Medical device recalls
39226 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2019-04-10FDA-DeviceAdvanced Bionics, LLCClass IINa¿da CI Q70 Sound Processor, Sand Beige, Model Cl-5245-120; Silver Gray, Model Cl-5245-140; and Velvet Black, Model Cl-5245-150.
The sound processors were loaded with the incorrect firmware.
- 2019-04-10FDA-DeviceAdvanced Bionics, LLCClass IINa¿da CL Q30 Sound Processor, Sand Beige, Model Cl-5260-120; Silver Gray, Model Cl-5260-140; and Velvet Black, Model Cl-5260-150.
The sound processors were loaded with the incorrect firmware.
- 2019-04-10FDA-DeviceZimmer Biomet, Inc.Class IIBiomet Small Diameter Cement Plug Inserter-Sterile, single-use accessory instrumentation to be used during surgical procedures utilizing the Small Diameter Cement Plug implant devices Item Number: 414989
Lack of an adequate sterilization validation
- 2019-04-10FDA-DeviceCypress Medical Products LLCClass IIIMcKesson Consult Diagnostic Immunochemical Fecal Occult Blood Test, Part numbers 178-CK30T, 178-CT30, 178¿TBSO, 178-CM25, 178-CK50, 178-CM50, 178-CT50, 178¿CK30
Products were denied Clinical Laboratory Improvement Amendments (CLIA) waiver approval by the FDA, due to the packaging being too similar to domestically sourced products under the same branded label.
- 2019-04-10FDA-DeviceConformis, Inc.Class IIiTotal Posterior Stabilized (PS) Knee Replacement System, Catalog Number TPS-111-1111 - 020102
The labeling of the boxes and sterile pouches did not match the contents.
- 2019-04-10FDA-DeviceBeckman Coulter Inc.Class IIUniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. System Name/ Catalog Number: DxC 660i/ A64871, DxC 680i /A64903, DxC 860i /A64935, DxC 880i /A59102
When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
- 2019-04-10FDA-DeviceBeckman Coulter Inc.Class IIThe UniCel DxI 600 and 800 Access Immunoassay Systems, that are connected to an automation line. UniCel DxI 600 Catalog Numbers: A30260, A71460, A71461, A92060. Unicel DxI 800 Catalog Numbers: 973100, A71456, A71457, A84545, A25288, A25285.
When racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
- 2019-04-10FDA-DeviceAbbott Laboratories, IncClass IIAbbott ARCHITECT Creatine Kinase, IVD, REF 7D6341.
The product has a stability issue which may lead to an error code indicating "Unable to calculate result, reaction check failure" and/or Quality Control results out of range low.
- 2019-04-10FDA-DeviceAbbott Laboratories, IncClass IIAbbott ARCHITECT Creatine Kinase, IVD, REF 7D63-21.
The product has a stability issue which may lead to an error code indicating "Unable to calculate result, reaction check failure" and/or Quality Control results out of range low.
- 2019-04-10FDA-DeviceAbbott LaboratoriesClass IICELL-DYN Emerald analyzer, REF 09H39-01. The CELL-DYN Emerald is an automated hematology analyzer intended for in-vitro diagnostic use in the clinical laboratory.
There is a potential for the device to generate Quality Control (QC) low or out-of-range low for parameters RBC and PLT.
- 2019-04-10FDA-DeviceO-Two Medical Technologies, Inc.Class Ie700 Automatic Transport Ventilator, Model Number 01EVE700
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
- 2019-04-10FDA-DeviceO-Two Medical Technologies, Inc.Class Ie600 Automatic Transport Ventilator, Model Number 01EVE600
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
- 2019-04-10FDA-DeviceO-Two Medical Technologies, Inc.Class Ie500 Automatic Transport Ventilator, Model Number 01EVE500
There is a potential for the unit to catch fire due to one of the screen retaining screws being incorrectly inserted. Due to this improper screw insertion the screw sat higher than normal allowing it to contact the underside of the main board causing the screw to rub against the bottom side of the board, wearing away the insulation layer and creating a short circuit.
- 2019-04-03FDA-DeviceCryoLife, Inc.Class IIOn-X Aortic Heart Valve with Anatomic Sewing Ring Product Usage: The On-X Prosthetic Heart Valve is indicated for the replacement of diseased, damaged, or malfunctioning native or prosthetic heart valves in the aortic position. It is a bileaflet mechanical heart valve, which consists of an orifice housing and two leaflets. The On-X¿ Prosthetic Heart Valve is available in 3 aortic configurations. All aortic configurations are available in sizes 19, 21, 23, 25, and 27/29 mm. Aortic valves, siz
On-X Valve was mislabeled with the incorrect serial number.
- 2019-04-03FDA-DeviceICU Medical, Inc.Class IIICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 60" Tubing, Disposable Transducer, 3 ml Intraflo Flush, Macrodrip (Pole Mount), REF: 42641-06, Sterile EO
Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
- 2019-04-03FDA-DeviceICU Medical, Inc.Class IIICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 60" Tubing, Disposable Transducer, 3 ml Squeeze Flush, Macrodrip (Pole Mount), REF: 42640-06, Sterile EO
Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
- 2019-04-03FDA-DeviceICU Medical, Inc.Class IIICU Medical, Transpac IV Monitoring Kit With Safeset Reservoir and Blood Sampling Port, 152 cm (60") Tubing, Disposable Transducer, 03 ml Squeeze Flush Device, Macrodrip (Pole Mount), REF: 01C-42640-06, Sterile EO
Reports of pressure tubing separation between the sampling port and the high pressure tubing on certain lots of extravascular blood pressure transducers, which could lead to delay in therapy, and in rare circumstances, pressure tubing separations could lead to blood loss, air embolism, or contamination of the fluid path.
- 2019-04-03FDA-DeviceMedtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)Class IIIMedtronic Pulmonary Vein Ablation Catheter(R) GOLD (PVAC GOLD) 990078 - Product Usage: The Pulmonary Vein Ablation Catheter GOLD is designed to be used for the creation of endocardial lesions (focal and linear) during cardiac ablation procedures for the treatment of symptomatic Atrial Fibrillation (AF). The PVAC GOLD is also intended to be used for cardiac electrophysiological (EP) mapping of pulmonary vein potentials, delivery of diagnostic pacing stimuli and verifying electrical isolation of t
The firm issued an updated instructions for use to reduce the potential for phrenic nerve injury.
- 2019-04-03FDA-DeviceMedical Action Industries IncClass IILaparotomy Sponge with the following product description and model numbers: 1. Lap 18X18 XR (5 each/pack, 20 pack/case) - Model Number 300, 450, SM40018, V300; 2. Lap 18X18 XR (5 each/pack, 40 pack/case) - Model Number 400, 404, V400, V450; 3. Lap 18X18 XR w/ring (5 each/pack, 40 pack/case) - Model Number 402; 4. Lap 4X18 XR (5 each/pack, 40 pack/case) - Model Number 407; 5. Lap 12X12 XR (5 each/pack, 40 pack/case) - Model Number 411; 6. Lap 12X12 XR w/ring (5 each
Potential lack of sterility assurance
- 2019-04-03FDA-DevicePhilips North America, LLCClass IILithium-ion Rechargeable Batteries for the Philips MX40 Wearable Monitor REF: 989803176201 - Product Usage: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals. The MX40 is to be used primarily as a traditional telemetry medical de
Lithium-ion Rechargeable Batteries for the Philips MX4O Wearable Monitor may have a manufacturing defect that may affect the operation of the monitor
- 2019-04-03FDA-DeviceBeaver VisitecClass IIBVI Visitec Soft Tip Cannula, 0.8MM .50 x 33mm (25G x 1.3 in). The BVI Soft Tip Cannula is intended to irrigate and/or aspirate ophthalmic solutions and liquids in or out of the eye, and remove materials where appropriate (e.g. retinal, subretinal procedures).
Certain lots are missing a protective sheath component over the silicone tip, which may result in the silicone tip breaking or detaching at the metal tube component. If the tip becomes detached during a procedure while in the eye, the fragment will need to be retrieved to mitigate the possibility of injury.
- 2019-04-03FDA-DeviceAccelerate Diagnostics IncClass IIAccelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The Accelerate Pheno system is intended to measure signal intensity of fluorescent probes bound to nucleic acid in target and non-target organisms and to take time-lapse dark-field images of immobilized growing bacterial cells when used with Accelerate PhenoTest kits.
Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to mitigate any risk associated with these rare isolates, the firm has implemented an Expert Rule that suppresses the meropenem result for isolates that display a profile consistent with those that generated a false-susceptible result. The firm has responded to this risk by implementing a set of suppression rules, in a software update, to prevent incorrect results.
- 2019-04-03FDA-DeviceKeystone Dental IncClass IIPaltop 1.25 Hex Drivers, Short Part Number: 60-70102
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
- 2019-04-03FDA-DeviceKeystone Dental IncClass IIPaltop 1.25 Hex Drivers, Long Part Number: 60-70101
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
- 2019-04-03FDA-DeviceKeystone Dental IncClass IIPaltop Fully Guided Surgical Kit- containing the 1.25mm Hex Driver For surgical and restorative applications prosthetic devices, such as artificial teeth Part Number: 30-70021
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
- 2019-04-03FDA-DeviceKeystone Dental IncClass IIPaltop Premium Surgical Kit-containing the 1.25mm Hex Driver For surgical and restorative applications prosthetic devices, such as artificial teeth Part Number: 60-70112
Paltop Surgical Kits contain 1.25mm Hex Driver(s) which may have a slightly oversized hex driving feature, make it difficult to engage screws and abutments containing the corresponding screw feature
- 2019-04-03FDA-DeviceGE Healthcare, LLCClass IIIIntegrated ECG cable with 3-lead leadwires labeled as the following: 1. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, AHA, 3.6 M/12 FT; 2. ECG TRUNK CABLE, 3-LD W/ INTEGRATED GRABBER LEAD WIRE, IEC, 3.6 M/12 FT Product Usage: The reusable ECG trunk cable enables ECG monitoring with GE monitoring equipment requiring rectangular 11-pin connector cable. These accessories are indicated for use by qualified medical personnel only.
The Integrated ECG cable with 3-lead leadwires can short circuit during defibrillation and conduct 25% of the defibrillation energy away from the patient.
- 2019-04-03FDA-DeviceFenwal IncClass IIVacuum Unit for CATSmart Vacuum Pump. Fresenius CATSmart (Continuous Autotransfusion System) is indicated for the processing of blood collected intraoperatively and postoperatively to obtain washed packed red blood cells for reinfusion.
The optional Vacuum Unit may stop working and display the failure message "Failure vacuum unit" during use.
- 2019-04-03FDA-DeviceSANRAI INTERNATIONAL LLCClass IISunset Compressor Nebulizer, Model NEB100 - Product Usage: The Sunset Compressor Nebulizer is an AC-powered air nebulizer compressor system intended to provide a source of compressed air for medical purposes for use in home healthcare. It is to be used with a nebulizer kit to produce medicated aerosol particles for respiratory therapy for both children and adults.
The product was designed for 120V usage and is being used in South Africa with 220V. There is a potential for the device to overheat and the mouthpiece can pop off.
- 2019-04-03FDA-DeviceRandox Laboratories Ltd.Class IIRandox Blood Gas Control Level 2 -Model BG5002. In-vitro Diagnostic Quality control of Blood Gas analysis This product is intended for in vitro diagnostic use, in the quality control of Blood Gas analysis. Target values and ranges are supplied for the following analytes: Calcium, Chloride, Glucose, Lactate, PCO2, pH, pO2, Potassium, Sodium and Total CO2.
An error in the assignment of the control range pH in Blood Gas Control