioplegia solution to th
Edwards Lifesciences has received a limited number of customer reports regarding an inter-lumen leak between the cardioplegia lumen and pressure lumen in the device.
- 2019-04-17FDA-DeviceLivaNova USA Inc.Class IISORIN Heater-Cooler System 3T, REF 16-02-85 Intended for use with a Stockert/Sorin S3/S5/C5 heart-lung machine and/or any other heart-lung machine featuring a separate temperature control for extracorporeal perfusion of durations of up to 6 hours.
For six units, the post-deep cleaning (DC) water sample tested positive for nontuberculous mycobacteria (NTM).
- 2019-04-17FDA-DeviceIntegra LimitedClass IIIntegra bioBLOCK Resorbable subtaler Implant, 12mm Catalog Number: 040012 - Product Usage: The bioBLOCK Implant is indicated for internal support to primary surgical interventions in the treatment of flatfoot.
Temperature indicator may have changed color to dark gray or black prior to shipment. Change in color may indicate the product exposure to temperatures greater than 110¿F
- 2019-04-17FDA-DeviceIntegra LimitedClass IIIntegra bioBLOCK Resorbable subtaler Implant, 11mm Catalog Number: 040011 - Product Usage: The bioBLOCK Implant is indicated for internal support to primary surgical interventions in the treatment of flatfoot.
Temperature indicator may have changed color to dark gray or black prior to shipment. Change in color may indicate the product exposure to temperatures greater than 110¿F
- 2019-04-17FDA-DeviceIntegra LimitedClass IIIntegra bioBLOCK Resorbable subtaler Implant, 10mm Catalog Number: 40010 - Product Usage: The bioBLOCK Implant is indicated for internal support to primary surgical interventions in the treatment of flatfoot.
Temperature indicator may have changed color to dark gray or black prior to shipment. Change in color may indicate the product exposure to temperatures greater than 110¿F
- 2019-04-10FDA-DeviceBeaver VisitecClass IIIMicro Knife 5.0mm 30¿, Part Number 0001522 The Beaver¿ Micro Knife 5.0mm 30¿ is a sharp pointed tip intended for ophthalmic, ENT and other forms of surgery stab incisions and other microsurgical techniques. .
The product contains a misprinted expiration date on the peel pack. Both were printed as "180615", which is the lot number. The Lot number is correct.
- 2019-04-10FDA-DeviceZimmer Biomet, Inc.Class IIBiomet RingLoc Hip System Self-Tapping Bone Screw 6.5mmx30mm-Low Profile Titanium, Accessory to total hip implant Item Number: 103533 Product Usage: Accessory to total hip implant
Two lengths of the 6.5mm RingLoc Hip System Self-Tapping Bone Screw may have the potential for the 30mm screws to be packaged as 20mm screws and vice versa.
- 2019-04-10FDA-DeviceZimmer Biomet, Inc.Class IIBiomet RingLoc Hip System Self-Tapping Bone Screw 6.5mmx20mm-Low Profile Titanium. Accessory to total hip implant Item Number: 103531 Product Usage: Accessory to total hip implant
Two lengths of the 6.5mm RingLoc Hip System Self-Tapping Bone Screw may have the potential for the 30mm screws to be packaged as 20mm screws and vice versa.
- 2019-04-10FDA-DeviceAdvanced Bionics, LLCClass IINa¿da CI Q90 Sound Processor, Silver Gray, Model Cl-5280-140.
The sound processors were loaded with the incorrect firmware.