Medical device recalls
39226 federal recalls on file. Federal recalls for medical devices — surgical equipment, monitors, diagnostic tools, implants, home health devices. Sourced from the FDA.
- 2019-02-06FDA-DeviceEdwards Lifesciences, LLCClass IEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination
These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
- 2019-02-06FDA-DeviceEdwards Lifesciences, LLCClass IEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determine hemodynamic pressures and cardiac output when used with a compatible cardiac output computer. Swan-Ganz thermodilution catheters are indicated for the assessment of a patient s hemodynamic condition through direct intra-cardiac and pulmonary artery pressure monitoring, cardiac output determination, and for infus
These Swan Ganz catheters have incorrect lumen assembly, which causes reversal of the lumens. If the lumens are reversed the clinician may note reverse PA and CVP pressure values and waveforms.
- 2019-01-30FDA-DeviceLeica Microsystems, Inc.Class IILeica M220 F12 surgical microscope The Leica M220 F12 surgical microscope is an optical instrument for improving the visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
The M220 optics may unintentionally drop into the surgical field, risking contact with the patient.
- 2019-01-30FDA-DeviceVarian Medical Systems, Inc.Class IIEclipse(TM) Treatment Planning System with Proton Convolution Superposition algorithm license only The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, elec- tron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatm
There is an anomaly with the Eclipse(TM) Treatment Planning System [TPS] Proton Convolution Superposition [PCS] dose calculation algorithm. The PCS algo- rithm calculates the water equivalent range incorrectly for non-square 3D CT images (either different number of pixels in X and Y, or non-square pixels). Before the dose is calculated, Eclipse resamples the CT images to create a calculation image with a maximum resolution of 256 x 256. The PCS algorithm assumes uniform resolution in X and Y directions, or that X = Y for all images, and erroneously sets the Y = X for dose calcu- lation. For images that are not square, where either X<>Y, or the length of X does not equal the length of Y, the computed water equivalent range R is erroneous compared to the correct range R: 1) X=Y; R =R (R is correct); 2) X<Y: R <R (R is too small) and 3) X>Y R >R (R is too big).
- 2019-01-30FDA-DeviceLivaNova USAClass IIPerfusion Pack, Rx only, Sterile EO
4-way stopcocks and stopcock manifold assembled in certain perfusion tubing systems, indicated for use in surgical procedures requiring extracorporeal support for periods of up to six hours, may leak or be occluded.
- 2019-01-30FDA-DeviceStryker Sustainability SolutionsClass IIBW Lasso 2515 ANV eco Variable Diagnostic EP Catheter, REF D134301 Diagnostic electrophysiology (EP) catheters are specially designed electrode catheters that transmit electrical impulses and can be positioned for endocardial recording or stimulation.
Stryker s Sustainability Solutions division (SSS) has received an increase in reports indicating that an EEPROM chip error code may occur when Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are used with CARTO(R) EP Navigation Systems.
- 2019-01-30FDA-DeviceStihler Electronic GmbhClass IIHeating Profile PRISMAFLO II blood return warmer for Prismaflex Product Code: PF2-WP2618 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
- 2019-01-30FDA-DeviceStihler Electronic GmbhClass IIHeating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP31 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
- 2019-01-30FDA-DeviceStihler Electronic GmbhClass IIHeating Profile PRISMAFLO II S blood return warmer for Prismaflex Product Code: PF2-WP33 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
PRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
- 2019-01-30FDA-DeviceFetzer Medical GmbH & Co. KGClass II12002 / Boss Instruments Simpson OB Forceps, short model, 13" (33cm), Item No./Product Code 46-5040 Fetzer Medical Obstetrical Forceps is a hand-held instruments intended to assist the birth of a foetus during difficult vaginal births through grasping the foetal head in order to facilitate its passage through the birth canal. The instruments are only used for female population. The product is non-powder, non-liquid, non-tablet, capsule or other.
The firm had a report of doctors being unable to pull the obstetric forceps apart.
- 2019-01-30FDA-DeviceFetzer Medical GmbH & Co. KGClass II16600 / Fetzer Surgical De Lee Obstetric Forceps, 11" (30cm), Item No./Product Code 46-5034 Fetzer Medical Obstetrical Forceps is a hand-held instruments intended to assist the birth of a foetus during difficult vaginal births through grasping the foetal head in order to facilitate its passage through the birth canal. The instruments are only used for female population. The product is non-powder, non-liquid, non-tablet, capsule or other.
The firm had a report of doctors being unable to pull the obstetric forceps apart.
- 2019-01-30FDA-DeviceFetzer Medical GmbH & Co. KGClass II12002 / Boss Instruments Ltd. Wrigley Obstetric Forceps, 11" (28cm), Item No./Product Code 46-5004 Fetzer Medical Obstetrical Forceps is a hand-held instruments intended to assist the birth of a foetus during difficult vaginal births through grasping the foetal head in order to facilitate its passage through the birth canal. The instruments are only used for female population. The product is non-powder, non-liquid, non-tablet, capsule or other.
The firm had a report of doctors being unable to pull the obstetric forceps apart.
- 2019-01-30FDA-DeviceFetzer Medical GmbH & Co. KGClass II16600 / Fetzer Surgical Luikart Simpson OB Forceps14" (35.5cm), Item No./Product Code 46-5050 Fetzer Medical Obstetrical Forceps is a hand-held instruments intended to assist the birth of a foetus during difficult vaginal births through grasping the foetal head in order to facilitate its passage through the birth canal. The instruments are only used for female population. The product is non-powder, non-liquid, non-tablet, capsule or other.
The firm had a report of doctors being unable to pull the obstetric forceps apart.
- 2019-01-30FDA-DeviceTeleflex MedicalClass IIPercuvance(R) Percutaneous Shaft 36cm, PCVSHL3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.
There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required.
- 2019-01-30FDA-DeviceTeleflex MedicalClass IIPercuvance(R) Percutaneous Shaft 29cm, PCVSH3 The Percutaneous Surgical System with 5mm attachments is indicated for the means to penetrate Intended Use: soft tissue to access certain areas of the abdomen. The system is used to grasp, manipulate, cut, cauterize and deliver Hem-o-lok ligating clips to soft tissue during laparoscopic surgery.
There is potential for a piece of the distal tip on the Percuvance shaft to become deformed and possibly break off during use. If a piece did break off it could compromise the functionality and security of the tool tip. Should this occur during use, surgical intervention may be required.
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IID Circuit Basic 180, Catalog Number MP01350 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of conden
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IID Circuit Basic 250, Catalog Number MP01348 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of conden
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IID Circuit Watertrap (P) 180, Catalog Number MP01341 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IVentStar Anesthesia WT(P) 180, Catalog Number MP00374 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IVentStar Watertrap (P) 180, Catalog Number MP00361 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IVentStar Basic (P) 250, Catalog Number MP00352 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of cond
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IVentStar Basic (P) 180, Catalog Number MP00351 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of cond
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IVentStar Basic 250, Catalog Number MP00350 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of condensat
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-30FDA-DeviceDraeger Medical, Inc.Class IVentStar Basic 180, Catalog Number MP00349 Breathing Circuits/Anesthesia Sets without a Water trap: Breathing circuit for conduction of respiratory gases between anesthesia machine or ventilator and patient. Intended for single use only. Breathing Circuits/Anesthesia Sets with Water trap: Breathing circuit for conveying breathing gases between an anesthesia delivery system or ventilator and a pediatric patient. Intended for single use only. Water traps are intended for collection of condensa
Incorrectly pre-assembled breathing circuit has potential for incorrect connection/short circuit of the breathing circuit
- 2019-01-23FDA-DeviceCollagen Matrix IncClass IIIPorcine Anorganic Bone Mineral 4.0cc
There is a possibility that the product labeled as 4.0cc volume may only contain 2.0cc volume and is therefore mislabeled.
- 2019-01-23FDA-DeviceKimberly-Clark CorporationClass IIU by Kotex¿ Sleek¿, Regular Tampons, 3 Count
Potential for tampon to come apart upon removal.
- 2019-01-23FDA-DeviceKimberly-Clark CorporationClass IIU by Kotex¿ Sleek¿, Regular/Super Tampons, 34 Count Multipack
Potential for tampon to come apart upon removal.
- 2019-01-23FDA-DeviceKimberly-Clark CorporationClass IIU by Kotex¿ Sleek¿, Regular Tampons, 34 Count
Potential for tampon to come apart upon removal.
- 2019-01-23FDA-DeviceKimberly-Clark CorporationClass IIU by Kotex¿ Sleek¿, Regular Tampons, 18 Count
Potential for tampon to come apart upon removal.
- 2019-01-23FDA-DeviceEthicon, Inc.Class IIPERMA-HAND Silk Suture Size 5, Black 30" (75cm) Double Armed LR Needles, Product Code 485T Product Usage: PERMA-HAND* suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
The sutures inside the packaging are not the same size or type as indicated on the label.