- 2019-01-23FDA-DeviceDeRoyal Industries IncClass IIDeRoyal(R) Enteral Safe Feed Tube, REF 54-2465 A Nasogastric/Oralgastric enteral feeding tube. The MED-RX Feeding Tube is intended to be used for nasogastric/oralgastric enteral feeding as directed by a physician
Enteral Safe Feeding Tubes labeled as 6.5 FR X 24" contained 5 FR X 24" Enteral Safe Feeding Tubes.
- 2019-01-23FDA-DeviceBayer Medical Care, Inc.Class IIMAVIG Portegra or Portegra2 Overhead Counterpoise System (OCS) used in conjunction with the Medrad Stellant CT Injection System (Certegra Workstation), Medrad Mark 7 Arterion" Injection System, or Medrad Mark V ProVis Injection System, Catalog Numbers: Catalog numbers: OCS CEIL (58-S, 58-P, 85-S, 85-P, 100-S, 100-P, 70Si, 97Si), OCS TRACK (58-S, 58-P, 80-S, 80-P), OCS WALL (S, P), OCS (115A, 115CT, 125CT, 125A, 135A, 135CT, 145, 155A, 165A, 115AS, 115CTS), EOC (700, 700S), EOW 700, EOA 700
If the head mount knob is not properly tightened and the operator directly uses the injector head to move the unit, it may become disengaged, causing the injector head to become loose or fall.
- 2019-01-23FDA-DeviceZimmer Biomet, Inc.Class IIMaximum Variable Pitch Compression Screw Caddy, Reference Number 231201003 Product Usage: Intended to carry implants and instruments during regular inventory distribution and handling and sterilization in the hospitals
There is a potential for silicone shedding during cleaning and sterilization.
- 2019-01-23FDA-DeviceZimmer Biomet, Inc.Class IIMaximum Variable Pitch Compression Tray Base, Reference Number 231201002 Product Usage: Intended to carry implants and instruments during regular inventory distribution and handling and sterilization in the hospitals
There is a potential for silicone shedding during cleaning and sterilization.
- 2019-01-23FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IILINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK¿ STEM with Microporous Surface MP¿ Reconstruction Prosthesis, CEMENTLESS System is indicated for
The reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.
- 2019-01-23FDA-DeviceMeridian Bioscience IncClass IIPREMIER Cryptococcal Antigen, Catalog Number 602096 Product Usage: The PREMIER Cryptococcal Antigen enzyme immunoassay (EIA) is a screening or a semi-quantitative test system for the detection of capsular polysaccharide antigens of Cryptococcus neoformans in serum and cerebrosoinal fluid (CSF).
An enzyme reagent included in the kit is not maintaining stability through claimed product expiration, resulting in the potential for false negative results when tested with patient specimens and the positive control.
- 2019-01-23FDA-DeviceConformis, Inc.Class IIiTotal Hip Replacement System, Total Hip Replacement System, Catalog Number HAA-050-D001-020101
The stage 1 and 2 reamer instruments used to prepare fixation for the cup in the acetabulum had a -1mm interference fit with acetabulum bone. The expected interference fit is -2mm.
- 2019-01-23FDA-DeviceSmith & Nephew, Inc.Class IISL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
The pin that holds the spring mechanism is not properly mounted. The pin could potentially disassemble.
- 2019-01-23FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IIBiMobile UHMWPE Liner: ID-28mm/OD-54mm, Reference Number 184-260/04 The LINK BiMobile Dual Mobility System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The device is intended for cemented and cementless use.
The internal labeling included with these implants may include incorrect part numbers and could result in these incorrect labels being added to patient records.
- 2019-01-23FDA-DeviceWaldemar Link GmbH & Co. KG (Mfg Site)Class IIBiMobile UHMWPE Liner: ID-28mm/OD-50mm, Reference Number 184-260/02 The LINK BiMobile Dual Mobility System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The device is intended for cemented and cementless use.
The internal labeling included with these implants may include incorrect part numbers and could result in these incorrect labels being added to patient records.
- 2019-01-23FDA-DeviceVolcano CorporationClass IIVisions PV .035 Digital IVUS Catheter, Catalog Number/Part Number 81234/400-0200.272 Product Usage: The Volcano PV .035 Digital IVUS Catheter is an over the wire imaging catheter with a digital ultrasound transducer on the distal tip. The transducer utilizes a 64-element cylindrical array that radiates acoustic energy into the surrounding tissue and detects the subsequent echoes. The information from the echoes is used to generate real-time images of the peripheral vessels. The devices are p
Visions PV .035 Intravascular Ultrasound (IVUS) Catheters (catalog number 88901 adnd81234) due to the potential inability to pass the guidewire through the y-port of the catheter.
- 2019-01-23FDA-DeviceVolcano CorporationClass IIVisions PV .035 Digital IVUS Catheter, Catalog Number/Part Number 88901/400-0200.173 Product Usage: The Volcano PV .035 Digital IVUS Catheter is an over the wire imaging catheter with a digital ultrasound transducer on the distal tip. The transducer utilizes a 64-element cylindrical array that radiates acoustic energy into the surrounding tissue and detects the subsequent echoes. The information from the echoes is used to generate real-time images of the peripheral vessels. The devices are p
Visions PV .035 Intravascular Ultrasound (IVUS) Catheters (catalog number 88901 adnd81234) due to the potential inability to pass the guidewire through the y-port of the catheter.
- 2019-01-23FDA-DeviceCanadian Hospital SpecialtiesClass IIMED-RX POLYURETHANE FEEDING TUBE For Pediatric Use with Orange Radioplaque Stripe and Enteral Connector, 6.5Fr X 24" (60cm), REF 54-2465-R Product Usage: A Nasogastric/Oralgastric enteral feeding tube. The MED-RX Feeding Tube is intended to be used for nasogastric/oralgastric enteral feeding as directed by a physician.
The product may be packaged with a 5FR X 24 length Feeding tube instead of a 6.5 FR X 24 Feeding tube as labeled.
- 2019-01-23FDA-DeviceArthrex, Inc.Class IIArthrex Suture Anchor, Nano Corkscrew FT (consists of an anchor pre-loaded on a handled inserter including suture, with needles.) Part Number AR-1317FT Product Usage: Soft tissue fixation to bone in the hand and wrist
Potentially lead to anchor breakage during insertion,
- 2019-01-23FDA-DeviceArrow International IncClass IIArrow CVC 2 Lumen, Pediatric Two-Lumen Central Venous Cauterization Set with Blue FlexTip Catheter, 4 Fr 2 Lumen 5cm, Reference # CS-12402 The Arrow CVC is intended to provide short-term (< 30 days) central venous access for treatment of diseases or conditions requiring central venous access, including, but not limited to the following: " Lack of usable peripheral IV sites " Central venous pressure monitoring " Total parenteral nutrition (TPN) " Infusions of fluids, medications, or chemothera
The lidstock states the incorrect priming volume and flow rates.
- 2019-01-16FDA-DeviceZimmer Biomet, Inc.Class IIBiomet DVR Crosslock Screws- 2.7mm x18mm, Sterile ITEM 131827118 Product Usage: Usage: The system is intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions (osteotomies) of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone
One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screws
- 2019-01-16FDA-DeviceZimmer Biomet, Inc.Class IIBiomet DVR Crosslock Screws- 2.7mm x 20mm, Sterile ITEM 131827120 Product Usage: The system is intended for stabilization and fixation of small bone fragments in fresh fractures, revision procedures, joint fusion and reconstructions (osteotomies) of small bones of the hand, foot, radius, ulna, ankle, humerus, scapula and pelvis, particularly in osteopenic bone
One lot of product labeled 2.7mm x18mm Lock Screw Square Drive (Lot 029970) contains 20mm locking screws and one lot of product labeled 2.7mm x 20mm Lock Screw Square Drive (Lot 029980) contains 18mm locking screws
- 2019-01-16FDA-DeviceSiemens Medical Solutions USA, Inc.Class IISiemens Biograph Horizon PET/CT System, Product Usage: The Siemens Biograph systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.
Improper design or specifications.
- 2019-01-16FDA-DeviceElekta LimitedClass IIElekta Unity
In rare circumstances, due to reasons such as inadequate hospital facility maintenance, water leakage, and weakening of speaker fixings during the lifetime of the system, the speakers can become dislodged and be pulled into the bore.
- 2019-01-16FDA-DeviceMevion Medical Systems, Inc.Class IIMEVION S250i
Treatment is allowed to continue (via partials) in situations where dDose1 interlock has tripped
- 2019-01-16FDA-DeviceGE Healthcare, LLCClass IIVariCam
A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.
- 2019-01-16FDA-DeviceGE Healthcare, LLCClass IIDiscover VH/Millennium VG Nuclear Medicine Imaging System
A potential hazardous situation may occur if the collimator locking handle is not properly placed in the lock position and the warning circuit does not detect that the collimator is unlocked because of the patient proximity to the collimator during a clinical scan.
- 2019-01-16FDA-DeviceCardinal Health Inc.Class IIPresource Kit Total Joint Pack containing a double-wrapped basin component. Product Usage: This kit is used on patients undergoing joint replacement knee or hip.
The double wrapping of this custom kit basin set is not supported by the current sterility assurance validation.
- 2019-01-16FDA-DeviceBoston Scientific CorporationClass IIIIntellaNav MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or old
Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
- 2019-01-16FDA-DeviceBoston Scientific CorporationClass IIIIntellaTip MiFi Open-Irrigated Ablation Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or old
Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
- 2019-01-16FDA-DeviceBoston Scientific CorporationClass IIIBlazer Open-Irrigated Ablation Batch Catheter Product Usage: Open Irrigated ablation catheters, when used with a compatible Radiofrequency (RF) Controller and Irrigation Pump, are indicated for: Cardiac electrophysiological mapping Delivering diagnostic pacing stimuli RF ablation of sustained or recurrent type 1 Atrial Flutter in patients age 18 years or older Treatment of drug refractory, recurrent, symptomatic, paroxysmal atrial fibrillation (PAF) in patients age 18 years or olde
Supplier had changed its resin sub-suppliers three (3) times without notifying BSC. These changes resulted in slight variations in resin material compositions which were not reflected in BSC s internal product documentation. Not approved in customer document.
- 2019-01-16FDA-DeviceIon Beam Applications S.A.Class IIProteus 235 Product Usage: The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.
There is a risk of detachment of the Compact Gantry Rolling Floor.
- 2019-01-16FDA-DeviceSwissray Medical AGClass IIddRCombi Plus FP, Model Number 8.2000.0070.0 (US Version), 8.2000.0060.0 (Europe Version) The ddRCombi Plus FP is a fully digital, multifunctional X-ray system that allows the direct digitization of X-ray information without use of cassettes or other media. It can perform all applications in orthopedics and trauma as well as pulmonary imaging.
A component of the device was recalled by a supplier. In rare cases of inadequate maintenance or very high clinical workload, the first steel wire rope of the lifting column, which is designed to take the load, can break without triggering the safety lock. This can lead to overloading of the safety cable, which is not designed to permanently carry the load of the lifting column or continuously move under load. This could cause material fatigue and cause the arm to drop unexpectedly during patient positioning. If the arm drops unexpectedly, patients and users may be seriously injured.
- 2019-01-16FDA-DeviceEthicon Endo-Surgery IncClass IIPROXIMATE¿ PPH Hemorrhoidal Circular Stapler Product Code:PPH03 PPH03: The PROXIMATE¿ PPH Hemorrhoidal Circular Stapler and Accessories have application throughout the anal canal to perform surgical treatment of hemorrhoidal disease
Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired
- 2019-01-16FDA-DeviceEthicon Endo-Surgery IncClass IIPROXIMATE¿ HCS Hemorrhoidal Circular Stapler Product Code: PPH01 PPH01: The PROXIMATE¿ HCS Hemorrhoidal Circular Stapler and Accessories have application for general surgical treatment of anorectal wall defects by means of transanal stapling and resection of mucosal and musculomucosal tissue.
Hemorrhoidal Circular Stapler may have been assembled without a washer, this may potentially cause bleeding or soft tissue injury during use as the device may not fully cut when fired